COMPARISON OF FIVE PORTABLE PEAK FLOW METERS

被引:10
|
作者
Takara, Glaucia Nency [1 ]
Ruas, Gualberto [1 ]
Pessoa, Bruna Varanda [1 ]
Jamami, Luciana Kawakami [2 ]
Pires Di Lorenzo, Valeria Amorim [1 ]
Jamami, Mauricio [1 ]
机构
[1] Univ Fed Sao Carlos, Phys Therapy Dept, Sao Carlos, SP, Brazil
[2] Ctr Univ Cent Paulista UNICEP, Sao Carlos, SP, Brazil
关键词
Peak expiratory flow rate; Asthma; Monitoring; Airway obstruction; Meters; MINI-WRIGHT; EXPIRATORY FLOW; ACCURACY; PERFORMANCE;
D O I
10.1590/S1807-59322010000500003
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
OBJECTIVE: To compare the measurements of spirometric peak expiratory flow (PEF) from five different PEF meters and to determine if their values are in agreement. Inaccurate equipment may result in incorrect diagnoses of asthma and inappropriate treatments. METHODS: Sixty-eight healthy, sedentary and insufficiently active subjects, aged from 19 to 40 years, performed PEF measurements using Air Zone (R), Assess (R), Galemed (R), Personal Best (R) and Vitalograph (R) peak flow meters. The highest value recorded for each subject for each device was compared to the corresponding spirometric values using Friedman's test with Dunn's post-hoc (p<0.05), Spearman's correlation test and Bland-Altman's agreement test. RESULTS: The median and interquartile ranges for the spirometric values and the Air Zone (R), Assess (R), Galemed (R), Personal Best (R) and Vitalograph (R) meters were 428 (263-688 L/min), 450 (350-800 L/min), 420 (310-720 L/min), 380 (300-735 L/min), 400 (310-685 L/min) and 415 (335-610 L/min), respectively. Significant differences were found when the spirometric values were compared to those recorded by the Air Zone (R) (p<0.001) and Galemed (R) (p<0.01) meters. There was no agreement between the spirometric values and the five PEF meters. CONCLUSIONS: The results suggest that the values recorded from Galemed (R) meters may underestimate the actual value, which could lead to unnecessary interventions, and that Air Zone (R) meters overestimate spirometric values, which could obfuscate the need for intervention. These findings must be taken into account when interpreting both devices' results in younger people. These differences should also be considered when directly comparing values from different types of PEF meters.
引用
收藏
页码:469 / 474
页数:6
相关论文
共 50 条
  • [31] HOME PEAK FLOW METERS
    LEVY, M
    [J]. BRITISH MEDICAL JOURNAL, 1991, 302 (6786): : 1209 - 1210
  • [32] Comparison of peak cough flows measured by pneumotachograph and a portable peak flow meter
    Sancho, J
    Servera, E
    Díaz, J
    Marín, J
    [J]. AMERICAN JOURNAL OF PHYSICAL MEDICINE & REHABILITATION, 2004, 83 (08) : 608 - 612
  • [33] Comparison of bench test results measuring the accuracy of peak flow meters
    Cristiano VanZeller
    Andrew Williams
    Ian Pollock
    [J]. BMC Pulmonary Medicine, 19
  • [34] Comparison of bench test results measuring the accuracy of peak flow meters
    VanZeller, Cristiano
    Williams, Andrew
    Pollock, Ian
    [J]. BMC PULMONARY MEDICINE, 2019, 19 (1)
  • [35] COMPARISON OF MINI-WRIGHT AND STANDARD WRIGHT PEAK FLOW METERS
    BROWN, LA
    SLY, RM
    [J]. ANNALS OF ALLERGY, 1980, 45 (02): : 72 - 74
  • [36] COMPARISON OF THE MINI-WRIGHT AND THE WRIGHT POCKET PEAK FLOW METERS
    BODANI, H
    PEARSON, J
    LEVY, M
    [J]. THORAX, 1992, 47 (03) : P230 - P230
  • [37] The comparison study on accuracy of the novel digital peak expiratory flow device compared to conventional peak flow meters
    Chaiparnich, S.
    Manasilp, N.
    Poachanukoon, O.
    Saiphoklang, N.
    Tantibundhit, C.
    Vatcharavongvan, P.
    Kunumpol, P.
    Phienphanich, P.
    Pugongchai, A.
    Leelasittikul, K.
    Prompoom, N.
    Puangarom, S.
    [J]. ALLERGY, 2023, 78 : 54 - 54
  • [38] MINIATURE PEAK-FLOW METERS
    WRIGHT, BM
    [J]. BRITISH MEDICAL JOURNAL, 1979, 1 (6161): : 489 - 489
  • [39] THE ACCURACY OF PEAK EXPIRATORY FLOW METERS
    VANAS, A
    [J]. CHEST, 1982, 82 (03) : 263 - 263
  • [40] MINIATURE PEAK-FLOW METERS
    FREEDMAN, BJ
    [J]. BRITISH MEDICAL JOURNAL, 1979, 1 (6158): : 268 - 268