COMPARISON OF FIVE PORTABLE PEAK FLOW METERS

被引:10
|
作者
Takara, Glaucia Nency [1 ]
Ruas, Gualberto [1 ]
Pessoa, Bruna Varanda [1 ]
Jamami, Luciana Kawakami [2 ]
Pires Di Lorenzo, Valeria Amorim [1 ]
Jamami, Mauricio [1 ]
机构
[1] Univ Fed Sao Carlos, Phys Therapy Dept, Sao Carlos, SP, Brazil
[2] Ctr Univ Cent Paulista UNICEP, Sao Carlos, SP, Brazil
关键词
Peak expiratory flow rate; Asthma; Monitoring; Airway obstruction; Meters; MINI-WRIGHT; EXPIRATORY FLOW; ACCURACY; PERFORMANCE;
D O I
10.1590/S1807-59322010000500003
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
OBJECTIVE: To compare the measurements of spirometric peak expiratory flow (PEF) from five different PEF meters and to determine if their values are in agreement. Inaccurate equipment may result in incorrect diagnoses of asthma and inappropriate treatments. METHODS: Sixty-eight healthy, sedentary and insufficiently active subjects, aged from 19 to 40 years, performed PEF measurements using Air Zone (R), Assess (R), Galemed (R), Personal Best (R) and Vitalograph (R) peak flow meters. The highest value recorded for each subject for each device was compared to the corresponding spirometric values using Friedman's test with Dunn's post-hoc (p<0.05), Spearman's correlation test and Bland-Altman's agreement test. RESULTS: The median and interquartile ranges for the spirometric values and the Air Zone (R), Assess (R), Galemed (R), Personal Best (R) and Vitalograph (R) meters were 428 (263-688 L/min), 450 (350-800 L/min), 420 (310-720 L/min), 380 (300-735 L/min), 400 (310-685 L/min) and 415 (335-610 L/min), respectively. Significant differences were found when the spirometric values were compared to those recorded by the Air Zone (R) (p<0.001) and Galemed (R) (p<0.01) meters. There was no agreement between the spirometric values and the five PEF meters. CONCLUSIONS: The results suggest that the values recorded from Galemed (R) meters may underestimate the actual value, which could lead to unnecessary interventions, and that Air Zone (R) meters overestimate spirometric values, which could obfuscate the need for intervention. These findings must be taken into account when interpreting both devices' results in younger people. These differences should also be considered when directly comparing values from different types of PEF meters.
引用
收藏
页码:469 / 474
页数:6
相关论文
共 50 条
  • [1] Portable peak flow meters
    McNaughton, JP
    [J]. EUROPEAN RESPIRATORY JOURNAL, 1997, 10 : S26 - S28
  • [2] THE ACCURACY OF PORTABLE PEAK FLOW METERS
    MILLER, MR
    DICKINSON, SA
    HITCHINGS, DJ
    [J]. THORAX, 1992, 47 (11) : 904 - 909
  • [3] Peak expiratory flow associated with change in positioning of the instrument: Comparison of five peak flow meters
    Nolan, S
    Tolley, E
    Leeper, K
    Smith, VS
    Self, T
    [J]. JOURNAL OF ASTHMA, 1999, 36 (03) : 291 - 294
  • [4] PEAK EXPIRATORY FLOW MEASUREMENTS IN YOUNG-CHILDREN - COMPARISON OF 4 PORTABLE PEAK FLOW METERS
    TRUONG, M
    INIGUEZ, JL
    CHOUHOU, D
    DESSANGE, JF
    GENDREL, D
    CHAUSSAIN, M
    [J]. ARCHIVES DE PEDIATRIE, 1995, 2 (04): : 324 - 327
  • [5] AN ASSESSMENT OF 3 PORTABLE PEAK FLOW METERS
    EICHENHORN, MS
    BEAUCHAMP, RK
    HARPER, PA
    WARD, JC
    [J]. CHEST, 1982, 82 (03) : 306 - 309
  • [6] COMPARISON OF PEAK FLOW METERS
    LEVY, M
    BODANI, H
    [J]. BRITISH JOURNAL OF GENERAL PRACTICE, 1993, 43 (366): : 36 - 37
  • [7] Interinstrument variability of three portable peak flow meters
    Schmekel, B
    [J]. CHEST, 1996, 109 (06) : 1668 - 1668
  • [8] PORTABLE PEAK-FLOW METERS - INTRASUBJECT COMPARISONS
    HARM, DL
    KOTSES, H
    CREER, TL
    [J]. JOURNAL OF ASTHMA, 1984, 21 (01) : 9 - 13
  • [9] Inaccuracy of portable peak flow meters: correction is not needed
    Brand, PLP
    Waalkens, HJ
    Duiverman, EJ
    vanEssenZandvliet, EEM
    [J]. ACTA PAEDIATRICA, 1997, 86 (08) : 888 - 889
  • [10] The peak flow working group: Test of portable peak flow meters by explosive decompression
    Pedersen, OF
    Miller, MR
    [J]. EUROPEAN RESPIRATORY JOURNAL, 1997, 10 : S23 - S25