Repeatability and Reproducibility of Retinal Blood Flow Measurement Using a Doppler Optical Coherence Tomography Flowmeter in Healthy Subjects

被引:16
|
作者
Tani, Tomofumi [1 ]
Song, Young-Seok [1 ]
Yoshioka, Takafumi [1 ]
Omae, Tsuneaki [1 ]
Ishibazawa, Akihiro [1 ]
Akiba, Masahiro [2 ]
Yoshida, Akitoshi [1 ]
机构
[1] Asahikawa Med Univ, Dept Ophthalmol, Midorigaoka Higashi 2-1-1, Asahikawa, Hokkaido 0788510, Japan
[2] Topcon Corp, R&D Div, Tokyo, Japan
关键词
retinal blood flow; Doppler OCT flowmeter; laser Doppler velocimetry; retinal blood velocity profile; SHEAR-STRESS; ENDOTHELIAL-CELLS; CIRCULATION; VELOCIMETRY; VELOCITY; SPEED;
D O I
10.1167/iovs.16-21389
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE. To evaluate the repeatability and reproducibility of retinal blood flow (RBF) measurements in humans by using new auto-alignment and measurement software in a commercially available Doppler optical coherence tomography (DOCT) system. METHODS. The DOCT flowmeter assessed the intrasession repeatability and the intersession and interobserver reproducibility of the RBF measurements. For intrasession repeatability, the coefficients of variation (CVs) of five repeated RBF measurements were calculated at the retinal arteries and veins in 20 normal eyes of 20 healthy volunteers. For intersession reproducibility, two sets of three measurements obtained by one observer on 2 different days were compared. For interobserver reproducibility, two sets of three measurements obtained by two observers on the same day were compared. Intraclass correlation coefficients (ICCs) also were used to evaluate the repeatability and reproducibility. The relevance of the DOCT flowmeter and laser Doppler velocimetry (LDV) also was assessed. RESULTS. Regarding intrasession repeatability, the ICC of the RBF exceeded 0.90 in arterioles and venules (ICC: 0.994 and 0.970, respectively). The CVs of the RBF in the arterioles and venules were 6.0% +/- 3.4% and 8.8% +/- 5.1%, respectively. The intersession and interobserver RBF values had high reproducibility in the arterioles (ICC: 0.980 and 0.993, respectively) and venules (ICC: 0.982 and 0.986, respectively). The RBF measured with the DOCT flowmeter was correlated strongly with LDV in the arterioles (r = 0.76; P < 0.001). CONCLUSIONS. The DOCT flowmeter had good reproducibility in the arterioles and venules and precisely measured the RBF as compared to the LDV in the arterioles.
引用
收藏
页码:2891 / 2898
页数:8
相关论文
共 50 条
  • [1] Evaluation of Retinal Arteriosclerosis using Doppler Optical Coherence Tomography Flowmeter in Healthy Subjects
    Murono, Masataka
    Tani, Tomofumi
    Song, Youngseok
    Sogawa, Kenji
    Yoshioka, Takafumi
    Takahashi, Kengo
    Wada, Takanari
    Akiba, Masahiro
    Sakai, Jun
    Nakamura, Syunsuke
    Minamide, Kana
    Ishiko, Satoshi
    Yanagi, Yasuo
    Yoshida, Akitoshi
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2019, 60 (09)
  • [2] Measurement of retinal blood flow using a Doppler optical coherence tomography auto-flowmeter in humans
    Yoshida, Akitoshi
    Tani, Tomofumi
    Yoshioka, Takafumi
    Sogawa, Kenji
    Nakabayashi, Seigo
    Ishibazawa, Akihiro
    Omae, Tsuneaki
    Song, Youngseok
    Sakai, Jun
    Akiba, Masahiro
    Nagaoka, Taiji
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2016, 57 (12)
  • [3] Repeatability and reproducibility of automatic segmentation of retinal layers in healthy subjects using Spectralis optical coherence tomography
    Cetinkaya, Ersan
    Duman, Rahmi
    Duman, Resat
    Sabaner, Mehmet Cem
    [J]. ARQUIVOS BRASILEIROS DE OFTALMOLOGIA, 2017, 80 (06) : 378 - 381
  • [4] Retinal Blood Flow Biomarkers in Healthy Human Subjects Measured with Clinical Doppler Optical Coherence Tomography
    Otero-Marquez, Oscar
    Ahsanuddin, Sofia
    Migacz, Justin V.
    Minamide, Kana
    Akiba, Masahiro
    Harris, Alon
    Chui, Toco Yuen Ping
    Rosen, Richard B.
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2022, 63 (07)
  • [5] Reproducibility of Retinal Thickness Measurements in Healthy Subjects Using Spectralis Optical Coherence Tomography
    Menke, Marcel N.
    Dabov, Simeon
    Knecht, Pascal
    Sturm, Veit
    [J]. AMERICAN JOURNAL OF OPHTHALMOLOGY, 2009, 147 (03) : 467 - 472
  • [6] Retinal blood flow measurement by using optical coherence tomography
    Makita, Shuichi
    Fabritius, Tapio
    Miura, Masahiro
    Yatagai, Toyohiko
    Yasuno, Yoshiaki
    [J]. COHERENCE DOMAIN OPTICAL METHODS AND OPTICAL COHERENCE TOMOGRAPHY IN BIOMEDICINE XII, 2008, 6847
  • [7] EVALUATION OF RETINAL BLOOD FLOW VELOCITY PROFILES USING SPECTRAL-DOMAIN DOPPLER OPTICAL COHERENCE TOMOGRAPHY IN HEALTHY SUBJECTS
    Nagaoka, Taiji
    Tani, Tomofumi
    Ishibazawa, Akihiro
    Nakabayashi, Seigo
    Omae, Tsuneaki
    Sogawa, Kenji
    Yoshida, Akitoshi
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2014, 55 (13)
  • [8] Variability and Repeatability of Quantitative, Fourier-Domain Optical Coherence Tomography Doppler Blood Flow in Young and Elderly Healthy Subjects
    Tayyari, Faryan
    Yusof, Firdaus
    Vymyslicky, Michal
    Tan, Ou
    Huang, David
    Flanagan, John G.
    Hudson, Christopher
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2014, 55 (12) : 7716 - 7725
  • [9] Deterioration of Retinal Blood Flow Parameters in Branch Retinal Vein Occlusion Measured by Doppler Optical Coherence Tomography Flowmeter
    Takahashi, Kengo
    Song, Youngseok
    Sogawa, Kenji
    Yoshioka, Takafumi
    Tani, Tomofumi
    Ishiko, Satoshi
    Yoshida, Akitoshi
    [J]. JOURNAL OF CLINICAL MEDICINE, 2020, 9 (06) : 1 - 8
  • [10] In vivo retinal blood flow measurement by Fourier domain Doppler optical coherence tomography
    Wang, Yimin
    Tan, Ou
    Huang, David
    [J]. COHERENCE DOMAIN OPTICAL METHODS AND OPTICAL COHERENCE TOMOGRAPHY IN BIOMEDICINE XII, 2008, 6847