Development of miniaturized mass flow meter for an axial flow blood pump

被引:1
|
作者
Kosaka, Ryo [1 ]
Maruyama, Osamu [1 ]
Nishida, Masahiro [1 ]
Yamane, Takashi [1 ]
机构
[1] AIST, Inst Human Sci & Biomed Engn, Tsukuba, Ibaraki 3058564, Japan
关键词
axial flow blood pump; centrifugal force; flow meter; mass flow; miniaturization; strain gauge;
D O I
10.1111/j.1525-1594.2007.00402.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
To grasp the conditions of patients and implantable artificial hearts, it is essential to monitor the blood flow rate continuously and noninvasively. However, it is difficult to monitor the pump flow rate in an implantable artificial heart, because the conventional flow meter is too large to implant into the human body, and the flow estimation method is influenced by changes in the blood characteristics and the pump performance. In particular, the power consumption has neither linearity nor uniqueness with respect to the pump flow rate in an axial flow blood pump. In this research, we develop a prototype miniaturized mass flow meter that uses centrifugal force F-c for discharged patients with an axial flow blood pump. This flow meter measures the F-c corresponding to the mass flow rate, and implements compensation for static pressure. Because the strain gauges are attached outside of the curved tube, this mass flow meter has no blood contact point, resulting in a compact design. To evaluate the measurement accuracy and the tracking performance, the mass flow meter was compared with the conventional ultrasonic flow meter in a mock-up circulation study. As a result, the measurement error ranging from 0.5 to 5.0 L/min was less than +/- 10% with respect to the maximum flow rate. The tracking performance of pulsation flow was approximately equivalent to that of the conventional flow meter. These experiments demonstrated that the prototype miniaturized mass flow meter using F-c could accurately measure the mass flow rate continuously and noninvasively.
引用
收藏
页码:412 / 419
页数:9
相关论文
共 50 条
  • [1] Development of a miniaturized mass-flow meter for an axial flow blood pump based on computational analysis
    Ryo Kosaka
    Masahiro Nishida
    Osamu Maruyama
    Takashi Yamane
    [J]. Journal of Artificial Organs, 2011, 14 : 178 - 184
  • [2] Development of Miniaturized Mass-Flow Meter for an Axial Flow Blood Pump using a Curved Cannula
    Kosaka, Ryo
    Nishida, Masahiro
    Maruyama, Osamu
    Yamane, Takashi
    [J]. 2010 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2010, : 2513 - 2516
  • [3] Development of a miniaturized mass-flow meter for an axial flow blood pump based on computational analysis
    Kosaka, Ryo
    Nishida, Masahiro
    Maruyama, Osamu
    Yamane, Takashi
    [J]. JOURNAL OF ARTIFICIAL ORGANS, 2011, 14 (03) : 178 - 184
  • [4] Noninvasive Miniaturized Mass-flow Meter using a Curved Cannula for Implantable Axial Flow Blood Pump
    Kosaka, Ryo
    Nishida, Masahiro
    Maruyama, Osamu
    Yamane, Takashi
    [J]. 2011 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2011, : 1343 - 1346
  • [5] Fluid dynamics aspects of miniaturized axial-flow blood pump
    Kang, Can
    Huang, Qifeng
    Li, Yunxiao
    [J]. BIO-MEDICAL MATERIALS AND ENGINEERING, 2014, 24 (01) : 723 - 729
  • [6] Noninvasive Blood-Flow Meter Using a Curved Cannula with Zero Compensation for an Axial Flow Blood Pump
    Kosaka, Ryo
    Fukuda, Kyohei
    Nishida, Masahiro
    Maruyama, Osamu
    Yamane, Takashi
    [J]. 2013 35TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2013, : 4090 - 4093
  • [7] Development of rear-impeller axial flow blood pump for realization of axial flow blood pump installed at aortic valve position
    Okamoto, Eiji
    Yano, Tetsuya
    Inoue, Yusuke
    Shiraishi, Yasuyuki
    Yambe, Tomoyuki
    Mitamura, Yoshinori
    [J]. ARTIFICIAL ORGANS, 2019, 43 (09) : 828 - 833
  • [8] Controller for an axial flow blood pump
    Konishi, H
    Antaki, JF
    Amin, DV
    Boston, JR
    Kerrigan, JP
    Mandarino, WA
    Litwak, P
    Yamazaki, K
    Macha, M
    Butler, KC
    Borovetz, HS
    Kormos, RL
    [J]. ARTIFICIAL ORGANS, 1996, 20 (06) : 618 - 620
  • [9] MECHANICAL CIRCULATORY SUPPORT USING A MINIATURIZED ARCHIMEDES SCREW (AXIAL-FLOW BLOOD PUMP)
    VONSEGESSER, LK
    BISANG, B
    LESKOSEK, B
    TURINA, M
    [J]. HELVETICA CHIRURGICA ACTA, 1991, 57 (04) : 625 - 629
  • [10] Development of an axial flow blood pump: Characteristics of a magnetic fluid seal
    Sekine K.
    Asakawa M.
    Mitamura Y.
    [J]. Journal of Artificial Organs, 2001, 4 (3) : 245 - 251