High Resolution 125I Pinhole SPECT Imaging of the Mouse Thyroid With the MediSPECT Small Animal CdTe Scanner

被引:6
|
作者
Mettivier, Giovanni [1 ,2 ]
Montesi, Maria Cristina [1 ,2 ]
Curion, Assunta Simona [1 ,2 ]
Lauria, Adele [1 ,2 ]
Marotta, Marcello [3 ]
Russo, Paolo [1 ,2 ]
机构
[1] Univ Naples Federico II, Dipartimento Sci Fis, I-80126 Naples, Italy
[2] INFN Sezione Napoli, I-80126 Naples, Italy
[3] Univ Naples Federico II, Dipartimento Med Clin & Sperimentale, I-80131 Naples, Italy
关键词
SYSTEM; SINGLE; TOMOGRAPHY; DETECTOR; DESIGN; PERFORMANCE;
D O I
10.1109/TNS.2010.2042815
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
The first in vivo tomographic I-125 imaging of the mouse thyroid carried out with the new MediSPECT small animal SPECT scanner is presented. The scanner is based on a fine pitch CdTe semiconductor pixel detector (14 x 14 mm(2), 256 x 256 square pixel with a 55 mu m side) and equipped with a set of high resolution collimators. The collimation and detection units of the scanner are mounted on a gantry, rotating around a horizontal axis, along which is placed the small animal housing. In an in vivo test, the mouse was injected with a (NaI)-I-125 solution having a total activity of 31.8 MBq. The planar projections for SPECT reconstruction were acquired with a 300 mu m pinhole (magnification 1.47 and field of view of 9.6 x 9.6 mm(2)). The projections were captured in a step-and-shoot fashion and were processed with an Ordered Subsets-Expectation Maximization reconstruction algorithm in order to obtain the SPECT images. Several I-125 imaging tests have been made by using phantoms to assess the detector spatial resolution. The measured spatial resolution with a 300 mu m pinhole is about 0.5 mm in planar imaging and better than 1 mm in tomographic imaging.
引用
收藏
页码:1029 / 1037
页数:9
相关论文
共 50 条
  • [31] Imaging arthritis in mouse-models using high-resolution multi-pinhole spect
    Ostendorf, B
    Wirrwar, A
    Scherer, A
    Joosten, L
    Hoppin, J
    Schramm, N
    Van den Berg, W
    Müller, H
    Schneider, M
    [J]. ANNALS OF THE RHEUMATIC DISEASES, 2005, 64 : 155 - 155
  • [32] High-resolution multi pinhole spect:: A new method for imaging arthritis in mouse-models.
    Ostendorf, B
    Wirrwar, A
    Scherer, A
    Joosten, L
    Hoppin, J
    Schramm, N
    van den Berg, W
    Müller, HW
    Schneider, M
    [J]. ARTHRITIS AND RHEUMATISM, 2005, 52 (09): : S360 - S361
  • [33] Evaluation of the high resolution neurofocus SPECT device for small animal imaging.
    Seibyl, JP
    Stoddart, HA
    Martin, D
    Smith, E
    Wisniewski, G
    Stoddart, HF
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2002, 43 (05) : 229P - 230P
  • [34] State-of-the-art of small animal imaging with high-resolution SPECT
    Nikolaus, S
    Wirrwar, A
    Antke, C
    Kley, K
    Müller, HW
    [J]. NUKLEARMEDIZIN-NUCLEAR MEDICINE, 2005, 44 (06): : 257 - 266
  • [35] Performance evaluation and quantitative accuracy of a multi-pinhole small animal SPECT/CT scanner for theranostic Lu-177 imaging
    Gupta, Arun
    Kim, Kyeong Yun
    Lee, Min Sun
    Hwang, Donghwi
    Lee, Dong Soo
    Lee, Jae Sung
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2018, 59
  • [36] Comparison between pixel array CdTe detector and NaI(Tl) detector for ultra-high resolution small animal SPECT imaging
    Park, S.
    Lee, C.
    Kim, H.
    [J]. MEDICAL IMAGING 2012: PHYSICS OF MEDICAL IMAGING, 2012, 8313
  • [37] TandemPET-A High Resolution, Small Animal, Virtual Pinhole-Based PET Scanner: Initial Design Study
    Raylman, Raymond R.
    Stolin, Alexander V.
    Martone, Peter F.
    Smith, Mark F.
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2016, 63 (01) : 75 - 83
  • [38] Aperture design for ultra-high resolution SPECT systems for small animal imaging
    Fu, Geng
    Tan, Jia-Wei
    Meng, Ling-Iian
    [J]. 2007 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-11, 2007, : 2785 - 2790
  • [39] An accurate and efficient system model of iterative image reconstruction in high-resolution pinhole SPECT for small animal research
    Huang, P. -C.
    Hsu, C. -H.
    Hsiao, I. -T.
    Lin, K. M.
    [J]. JOURNAL OF INSTRUMENTATION, 2009, 4
  • [40] Multi-pinhole collimator design for small-object imaging with SiliSPECT: a high-resolution SPECT
    Shokouhi, S.
    Metzler, S. D.
    Wilson, D. W.
    Peterson, T. E.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (02): : 207 - 225