Comparison of two different EPID-based solutions performing pretreatment quality assurance: 2D portal dosimetry versus 3D forward projection method

被引:21
|
作者
Bresciani, Sara [1 ]
Poli, Matteo [1 ]
Miranti, Anna [1 ]
Maggio, Angelo [1 ]
Di Dia, Amalia [1 ]
Bracco, Christian [1 ]
Gabriele, Pietro [2 ]
Stasi, Michele [1 ]
机构
[1] IRCCS, Candiolo Canc Inst FPO, Med Phys Div, Str Prov 142 Km 3-95, I-10060 Candiolo, TO, Italy
[2] IRCCS, Candiolo Canc Inst FPO, Radiotherapy Div, Str Prov 142 Km 3-95, I-10060 Candiolo, TO, Italy
关键词
Portal dosimetry; Gamma metric; Pre-treatment measurements; Delivery errors; RADIATION-THERAPY; TRANSIT DOSIMETRY; VERIFICATION; IMRT; SYSTEM; RADIOTHERAPY; BEAMS; MODEL; VMAT;
D O I
10.1016/j.ejmp.2018.06.005
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The aim of this paper is to characterize two different EPID-based solutions for pre-treatment VMAT quality assurance, the 2D portal dosimetry and the 3D projection technique. Their ability to catch the main critical delivery errors was studied. Methods: Measurements were performed with a linac accelerator equipped with EPID aSi1000, Portal Dose Image Prediction (PDIP), and PerFRACTION softwares. Their performances were studied simulating perturbations of a reference plan through systematic variations in dose values and micromultileaf collimator position. The performance of PDIP, based on 2D forward method, was evaluated calculating gamma passing rate (% GP) between no-error and error-simulated measurements. The impact of errors with PerFRACTION, based on 3D projection technique, was analyzed by calculating the difference between reference and perturbed DVH (%Delta D). Subsequently pre-treatment verification with PerFRACTION was done for 27 patients of different pathologies. Results: The sensitivity of PerFRACTION was slightly higher than sensitivity of PDIP, reaching a maximum of 0.9. Specificity was 1 for PerFRACTION and 0.6 for PDIP. The analysis of patients' DVHs indicated that the mean %Delta D was (1.2 +/- 1.9)% for D2%, (0.6 +/- 1.7)% for D95% and (- 0.0 +/- 1.2)% for Dmean of PTV. Regarding OARs, we observed important discrepancies on DVH but that the higher dose variations were in low dose area (< 10 Gy). Conclusions: This study supports the introduction of the new 3D forward projection method for pretreatment QA raising the claim that the visualization of the delivered dose distribution on patient anatomy has major advantages over traditional portal dosimetry QA systems.
引用
收藏
页码:65 / 71
页数:7
相关论文
共 36 条
  • [21] Comparison of interval breast cancers with 2D digital mammography versus 3D digital breast tomosynthesis in a large community-based practice
    Winter, Andrea M.
    Kazmi, Sakina
    Hardy, Anna K.
    Bennett, Debbie L.
    [J]. BREAST JOURNAL, 2020, 26 (10): : 1953 - 1959
  • [22] Two-level Newton iterative method based on nonconforming finite element discretization for 2D/3D stationary MHD equations
    Xu, Jiali
    Feng, Xinlong
    Su, Haiyan
    [J]. Computers and Fluids, 2022, 238
  • [23] A parallel two-grid method based on finite element approximations for the 2D/3D Navier-Stokes equations with damping
    Wassim, Eid
    Zheng, Bo
    Shang, Yueqiang
    [J]. ENGINEERING WITH COMPUTERS, 2024, 40 (01) : 541 - 554
  • [24] Virtual Clinical Trials in 2D and 3D X-ray Breast Imaging and Dosimetry: Comparison of CPU-Based and GPU-Based Monte Carlo Codes
    Mettivier, Giovanni
    Sarno, Antonio
    Lai, Youfang
    Golosio, Bruno
    Fanti, Viviana
    Italiano, Maria Elena
    Jia, Xun
    Russo, Paolo
    [J]. CANCERS, 2022, 14 (04)
  • [25] 2D versus 3D probabilistic homogenization of the metallic fiber-reinforced composites by the perturbation-based stochastic Finite Element Method
    Kaminski, Marcin
    Kazimierczak, Marlena
    [J]. COMPOSITE STRUCTURES, 2014, 108 : 1009 - 1018
  • [26] Inferring 3D Lung Tumor Positions Based On Projected 2D Positions: A Comparison of Different Real-Time Tumor Tracking Methods
    Montanaro, T.
    Nguyen, D.
    Keall, P.
    Booth, J.
    Caillet, V.
    Eade, T.
    Haddad, C.
    Shieh, C.
    [J]. MEDICAL PHYSICS, 2017, 44 (06) : 3245 - 3245
  • [27] A Parallel Subgrid Stabilized Finite Element Method Based on Two-Grid Discretization for Simulation of 2D/3D Steady Incompressible Flows
    Shang, Yueqiang
    Huang, Shumei
    [J]. JOURNAL OF SCIENTIFIC COMPUTING, 2014, 60 (03) : 564 - 583
  • [28] A Parallel Subgrid Stabilized Finite Element Method Based on Two-Grid Discretization for Simulation of 2D/3D Steady Incompressible Flows
    Yueqiang Shang
    Shumei Huang
    [J]. Journal of Scientific Computing, 2014, 60 : 564 - 583
  • [29] Machine-learning-based performance comparison of two-dimensional (2D) and three- dimensional (3D) CT radiomics features for intracerebral
    Chen, Q.
    Fu, C.
    Qiu, X.
    He, J.
    Zhao, T.
    Zhang, Q.
    Hu, X.
    Hu, H.
    [J]. CLINICAL RADIOLOGY, 2024, 79 (01) : e26 - e33
  • [30] 2D building change detection from high resolution satellite imagery: A two-step hierarchical method based on 3D invariant primitives
    Champion, Nicolas
    Boldo, Didier
    Pierrot-Deseilligny, Marc
    Stamon, Georges
    [J]. PATTERN RECOGNITION LETTERS, 2010, 31 (10) : 1138 - 1147