Design and Characterization of a Dedicated Cone-Beam CT Scanner for Detection of Acute Intracranial Hemorrhage

被引:1
|
作者
Xu, J. [1 ]
Sisniega, A. [1 ]
Zbijewski, W. [1 ]
Dang, H. [1 ]
Stayman, J. W. [1 ]
Wang, X. [2 ]
Foos, D. H. [2 ]
Aygun, N. [3 ]
Koliatsos, V. E. [4 ]
Siewerdsen, J. H. [1 ,3 ]
机构
[1] Johns Hopkins Univ, Dept Biomed Engn, Baltimore, MD 21218 USA
[2] Carestream Hlth Inc, Rochester, NY USA
[3] Johns Hopkins Univ, Dept Radiol, Baltimore, MD USA
[4] Johns Hopkins Univ, Dept Neurol, Baltimore, MD 21218 USA
关键词
Cone-Beam CT; intracranial hemorrhage; task based detectability; system design; OPTIMIZATION; PERFORMANCE;
D O I
10.1117/12.2216544
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
Purpose: Prompt and reliable detection of intracranial hemorrhage (ICH) has substantial clinical impact in diagnosis and treatment of stroke and traumatic brain injury. This paper describes the design, development, and preliminary performance characterization of a dedicated cone-beam CT (CBCT) head scanner prototype for imaging of acute ICH. Methods: A task-based image quality model was used to analyze the detectability index as a function of system configuration, and hardware design was guided by the results of this model-based optimization A robust artifact correction pipeline was developed using GPU-accelerated Monte Carlo (MC) scatter simulation, beam hardening corrections, detector veiling glare, and lag deconvolution. An iterative penalized weighted least-squares (PWLS) reconstruction framework with weights adjusted for artifact-corrected projections was developed. Various bowtie filters were investigated for potential dose and image quality benefits, with a MC-based tool providing estimates of spatial dose distribution. Results: The initial prototype will feature a source-detector distance of 1000 mm and source-axis distance of 550 mm, a 43x43 cm(2) flat panel detector, and a 15 rotating anode x-ray source with 15 kW power and 0.6 focal spot size. Artifact correction reduced image nonuniformity by similar to 250 HU, and PWLS reconstruction with modified weights improved the contrast to noise ratio by 20%. Inclusion of a bowtie filter can potentially reduce dose by 50% and improve CNR by 25%. Conclusions: A dedicated CBCT system capable of imaging millimeter-scale acute ICH was designed. Preliminary findings support feasibility of point-of-care applications in 1131 and stroke imaging, with clinical studies beginning on a prototype.
引用
收藏
页数:7
相关论文
共 50 条
  • [1] Monte Carlo Dosimetry for a Dedicated Cone-Beam CT Head Scanner
    Sisniega, A.
    Zbijewski, W.
    Xu, J.
    Dang, H.
    Stayman, J. W.
    Aygun, N.
    Koliatsos, V. E.
    Wang, X.
    Foos, D. H.
    Siewerdsen, J. H.
    MEDICAL PHYSICS, 2015, 42 (06) : 3681 - 3682
  • [2] Development and Clinical Translation of a Cone-Beam CT Scanner for High-Quality Imaging of Intracranial Hemorrhage
    Sisniega, A.
    Xu, J.
    Dang, H.
    Zbijewski, W.
    Stayman, J. W.
    Mow, M.
    Koliatsos, V. E.
    Aygun, N.
    Wang, X.
    Foos, D. H.
    Siewerdsen, J. H.
    MEDICAL IMAGING 2017: PHYSICS OF MEDICAL IMAGING, 2017, 10132
  • [3] Technical assessment of a prototype cone-beam CT system for imaging of acute intracranial hemorrhage
    Xu, Jennifer
    Sisniega, Alejandro
    Zbijewski, Wojciech
    Dang, Hao
    Stayman, J. Webster
    Mow, Michael
    Wang, Xiaohui
    Foos, David H.
    Koliatsos, Vassillis E.
    Aygun, Nafi
    Siewerdsen, Jeffrey H.
    MEDICAL PHYSICS, 2016, 43 (10) : 5745 - 5757
  • [4] Peripheral Quantitative CT (pQCT) Using a Dedicated Extremity Cone-Beam CT Scanner
    Muhit, A. A.
    Arora, S.
    Ogawa, M.
    Ding, Y.
    Zbijewski, W.
    Stayman, J. W.
    Thawait, G.
    Packard, N.
    Senn, R.
    Yang, D.
    Yorkston, J.
    Bingham, C. O., III
    Means, K.
    Carrino, J. A.
    Siewerdsen, J. H.
    MEDICAL IMAGING 2013: BIOMEDICAL APPLICATIONS IN MOLECULAR, STRUCTURAL, AND FUNCTIONAL IMAGING, 2013, 8672
  • [5] Dedicated scanner for laboratory investigations on cone-beam CT/SPECT imaging of the breast
    Mettivier, Giovanni
    Russo, Paolo
    Cesarelli, Mario
    Ospizio, Roberto
    Passeggio, Giuseppe
    Roscilli, Lorenzo
    Pontoriere, Giuseppe
    Rocco, Raffaele
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2011, 629 (01): : 350 - 356
  • [6] Cone-Beam Artifacts in Dedicated Breast CT
    Vedantham, S.
    Shi, L.
    Noo, F.
    Glick, S.
    Karellas, A.
    MEDICAL PHYSICS, 2011, 38 (06)
  • [7] Diagnostic Accuracy of Cone-Beam CT for Acute Intracranial Hemorrhage: A Systematic Review and Meta-Analysis
    Ferrone, Nicholas G.
    Sanmartin, Maria X.
    O'Hara, Joseph
    Jimenez, Jean
    Ferrone, Sophia R.
    Lodato, Zachary
    Lacher, Gregory
    Bandi, Sanjana
    Convey, Alicia
    Bastani, Mehrad
    Lee, Un Jung
    Vialet, Jaclyn Morales
    White, Timothy
    Wang, Jason J.
    Katz, Jeffrey M.
    Sanelli, Pina C.
    JOURNAL OF THE AMERICAN COLLEGE OF RADIOLOGY, 2024, 21 (12) : 1841 - 1850
  • [8] Design and Initial Performance Characterization of a Dedicated Cone-Beam CT System for Musculoskeletal Extremities Imaging
    Zbijewski, W.
    De Jean, P.
    Prakash, P.
    Stayman, J.
    Packard, N.
    Senn, R.
    Yang, D.
    Yorkston, J.
    Machado, A.
    Carrino, J.
    Siewerdsen, J.
    MEDICAL PHYSICS, 2010, 37 (06)
  • [9] Regularization Design for High-Quality Cone-Beam CT of Intracranial Hemorrhage using Statistical Reconstruction
    Dang, H.
    Stayman, J. W.
    Xu, J.
    Sisniega, A.
    Zbijewski, W.
    Wang, X.
    Foos, D. H.
    Aygun, N.
    Koliatsos, V. E.
    Siewerdsen, J. H.
    MEDICAL IMAGING 2016: PHYSICS OF MEDICAL IMAGING, 2016, 9783
  • [10] Task-Based Modeling and Optimization of a Dedicated Cone-Beam CT Scanner for Musculoskeletal Imaging
    Prakash, P.
    Gang, G.
    Zbijewski, W.
    Yorkston, J.
    Machado, A.
    Carrino, J.
    Siewerdsen, J.
    MEDICAL PHYSICS, 2011, 38 (06)