The initial stability and contact mechanics of a press-fit resurfacing arthroplasty of the hip

被引:45
|
作者
Udofia, I.
Liu, F.
Jin, Z.
Roberts, P.
Grigoris, P.
机构
[1] Univ Leeds, Inst Med & Biol Engn, Sch Mech Engn, Leeds LS2 9JT, W Yorkshire, England
[2] Royal Gwent Hosp, Newport NP20 2UB, Gwent, Wales
[3] Univ Bradford, Sch Engn Design & Technol, Bradford BD7 1DP, W Yorkshire, England
来源
关键词
D O I
10.1302/0301-620X.89B4.18055
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Finite element analysis was used to examine the initial stability after hip resurfacing and the effect of the procedure on the contact mechanics at the articulating surfaces. Models were created with the components positioned anatomically and loaded physiologically through major muscle forces. Total micromovement of less than 10 mu m was predicted for the press-fit acetabular components models, much below the 50 mu m limit required to encourage osseointegration. Relatively high compressive acetabular and contact stresses were observed in these models. The press-fit procedure showed a moderate influence on the contact mechanics at the bearing surfaces, but produced marked deformation of the acetabular components. No edge contact was predicted for the acetabular components studied. It is concluded that the frictional compressive stresses generated by the 1 mm to 2 mm interference-fit acetabular components, together with the minimal micromovement, would provide adequate stability for the implant, at least in the immediate post-operative situation.
引用
收藏
页码:549 / 556
页数:8
相关论文
共 50 条
  • [1] Press-fit hip arthroplasty - A European alternative
    Romagnoli, S
    JOURNAL OF ARTHROPLASTY, 2002, 17 (04): : 108 - 112
  • [2] Global fit concept in revision hip arthroplasty for cementless press-fit femoral stems
    Canovas, F.
    LeBeguec, P.
    Batard, J.
    Gaillard, F.
    Dagneaux, L.
    ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2017, 103 (04) : 579 - 581
  • [3] IMPROVING THE FIT OF PRESS-FIT HIP STEMS
    ROBERTSON, DD
    WALKER, PS
    HIRANO, SK
    ZHOU, XM
    GRANHOLM, JW
    POSS, R
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1988, (228) : 134 - 140
  • [4] Influence of press-fit parameters on the primary stability of uncemented femoral resurfacing implants
    Gebert, A.
    Peters, J.
    Bishop, N. E.
    Westphal, F.
    Morlock, M. M.
    MEDICAL ENGINEERING & PHYSICS, 2009, 31 (01) : 160 - 164
  • [5] Outcomes of press-fit uncemented versus cemented hip arthroplasty in the oncologic patient
    Larsen, Christopher G.
    Crockatt, William K.
    Fitzgerald, Michael
    Matos, Nanette
    Goodman, Howard J.
    Kenan, Samuel
    Kenan, Shachar
    JOURNAL OF ORTHOPAEDICS, 2020, 22 : 198 - 202
  • [6] New instruments for socket preparation, and initial stability of press-fit cups
    Pitto, RP
    Hofmeister, V
    BIOMEDIZINISCHE TECHNIK, 1999, 44 (04): : 92 - 97
  • [7] Initial stability of press-fit acetabular cups - An in-vitro study
    Wetzel, R
    Simnacher, M
    Scheller, G
    BIOMEDIZINISCHE TECHNIK, 2005, 50 (12): : 400 - 403
  • [8] Initial Stability of Press-Fit Acetabular Components Under Rotational Forces
    Fehring, Keith A.
    Owen, John R.
    Kurdin, Anton A.
    Wayne, Jennifer S.
    Jiranek, William A.
    JOURNAL OF ARTHROPLASTY, 2014, 29 (05): : 1038 - 1042
  • [9] The influence of bone damage on press-fit mechanics
    Bishop, Nicholas E.
    Hoehn, Jan-Christian
    Rothstock, Stephan
    Damm, Niklas B.
    Morlock, Michael M.
    JOURNAL OF BIOMECHANICS, 2014, 47 (06) : 1472 - 1478
  • [10] STABILITY OF PRESS-FIT ACETABULAR CUPS
    KUMMER, FJ
    STUCHIN, SA
    ADLER, A
    BULLETIN OF THE HOSPITAL FOR JOINT DISEASES ORTHOPAEDIC INSTITUTE, 1991, 51 (01): : 116 - 116