Controller for an axial flow blood pump

被引:50
|
作者
Konishi, H
Antaki, JF
Amin, DV
Boston, JR
Kerrigan, JP
Mandarino, WA
Litwak, P
Yamazaki, K
Macha, M
Butler, KC
Borovetz, HS
Kormos, RL
机构
[1] UNIV PITTSBURGH,MED CTR,PRESBYTERIAN UNIV HOSP,DIV CARDIOTHORAC SURG,PITTSBURGH,PA 15213
[2] UNIV PITTSBURGH,SCH MED,ARTIFICIAL HEART & LUNG PROGRAM,PITTSBURGH,PA
[3] UNIV PITTSBURGH,SCH ENGN,ARTIFICIAL HEART & LUNG PROGRAM,PITTSBURGH,PA
[4] NIMBUS INC,RANCHO CORDOVA,CA
关键词
axial flow blood pump; noninvasive pump regulation;
D O I
10.1111/j.1525-1594.1996.tb04491.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A rotary blood pump inherently provides only one noninvasive ''observable'' parameter (motor current) and allows for only one ''controllable'' parameter (pump speed). To maintain the systemic circulation properly, the pump speed must be controlled to sustain appropriate outlet flows and perfusion pressure while preventing pulmonary damage caused by extremes in preload. Steady-state data were collected at repeated intervals during chronic trials of the Nimbus AxiPump (Nimbus, Inc., Rancho Cordova, California, U.S.A.) in sheep (n = 7) and calves (n = 12). For each data set, the pump speed was increased at increments of 500 rpm until left ventricular and left atrial emptying was observed by left atrial pressure diminishing to zero. The effect of decreasing preload was evaluated perioperatively by inferior vena cava occlusion at a constant pump speed. Fourier analysis established a relationship between changes in the pump preload and the power spectra of the pump current waveform. Based on these results, a control method was devised to avoid ventricular collapse and maintain the preload within a physiologic range. The objective of this controller is the minimization of the second and third harmonic of the periodic current waveform. This method is intended to provide a noninvasive regulation of the pump by eliminating the need for extraneous transducers.
引用
收藏
页码:618 / 620
页数:3
相关论文
共 50 条
  • [1] A Controller Design for an Axial Flow Blood Pump Based on Back-EMF without Delay
    Gao, Bin
    Song, Miao
    Chang, Yu
    [J]. 2009 3RD INTERNATIONAL CONFERENCE ON BIOINFORMATICS AND BIOMEDICAL ENGINEERING, VOLS 1-11, 2009, : 1242 - 1245
  • [2] Modeling and identification of an axial flow blood pump
    Choi, SJ
    Boston, JR
    Thomas, D
    Antaki, JF
    [J]. PROCEEDINGS OF THE 1997 AMERICAN CONTROL CONFERENCE, VOLS 1-6, 1997, : 3714 - 3715
  • [3] An automatic flow controller for a centrifugal blood pump
    Beppu, T
    Seo, K
    Imai, Y
    Takanashi, Y
    Hoshino, S
    Ohta, J
    Aoki, M
    Uchita, S
    Suzuki, S
    Tajima, Y
    Tooyama, N
    Fukui, Y
    [J]. ARTIFICIAL ORGANS, 1997, 21 (07) : 630 - 634
  • [4] Development of rear-impeller axial flow blood pump for realization of axial flow blood pump installed at aortic valve position
    Okamoto, Eiji
    Yano, Tetsuya
    Inoue, Yusuke
    Shiraishi, Yasuyuki
    Yambe, Tomoyuki
    Mitamura, Yoshinori
    [J]. ARTIFICIAL ORGANS, 2019, 43 (09) : 828 - 833
  • [5] Induction of ventricular collapse by an axial flow blood pump
    Amin, DV
    Antaki, JF
    Litwak, P
    Thomas, D
    Wu, ZJJ
    Watach, M
    [J]. ASAIO JOURNAL, 1998, 44 (05) : M685 - M690
  • [6] Experiment and design improvement of an axial flow blood pump
    School of Jet Propulsion, Beijing University of Aeronautics and Astronautics, Beijing 100083, China
    不详
    [J]. Hangkong Dongli Xuebao, 2008, 6 (1061-1066):
  • [7] Flow Visualization in the Outflow Cannula of an Axial Blood Pump
    Liu, Guangmao
    Zhang, Yan
    Chen, Haibo
    Sun, Hansong
    Zhou, Jianye
    Hu, Shengshou
    [J]. BIO-MEDICAL MATERIALS AND ENGINEERING, 2014, 24 (01) : 117 - 122
  • [8] Miniaturization of a Magnetically Levitated Axial Flow Blood Pump
    Cheng, Shanbao
    Olles, Mark W.
    Olsen, Don B.
    Joyce, Lyle D.
    Day, Steven W.
    [J]. ARTIFICIAL ORGANS, 2010, 34 (10) : 807 - 815
  • [9] Induction of ventricular collapse by an axial flow blood pump
    Amin, Devin V.
    Antaki, James F.
    Litwak, Philip
    Thomas, Douglas
    Wu, Zhongjun J.
    Watach, Mary
    [J]. ASAIO Journal, 44 (05):
  • [10] In vitro performance of a centrifugal, a mixed flow, and an axial flow blood pump
    Medical Informatics, Miyazaki Medical College Hospital, Miyazaki, Japan
    不详
    不详
    不详
    [J]. Artif. Organs, 5 (366-370):