Axial vibration of threaded external fixation pins in vivo: Detection of pin instability and pin tract lysis

被引:0
|
作者
Kay, MW [1 ]
Roe, SC [1 ]
Abrams, F [1 ]
机构
[1] N Carolina State Univ, Dept Biol & Agr Engn, Raleigh, NC 27695 USA
关键词
D O I
暂无
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The objective of this study was to investigate the ability of a non-destructive vibrational method to detect bone lysis at the external fixator pin-bone interface. The right hind leg of 12 dogs was fractured and stabilized with a 4 pin unilateral frame. The resonant frequency of each pin was measured and radiographs were taken at 0, 2, 4, 6, 8, and 10 weeks post-op. At 11 weeks dogs were sacrificed and quasi-static stiffness was measured for each pin. There was good correlation between frequency and quasi-static stiffness. The vibrational method was unsuccessful 28% of the time in detecting the changes in bone quality around a pin that had been determined using standard radiographs. Resonant frequency was not sensitive to small changes in pin tract bone quality that had not resulted in a significant decrease in pin stability. It was concluded that the vibrational method used in this investigation was not an improvement over standard radiographs for diagnosing pin tract bone lysis. However, the vibrational method was more sensitive to changes at the pinbone interface than low-load quasi-static tests.
引用
收藏
页码:485 / 486
页数:2
相关论文
共 31 条
  • [21] Pin-tract myiasis after external bone fixation: A case report and review of literature
    Lageju, Nimesh
    Neupane, Durga
    Jaiswal, Lokesh Shekher
    Phuyal, Upama
    INTERNATIONAL JOURNAL OF SURGERY CASE REPORTS, 2022, 95
  • [22] Routine pin tract care in external fixation is unnecessary: A randomised, prospective, blinded controlled study
    Camathias, C.
    Valderrabano, V.
    Oberli, H.
    SWISS MEDICAL WEEKLY, 2011, 141 : 7S - 7S
  • [23] Percutaneous external fixator pins with bactericidal micron-thin sol-gel films for the prevention of pin tract infection
    Qu, Haibo
    Knabe, Christine
    Radin, Shula
    Garino, Jonathan
    Ducheyne, Paul
    BIOMATERIALS, 2015, 62 : 95 - 105
  • [24] Antimicrobial efficacy of external fixator pins coated with a lipid stabilized hydroxyapatite/chlorhexidine complex to prevent pin tract infection in a goat model
    DeJong, ES
    DeBerardino, TM
    Brooks, DE
    Nelson, BJ
    Campbell, AA
    Bottoni, CR
    Pusateri, AE
    Walton, RS
    Guymon, CH
    McManus, AT
    JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 2001, 50 (06): : 1008 - 1013
  • [25] Safety of the limited open technique of bone-transfixng threaded-pin placement for external fixation of distal radial fractures: A cadaver study
    Hassan, DM
    Johnston, GHF
    CANADIAN JOURNAL OF SURGERY, 1999, 42 (05) : 363 - 365
  • [26] Antimicrobial efficacy of external fixator pins coated with a lipid stabilized hydroxyapatite/chlorhexidine complex to prevent pin tract infection in a goat model - Discussion
    Browner, BD
    DeJong, ES
    JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 2001, 50 (06): : 1013 - 1014
  • [27] Hydroxyapatite-coated compared with stainless steel external fixation pins did not show impact in the rate of pin track infection: a multicenter prospective study
    Stoffel, Cristhopher
    de Lima, Everton
    Salles, Mauro Jose
    INTERNATIONAL ORTHOPAEDICS, 2023, 47 (05) : 1163 - 1169
  • [28] Hydroxyapatite-coated compared with stainless steel external fixation pins did not show impact in the rate of pin track infection: a multicenter prospective study
    Cristhopher Stoffel
    Everton de Lima
    Mauro José Salles
    International Orthopaedics, 2023, 47 : 1163 - 1169
  • [29] Designing of double-threaded intramedullary pin and evaluation of ex vivo biomechanical resistance to axial compression load on the canine long bone fracture gap model
    Vora, Shruti
    Kumar, Ashwani
    Kaur, Kirandip
    INDIAN JOURNAL OF ANIMAL SCIENCES, 2022, 92 (07): : 819 - 824
  • [30] Improvement of the bone-pin interface strength in osteoporotic bone with use of hydroxypatite-coated tapered external-fixation pins - A prospective, randomized clinical study of wrist fractures
    Moroni, A
    Faldini, C
    Marchetti, S
    Manca, M
    Consoli, V
    Giannini, S
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2001, 83A (05): : 717 - 721