New technique for mediastinal temperature measurement in hyperthermic cancer treatment: Balloon catheter in the azygos vein

被引:3
|
作者
Murata, T [1 ]
Nagata, K [1 ]
Akagi, K [1 ]
Nasu, R [1 ]
Imamura, M [1 ]
Kimura, H [1 ]
Tanaka, Y [1 ]
机构
[1] Kansai Med Univ, Dept Radiol, Osaka 570, Japan
关键词
hyperthermia; temperature measurement; angiographics; mediastinal tumours; lung cancer; oesophageal cancer;
D O I
10.1177/030006059802600106
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
The temperature in the mediastinum during hyperthermia is difficult to determine accurately. We measured the temperature in the azygos vein, using a new technique, and compared the measurements with temperatures in the oesophagus. Eight patients with mediastinal tumours resulting from lung cancer or oesophageal cancer were given hyperthermo-radiotherapy. The temperatures in the azygos vein and in the oesophagus were measured before and during blockage of the blood now of the azygos vein using an angiographic balloon catheter. None of the patients had complications as a result of these procedures, and hyperthermia by capacitative heating was safely performed. The temperature in the azygos vein increased by a mean of 1.7 degrees C (0.2-2.8 degrees C) after blockage of the blood flow. The temperature in the oesophagus was 0.83 +/- 1.09 degrees C (mean +/- SD) higher than that in the azygos vein. Measurement of the temperature in the azygos vein gives a more accurate estimate of mediastinal temperature than does oesophageal temperature but it is an invasive procedure.
引用
收藏
页码:50 / 56
页数:7
相关论文
共 50 条
  • [31] Transmural pulmonary vein lesions created by using a new ultrasound balloon catheter - an experimental study
    Maia, H. C.
    Seixas, T. N.
    Leite, L.
    Neto, J. Sobral
    Barreto, J. R.
    Septimio, C.
    Oliveira, E. M.
    Peres, A. K.
    EUROPEAN HEART JOURNAL, 2005, 26 : 597 - 598
  • [32] Usefulness of a new radiofrequency thermal balloon catheter for pulmonary vein isolation to treat atrial fibrillation
    Satake, S
    EUROPEAN HEART JOURNAL, 2003, 24 : 166 - 166
  • [33] A new balloon catheter EMR method for mucosal detachment in cancer resection
    Wada, M
    Takimoto, Y
    Orino, A
    GASTROINTESTINAL ENDOSCOPY, 2005, 61 (05) : AB243 - AB243
  • [34] New technique for superior guiding catheter support during advancement of a balloon in coronary angioplasty: The anchor technique
    Fujita, S
    Tamai, H
    Kyo, E
    Kosuga, K
    Hata, T
    Okada, M
    Nakamura, T
    Tsuji, T
    Takeda, S
    bin Hu, F
    Masunaga, N
    Motohara, S
    Uehata, H
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2003, 59 (04) : 482 - 488
  • [36] RATIONAL BASIS OF A NEW TECHNIQUE OF RADICAL AZYGOS PORTAL DISCONNECTION FOR TREATMENT OF PORTAL HYPERTENSION
    TORRES, UL
    DEGNI, M
    JOURNAL OF CARDIOVASCULAR SURGERY, 1966, : 82 - &
  • [37] A NEW TECHNIQUE IN REPAIR OF FALSE FEMORAL ANEURYSM WITH A PERCUTANEOUS BALLOON CATHETER FOR INFLOW CONTROL
    DIETHRICH, EB
    HANAFY, HM
    JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 1991, 102 (04): : 645 - 648
  • [38] Assessment of a new transurethral balloon dilation catheter in the treatment of urethral stricture disease
    MacDiarmid, SA
    Harrigan, CT
    Cottone, JL
    McIntyre, WJ
    Johnson, DE
    UROLOGY, 2000, 55 (03) : 408 - 413
  • [39] Efficacy and Safety of a New Type of Intraairway Hemostatic Balloon Catheter in the Treatment of Hemoptysis
    Dong, L.
    Lin, Y.
    Chen, C.
    AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2024, 209
  • [40] TEMPERATURE-GUIDED RADIOFREQUENCY ABLATION USING A NEW ELECTRODE CATHETER WITH A GOLD BALLOON
    SATAKE, S
    OKISHIGE, K
    AZEGAMI, K
    OHIRA, H
    YAMASHITA, K
    CIRCULATION, 1995, 92 (08) : 2926 - 2926