Comparison of lactate, bilirubin and hemoglobin F concentrations obtained by the ABL 700 series blood gas analyzers with laboratory methods

被引:19
|
作者
Suen, WW [1 ]
Ridley, B [1 ]
Blakney, G [1 ]
Higgins, TN [1 ]
机构
[1] Dynacare Kasper Med Labs, Dept Chem, Edmonton, AB T5J 5E2, Canada
关键词
lactate; hemglobin F; bilirubin; blood gas analyzer; comparisons;
D O I
10.1016/S0009-9120(02)00449-6
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Objective: To compare lactate, bilirubin and Hemoglobin F concentrations obtained on ABL 700 series blood gas analyzers with those from laboratory methods. Design and Method: Pooled neonatal plasma, cord blood and adult plasma samples were used for comparison of bilirubin, hemoglobin F and lactate concentrations respectively. Results: Results obtained on the ABL 700 series compared favorably (Deming regression slopes 0.97-1.13) with those from laboratory methods. For lactate ABL (y) = 1.13 Vitros (x) -0.43 with a CI (slope) of 1.10 to 1.16, CI (int) of -0.61 to -0.28. For hemoglobin F ABL(y) = 1.11 Variant (x) -8.0 with a CI (slope) of 0.88 to 1.33, CI (int) of -25.3 to 9.3. The three bilirubin comparisons are as follows: 1) Unistat (y) = 1.10 Vitros (x) -16.12 with CI (slope) of 1.06 to 1.14 and CI (int) of -25.3 to -6.9. 2), ABL (y) = 0.97 Vitros(x) -10.16 with CI (slope) of 0.94 to 1.00 and CI (int) of -17.6 to 2.73. Unistat (y) = 1.14 (x)-4.58 with CI (slope) of 1.09 to 1.18 and Cl (int) of -13.6 to 4.5 Conclusion: The ABL 700 series gave comparable results for lactate, bilirubin and hemoglobin F with laboratory methods and may be used in patient care. (C) 2003 The Canadian Society of Clinical Chemists. All rights reserved.
引用
下载
收藏
页码:103 / 107
页数:5
相关论文
共 5 条
  • [1] Hemoglobin in samples with leukocytosis can be measured on ABL 700 series blood gas analyzers
    Scharnhorst, V
    van de Laar, PJ
    Vader, HL
    CLINICAL CHEMISTRY, 2003, 49 (12) : 2107 - 2108
  • [2] Comparison of whole blood total bilirubin determination in neonates on the Siemens Rapidlab and radiometer ABL blood gas analyzers
    Dietzen, D. J.
    Wilhite, T. R.
    CLINICAL CHEMISTRY, 2008, 54 (06) : A235 - A235
  • [3] Comparison of point-of-care and central laboratory analyzers for blood gas and lactate measurements
    Indrasari, Nuri Dyah
    Wonohutomo, Jessica Purwanti
    Sukartini, Ninik
    JOURNAL OF CLINICAL LABORATORY ANALYSIS, 2019, 33 (05)
  • [4] Blood-gas vs. Central-Laboratory analyzers: interchangeability and reference intervals for sodium, potassium, glucose, lactate and hemoglobin
    Marija, Kocijancic
    Bernhard, Kraus Frank
    Beatrice, Ludwig-Kraus
    HELIYON, 2021, 7 (11)
  • [5] Comparison of blood gas results obtained on Abbott i-Stat® and on Radiometer ABL 800 Flex® analyzers Impact for the clinical decision
    Jouffroy, Romain
    Laney, Maude
    Leguillier, Teddy
    Nivet-Antoine, Valerie
    Beaudeux, Jean-Louis
    ANNALES DE BIOLOGIE CLINIQUE, 2022, 80 (06) : 521 - 525