Evaluation of the accuracy of deformable image registration on MRI with a physical phantom

被引:13
|
作者
Wu, Richard Y. [1 ]
Liu, Amy Y. [1 ]
Yang, Jinzhong [1 ]
Williamson, Tyler D. [1 ]
Wisdom, Paul G. [1 ]
Bronk, Lawrence [1 ]
Gao, Song [1 ]
Grosshan, David R. [2 ]
Fuller, David C. [2 ]
Gunn, Gary B. [2 ]
Ronald Zhu, X. [1 ]
Frank, Steven J. [2 ]
机构
[1] Univ Texas MD Anderson Canc Ctr, Dept Radiat Phys, Houston, TX 77030 USA
[2] Univ Texas MD Anderson Canc Ctr, Dept Radiat Oncol, Houston, TX 77030 USA
来源
基金
美国国家卫生研究院;
关键词
deformable image registration; deformable phantom; MR multimodality image registration; RADIOTHERAPY; VALIDATION; ALGORITHMS;
D O I
10.1002/acm2.12789
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background and purpose Magnetic resonance imaging (MRI) has gained popularity in radiation therapy simulation because it provides superior soft tissue contrast, which facilitates more accurate target delineation compared with computed tomography (CT) and does not expose the patient to ionizing radiation. However, image registration errors in commercial software have not been widely reported. Here we evaluated the accuracy of deformable image registration (DIR) by using a physical phantom for MRI. Methods and materials We used the "Wuphantom" for end-to-end testing of DIR accuracy for MRI. This acrylic phantom is filled with water and includes several fillable inserts to simulate various tissue shapes and properties. Deformations and changes in anatomic locations are simulated by changing the rotations of the phantom and inserts. We used Varian Velocity DIR software (v4.0) and CT (head and neck protocol) and MR (T1- and T2-weighted head protocol) images to test DIR accuracy between image modalities (MRI vs CT) and within the same image modality (MRI vs MRI) in 11 rotation deformation scenarios. Large inserts filled with Mobil DTE oil were used to simulate fatty tissue, and small inserts filled with agarose gel were used to simulate tissues slightly denser than water (e.g., prostate). Contours of all inserts were generated before DIR to provide a baseline for contour size and shape. DIR was done with the MR Correctable Deformable DIR method, and all deformed contours were compared with the original contours. The Dice similarity coefficient (DSC) and mean distance to agreement (MDA) were used to quantitatively validate DIR accuracy. We also used large and small regions of interest (ROIs) during between-modality DIR tests to simulate validation of DIR accuracy for organs at risk (OARs) and propagation of individual clinical target volume (CTV) contours. Results No significant differences in DIR accuracy were found for T1:T1 and T2:T2 comparisons (P > 0.05). DIR was less accurate for between-modality comparisons than for same-modality comparisons, and was less accurate for T1 vs CT than for T2 vs CT (P < 0.001). For between-modality comparisons, use of a small ROI improved DIR accuracy for both T1 and T2 images. Conclusion The simple design of the Wuphantom allows seamless testing of DIR; here we validated the accuracy of MRI DIR in end-to-end testing. T2 images had superior DIR accuracy compared with T1 images. Use of small ROIs improves DIR accuracy for target contour propagation.
引用
收藏
页码:166 / 173
页数:8
相关论文
共 50 条
  • [21] The evaluation of a hybrid biomechanical deformable registration method on a multistage physical phantom with reproducible deformation
    Qin, An
    Ionascu, Dan
    Liang, Jian
    Han, Xiao
    O'Connell, Nicolette
    Yan, Di
    [J]. RADIATION ONCOLOGY, 2018, 13
  • [22] The evaluation of a hybrid biomechanical deformable registration method on a multistage physical phantom with reproducible deformation
    An Qin
    Dan Ionascu
    Jian Liang
    Xiao Han
    Nicolette O’Connell
    Di Yan
    [J]. Radiation Oncology, 13
  • [23] Development of a dynamic deformable thorax phantom for the quality management of deformable image registration
    Sugawara, Yasuharu
    Kadoya, Noriyuki
    Kotabe, Kazuki
    Nakajima, Yujiro
    Ikeda, Ryutaro
    Tanabe, Shunpei
    Ohashi, Haruna
    Jingu, Keiichi
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2020, 77 : 100 - 107
  • [24] Automated Evaluation of Deformable Image Registration Using Physical Tissue Characteristics
    Watkins, W.
    Bzdusek, K.
    Siebers, J. V.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2014, 90 : S847 - S847
  • [25] Image registration accuracy of GammaPlan: a phantom study
    Watanabe, Yoichi
    Han, Eunyoung
    [J]. JOURNAL OF NEUROSURGERY, 2008, 109 : 21 - 24
  • [26] Development of a physical geometric phantom for deformable image registration credentialing of radiotherapy centers for a clinical trial
    Kadoya, Noriyuki
    Sakulsingharoj, Siwaporn
    Kron, Tomas
    Yao, Adam
    Hardcastle, Nicholas
    Bergman, Alanah
    Okamoto, Hiroyuki
    Mukumoto, Nobutaka
    Nakajima, Yujiro
    Jingu, Keiichi
    Nakamura, Mitsuhiro
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2021, 22 (07): : 255 - 265
  • [27] Dosimetric impact of deformable image registration using radiophotoluminescent glass dosimeters with a physical geometric phantom
    Sakulsingharoj, Siwaporn
    Kadoya, Noriyuki
    Tanaka, Shohei
    Sato, Kiyokazu
    Nakamura, Mitsuhiro
    Jingu, Keiichi
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2023, 24 (04):
  • [28] Validation of accuracy deformable image registration contour propagation using a benchmark virtual HN phantom dataset
    Boyd, Robert
    Basavatia, Amar
    Tome, Wolfgang A.
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2021, 22 (05): : 58 - 68
  • [29] Assessing Deformable Image Registration Accuracy in Cervical Brachytherapy Using A 3D Model Phantom
    Williams, Vonetta
    Geneser, Sarah
    Brunnquell, Christina
    Wootton, Landon
    Young, Lori
    Meyer, Juergen
    Dyer, Brandon
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2020, 108 (02): : E17 - E17
  • [30] The Potential for Patient-Specific, Automated Evaluation of Deformable Image Registration Accuracy
    Kirby, N.
    Morin, O.
    Nie, K.
    Chen, J.
    Pouliot, J.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2014, 90 : S878 - S878