Neutron dose rate evaluation for medical linear accelerators

被引:8
|
作者
Facure, A
Falcao, RC
Da Silva, AX
Crispim, VR
机构
[1] Comis Nacl Energia Atom, BR-22294900 Rio De Janeiro, Brazil
[2] UFRJ, CT, EE, DNC,COPPE,PEN, BR-21945970 Rio De Janeiro, Brazil
关键词
D O I
10.1093/rpd/nch369
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
During X-ray therapeutic irradiation with energies above the threshold of (X,n) reactions in the structural materials of medical accelerators, a photoneutron fluence is generated. In Brazil, no measurements of neutron doses in radiotherapy rooms are being done yet, when licensing these equipment. Consequently, it is very important to obtain accurate analytical formulae and/or simulation of these dose rates, in order to estimate the increase in dose received by the patient and staff, as well as to correctly project the additional shielding for the treatment room. In this work, we present MCNP simulation of dosimetric quantities at the isocentre of some models of high-energy linear accelerators, and compare it with the values given by the manufacturers, finding good agreement between both.
引用
收藏
页码:101 / 103
页数:3
相关论文
共 50 条
  • [31] Photon skyshine from medical linear accelerators
    McDermott, Patrick N.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2020, 21 (03): : 108 - 114
  • [32] THE TREATMENT HEAD DESIGN FOR MEDICAL LINEAR ACCELERATORS
    TAUMANN, L
    BULLETIN OF THE AMERICAN PHYSICAL SOCIETY, 1980, 25 (07): : 803 - 803
  • [33] ELECTRON BEAM MATCHING FOR MEDICAL LINEAR ACCELERATORS
    Behrens, C.
    Bjelkengren, U.
    Ottosson, W.
    Sjostrom, D.
    RADIOTHERAPY AND ONCOLOGY, 2008, 88 : S384 - S385
  • [34] Quality control of modern medical linear accelerators
    Tsitovich, Y., V
    Tarutin, I. G.
    Kiselev, M. G.
    DEVICES AND METHODS OF MEASUREMENTS, 2011, (01): : 112 - 116
  • [35] RADIOACTIVITY IN THE SURROUNDING AIR OF MEDICAL LINEAR ACCELERATORS
    EWEN, K
    STRAHLENTHERAPIE UND ONKOLOGIE, 1989, 165 (08): : 617 - 621
  • [36] Beam spot motion of medical linear accelerators
    Yeboah, C
    Challacombe, J
    Pang, G
    O'Brien, P
    MEDICAL PHYSICS, 2005, 32 (07) : 2411 - 2411
  • [37] Evaluation of neutron production in new accelerators for radiotherapy
    Martinez, S. A.
    Barquero, R.
    Gomez-Ros, J. M.
    Lallena, A. M.
    Andres, C.
    Tortosa, R.
    RADIATION MEASUREMENTS, 2010, 45 (10) : 1402 - 1405
  • [38] An Evaluation of the Overlap Between the Acceptance Testing and Commissioning Processes for Conventional Medical Linear Accelerators
    Morrow, A.
    Rangaraj, D.
    Perez-Andujar, A.
    Krishnamurthy, N.
    MEDICAL PHYSICS, 2016, 43 (06) : 3572 - 3572
  • [39] DOSE AND GAMMA-RAY SPECTRA FROM NEUTRON-INDUCED RADIOACTIVITY IN MEDICAL LINEAR ACCELERATORS FOLLOWING HIGH-ENERGY TOTAL BODY IRRADIATION
    Keehan, S.
    Taylor, M. L.
    Smith, R. L.
    Dunn, L.
    Kron, T.
    Franich, R. D.
    RADIATION PROTECTION DOSIMETRY, 2016, 172 (04) : 327 - 332
  • [40] Neutron dose rate in the facility at the Cyclotron Center of Chung Shan Medical University
    Lee, Jui-Ping
    Chen, Chien-Yi
    JOURNAL OF RADIATION RESEARCH, 2008, 49 (02) : 147 - 151