Biomechanical Evaluation of Strength and Stiffness of Subtalar Joint Arthrodesis Screw Constructs

被引:16
|
作者
Jastifer, James R. [1 ]
Alrafeek, Saif [2 ]
Howard, Peter [3 ]
Gustafson, Peter A. [2 ]
Coughlin, Michael J. [4 ]
机构
[1] Borgess Orthoped, 2490 South 11th St, Kalamazoo, MI 49009 USA
[2] Western Michigan Univ, Kalamazoo, MI 49008 USA
[3] Western Michigan Univ, Homer Stryker MD Sch Med, Kalamazoo, MI 49008 USA
[4] St Alphonsus Coughlin Foot & Ankle Clin, Boise, ID USA
关键词
subtalar arthrodesis; subtalar arthritis; subtalar biomechanics; INTRAARTICULAR CALCANEAL FRACTURES; ANKLE ARTHRODESIS; OS CALCIS; LATE COMPLICATIONS; BONE-GRAFT; FUSION; FIXATION; HINDFOOT; MOTION; MODEL;
D O I
10.1177/1071100715619680
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Subtalar arthrodesis is a common treatment for end-stage subtalar joint arthritis as well as many other clinical problems. The best method of subtalar arthrodesis fixation is unknown. The purpose of this study was to compare the strength of subtalar arthrodesis fixation methods including a single posterior screw (SP), 2 posterior minimally divergent screws (MD) and a 2 screw highly divergent screw (HD) construct for subtalar arthrodesis. Methods: A biomechanical study was performed including the three different screw configurations (SP, MD, HD). These surrogate bone specimens were subjected to applied inversion and eversion torques about the subtalar joint axis on a servo-hydraulic load frame. Torsional stiffness of the construct and the maximum torque for each configuration were measured. Additionally, a cadaver study was performed using 5 fresh-frozen cadaver specimens. The perpendicular distance from the divergent screw guide-wire placement was measured from anatomic structures. Results: The HD screw configuration was found to have the highest torsional stiffness in both inversion and eversion, followed by the MD construct and then the SP construct. Similarly, the HD construct had the highest maximum torque versus the MD and SP constructs. All between-group differences were statistically significant (P < .05). The mean distance from key structures to the divergent screw included the sural nerve (13 mm), peroneus brevis tendon (18 mm), tibialis anterior tendon (8 mm), and tibialis posterior tendon (21 mm). Conclusion: This biomechanical and cadaver study supports the use of 2 screws for fixation of subtalar arthrodesis over a single posterior screw. Additionally, we describe a biomechanically superior and potentially safe, alternative 2-screw divergent construct. Clinical Relevance: This study gives biomechanical support for 2 screw, divergent fixation of subtalar arthrodesis or a single over a single screw or two screw minimally divergent construct.
引用
收藏
页码:419 / 426
页数:8
相关论文
共 50 条
  • [31] Biomechanical evaluation of a second generation headless compression screw for ankle arthrodesis in a cadaver model
    Somberg, Andrew Max
    Whiteside, William K.
    Nilssen, Erik
    Murawski, Daniel
    Liu, Wei
    [J]. FOOT AND ANKLE SURGERY, 2016, 22 (01) : 50 - 54
  • [32] Is Subtalar Joint Cartilage Resection Necessary for Tibiotalocalcaneal Arthrodesis via Intramedullary Nail? A Multicenter Evaluation
    Mulhern, Jennifer L.
    Protzman, Nicole M.
    Levene, Maxwell J.
    Martin, Scott M.
    Fleming, Justin J.
    Clements, J. Randolph
    Brigido, Stephen A.
    [J]. JOURNAL OF FOOT & ANKLE SURGERY, 2016, 55 (03): : 572 - 577
  • [33] FURTHER EXPERIENCE WITH EXTRA-ARTICULAR ARTHRODESIS OF THE SUBTALAR JOINT
    GRICE, DS
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1955, 37 (02): : 246 - &
  • [34] Effect of Arthrodesis Device Type and Trajectory on Subtalar Joint Compression
    Beals, Caitlyn J.
    Dupont, Kenneth M.
    Gross, Christopher E.
    Safranski, David L.
    [J]. JOURNAL OF FOOT & ANKLE SURGERY, 2023, 62 (05): : 812 - 815
  • [35] Subtalar Joint Arthrodesis for Elective and Posttraumatic Foot and Ankle Deformities
    DiDomenico, Lawrence A.
    Butto, Danielle N.
    [J]. CLINICS IN PODIATRIC MEDICINE AND SURGERY, 2017, 34 (03) : 327 - +
  • [36] Subtalar Joint Arthrodesis Open and Arthroscopic Indications and Surgical Techniques
    Roster, Brent
    Kreulen, Christopher
    Giza, Eric
    [J]. FOOT AND ANKLE CLINICS, 2015, 20 (02) : 319 - +
  • [37] Optimizing Subtalar Arthrodesis: A Human Cadaveric Evaluation of a Novel Partially-Threaded Screw Combination in the Delta Configuration
    Raykov, Georgi
    Ivanov, Stoyan
    Gueorguiev, Boyko
    Pastor, Tatjana
    Berk, Till
    Pastor, Torsten
    Zderic, Ivan
    [J]. MEDICINA-LITHUANIA, 2024, 60 (06):
  • [38] Fixation of Anteromedial Coronoid Facet Fractures: A Biomechanical Evaluation of Plated Versus Screw Constructs
    Morellato, John
    Louati, Hakim
    Desloges, William
    Papp, Steven
    Pollock, J. Whitcomb
    [J]. JOURNAL OF ORTHOPAEDIC TRAUMA, 2018, 32 (11) : E451 - E456
  • [39] Erratum to: Optimal double screw configuration for subtalar arthrodesis: a finite element analysis
    Jun Young Lee
    Yeon Soo Lee
    [J]. Knee Surgery, Sports Traumatology, Arthroscopy, 2011, 19 : 867 - 867
  • [40] Accurate screw arthrodesis of the distal interphalangeal joint
    Pickering, G. T.
    Barnes, J. R.
    Bhatia, R.
    [J]. ANNALS OF THE ROYAL COLLEGE OF SURGEONS OF ENGLAND, 2014, 96 (03) : 245 - 246