Reduction in spread of excitation from current focusing at multiple cochlear locations in cochlear implant users

被引:36
|
作者
Padilla, Monica [1 ,2 ,3 ]
Landsberger, David M. [1 ,2 ]
机构
[1] House Res Inst, Dept Commun & Neurosci, 2100 West 3rd St, Los Angeles, CA 90057 USA
[2] NYU, Sch Med, Dept Otolaryngol, 550 1st Ave STE NBV 5E5, New York, NY 10016 USA
[3] Univ So Calif, Keck Sch Med, Dept Otolaryngol, 806 W Adams Blvd, Los Angeles, CA 90007 USA
关键词
Cochlear; Implants; Tripolar; Spread; Scaling; VIRTUAL CHANNEL DISCRIMINATION; ELECTRODE CONFIGURATION; TRIPOLAR STIMULATION; PLACE SPECIFICITY; MONOPOLAR; MASKING; PATTERNS; NOISE;
D O I
10.1016/j.heares.2016.01.002
中图分类号
R36 [病理学]; R76 [耳鼻咽喉科学];
学科分类号
100104 ; 100213 ;
摘要
Channel interaction from a broad spread of excitation is likely to be a limiting factor in performance by cochlear implant users. Although partial tripolar stimulation has been shown to reduce spread of excitation, the magnitude of the reduction is highly variable across subjects. Because the reduction in spread of excitation is typically only measured at one electrode for a given subject, the degree of variability across cochlear locations is unknown. The first goal of the present study was to determine if the reduction in spread of excitation observed from partial tripolar current focusing systematically varies across the cochlea. The second goal was to measure the variability in reduction of spread of excitation relative to monopolar stimulation across the cochlea. The third goal was to expand upon previous results that suggest that scaling of verbal descriptors can be used to predict the reduction in spread of excitation, by increasing the limited number of sites previously evaluated and verify the relationships remain with the larger dataset. The spread of excitation for monopolar and partial tripolar stimulation was measured at 5 cochlear locations using a psychophysical forward masking task. Results of the present study suggest that although partial tripolar stimulation typically reduces spread of excitation, the degree of reduction in spread of excitation was found to be highly variable and no effect of cochlear location was found. Additionally, subjective scaling of certain verbal descriptors (Clean/Dirty, Pure/Noisy) correlated with the reduction in spread of excitation suggesting sound quality scaling might be used as a quick clinical estimate of channels providing a reduction in spread of excitation. This quick scaling technique might help clinicians determine which patients would be most likely to benefit from a focused strategy. (C) 2016 Elsevier B.V. All rights reserved.
引用
收藏
页码:98 / 107
页数:10
相关论文
共 50 条
  • [1] Reducing current spread using current focusing in cochlear implant users
    Landsberger, David M.
    Padilla, Monica
    Srinivasan, Arthi G.
    HEARING RESEARCH, 2012, 284 (1-2) : 16 - 24
  • [2] Is the spread of excitation width correlated to the speech recognition in cochlear implant users?
    da Silva, Juliana Coutinho
    Schmidt Goffi-Gomez, Maria Valeria
    Magalhaes, Ana Tereza
    Tsuji, Robinson Koji
    Bento, Ricardo Ferreira
    EUROPEAN ARCHIVES OF OTO-RHINO-LARYNGOLOGY, 2021, 278 (06) : 1815 - 1820
  • [3] Is the spread of excitation width correlated to the speech recognition in cochlear implant users?
    Juliana Coutinho da Silva
    Maria Valéria Schmidt Goffi-Gomez
    Ana Tereza Magalhães
    Robinson Koji Tsuji
    Ricardo Ferreira Bento
    European Archives of Oto-Rhino-Laryngology, 2021, 278 : 1815 - 1820
  • [4] Improving speech perception in noise with current focusing in cochlear implant users
    Srinivasan, Arthi G.
    Padilla, Monica
    Shannon, Robert V.
    Landsberger, David M.
    HEARING RESEARCH, 2013, 299 : 29 - 36
  • [5] Influence of the spread of electric field on neural excitation in cochlear implant users: Transimpedance and spread of excitation measurements
    Kopsch, Anna C.
    Rahne, Torsten
    Plontke, Stefan K.
    Wagner, Luise
    HEARING RESEARCH, 2022, 424
  • [6] Current focusing sharpens local peaks of excitation in cochlear implant stimulation
    Srinivasan, Arthi G.
    Landsberger, David M.
    Shannon, Robert V.
    HEARING RESEARCH, 2010, 270 (1-2) : 89 - 100
  • [7] Intra-Cochlear Current Spread Correlates with Speech Perception in Experienced Adult Cochlear Implant Users
    Joly, Charles-Alexandre
    Reynard, Pierre
    Hermann, Ruben
    Seldran, Fabien
    Gallego, Stephane
    Idriss, Samar
    Thai-Van, Hung
    JOURNAL OF CLINICAL MEDICINE, 2021, 10 (24)
  • [8] Precompensating for spread of excitation in a cochlear implant coding strategy
    Bolner, Federico
    Magits, Sara
    van Dijk, Bas
    Wouters, Jan
    HEARING RESEARCH, 2020, 395
  • [9] Investigating the Effect of Blurring and Focusing Current in Cochlear Implant Users with the Panoramic ECAP Method
    Garcia, Charlotte
    Morse-Fortier, Charlotte
    Guerit, Francois
    Hislop, Scott
    Goehring, Tobias
    Carlyon, Robert P.
    Arenberg, Julie G.
    JARO-JOURNAL OF THE ASSOCIATION FOR RESEARCH IN OTOLARYNGOLOGY, 2024,
  • [10] Electrical Cochlear Response Consistency from different Cochlear Implant Users
    Comejo-Cruz, J. M.
    Granados-Trejo, P.
    Castaneda-Villa, N.
    2021 43RD ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE & BIOLOGY SOCIETY (EMBC), 2021, : 6594 - 6597