Microinjection of living adherent cells by using a semi-automatic microinjection system

被引:21
|
作者
Viigipuu, K [1 ]
Kallio, P [1 ]
机构
[1] Tampere Univ Technol, Inst Automat & Control, FIN-33101 Tampere, Finland
来源
关键词
adherent cells; capillary pressure microinjection; cell curvival rate; injection success rate; micromanipulator; penetration success rate; semi-automatic microinjection;
D O I
10.1177/026119290403200414
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
Testing in vitro is an alternative to animal experimentation. The capillary pressure microinjection technique is a supporting technology for efficient in vitro testing. The main benefit of the technique is the possibility of injecting large molecules into a single living cell. The ultimate goal of the research discussed in this paper is to increase the cell survival rate in capillary pressure microinjection. A method to reliably evaluate cell survival rate is therefore needed. A three-phase evaluation process is presented in this paper. The first phase determines the success rate of the injection capillary to penetrate the cell membrane. The second phase studies the success rate of delivering the injection substance inside the cell, while the third phase studies cell survival after the microinjection. In addition to the three-phase evaluation process, this paper describes the initial results of penetration and injection tests performed by using a semi-automatic capillary pressure microinjection system developed by the research group. Three adherent cell lines, namely, retinal pigment epithelia] cells, MCF-7 human breast cancer cells and SH-SY5Y neuroblastoma cells, were used in the experiments. The results of the penetration tests show that the average success rate of penetrating the cell membrane using the micromanipulator was 87%. The goal of the injection tests was to demonstrate the successful microinjection of living cells and to study the injection success rate. Fluorescein dextran was injected into MCF-7 cells, and preliminary results showed an injection success rate of 49%. In the survival tests, the neuronal cells were microinjected with KCl. During long-term observation after the microinjection, the microinjected cells first decreased their adhesion to the plate, but later adhered to the bottom of the plate and even grew some dendrites. In the next phase of the study, more tests will be performed in order to obtain a statistically reliable value for the survival rate.
引用
收藏
页码:417 / 423
页数:7
相关论文
共 50 条
  • [1] Electronic Controller Design of a Semi-Automatic Cell Microinjection System
    Hameed, Asad
    Hasan, Osman
    2019 13TH ANNUAL IEEE INTERNATIONAL SYSTEMS CONFERENCE (SYSCON), 2019,
  • [2] A Fully Automated System for Adherent Cells Microinjection
    Becattini, Gabriele
    Mattos, Leonardo S.
    Caldwell, Darwin G.
    IEEE JOURNAL OF BIOMEDICAL AND HEALTH INFORMATICS, 2014, 18 (01) : 83 - 93
  • [3] A system for high-speed microinjection of adherent cells
    Wang, Wenhui
    Sun, Yu
    Zhang, Ming
    Anderson, Robin
    Langille, Lowell
    Chan, Warren
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2008, 79 (10):
  • [4] A visual targeting system for the microinjection of unstained adherent cells
    Becattini, Gabriele
    Mattos, Leonardo S.
    Caldwell, Darwin G.
    COMPUTERS IN BIOLOGY AND MEDICINE, 2013, 43 (02) : 109 - 120
  • [5] Automated High-Productivity Microinjection System for Adherent Cells
    Pan, Fei
    Chen, Shuxun
    Jiao, Yang
    Guan, Zhangyan
    Shakoor, Adnan
    Sun, Dong
    IEEE ROBOTICS AND AUTOMATION LETTERS, 2020, 5 (02) : 1167 - 1174
  • [6] Automatic microinjection system using stereoscopic microscope
    Yamamoto, H
    Sano, T
    Hasebe, S
    IDAACS'2001: PROCEEDINGS OF THE INTERNATIONAL WORKSHOP ON INTELLIGENT DATA ACQUISITION AND ADVANCED COMPUTING SYSTEMS: TECHNOLOGY AND APPLICATION, 2001, : 83 - 86
  • [7] MICROINJECTION OF GELSOLIN INTO LIVING CELLS
    COOPER, JA
    BRYAN, J
    FRIEDEN, C
    LOFTUS, DJ
    SCHWAB, B
    ELSON, EL
    JOURNAL OF CELL BIOLOGY, 1986, 103 (05): : A286 - A286
  • [8] MICROINJECTION OF GELSOLIN INTO LIVING CELLS
    COOPER, JA
    BRYAN, J
    SCHWAB, B
    FRIEDEN, C
    LOFTUS, DJ
    ELSON, EL
    JOURNAL OF CELL BIOLOGY, 1987, 104 (03): : 491 - 501
  • [9] Adherent Cells Localization Based on Gaussian Filtering in Automated Microinjection System
    Gao, Ping
    Huang, Haibo
    Chen, Guodong
    Chen, Liguo
    Chen, Tao
    2013 INTERNATIONAL CONFERENCE ON MANIPULATION, MANUFACTURING AND MEASUREMENT ON THE NANOSCALE (3M-NANO), 2013, : 166 - 169
  • [10] A Hyperelastic Model for Mechanical Responses of Adherent Cells in Microinjection
    Shen, Tianyao
    Shirinzadeh, Bijan
    Zhong, Yongmin
    Smith, Julian