Initial stability of fully and partially cemented femoral stems

被引:35
|
作者
Claes, L
Fiedler, S
Ohnmacht, M
Duda, GN
机构
[1] Univ Ulm, Inst Unfallchirurg Forsch & Biomech, D-89081 Ulm, Germany
[2] Humboldt Univ, Virchow Klinikum, Abt Unfallchirurg & Wiederherstellungschirurg, D-13353 Berlin, Germany
关键词
total hip replacement; initial stability; prosthesis design; mechanical testing;
D O I
10.1016/S0268-0033(00)00044-9
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Objective. To test the initial stability of a newly designed partially cemented femoral stem in comparison with a fully cemented conventional stem. Design. An in vitro study to determine the interface motion between femoral stem and bone as a response to loading. Background The aim of the new prosthesis design is a proximal load transfer by a defined partial cement fixation in the proximal femur region and a slim prosthesis stem in the distal region. Before a clinical study can be started, the new stem has to show an initial stability comparable to that of fully cemented prostheses. Method Six paired fresh cadaveric femora were used for the testing of the new partially cemented stem (Option 3000, Mathys Orthopaedics, Bettlach, Switzerland) and a fully cemented stein (Weber Shaft, AlloPro, Baar, Swizerland). Under cyclic loading up to 1600 N hip joint forces, the interface motion between implants and bone was measured at six locations. Results. Bath stems showed uncritical interface motions below 43 mum. However, the Option 3000 stem exhibited significantly smaller motions in the proximal region and slightly larger movements in the distal regions than the Weber prosthesis. Conclusions. The new type of partially cemented stem provided a comparable initial stability to the fully cemented Weber prosthesis.
引用
收藏
页码:750 / 755
页数:6
相关论文
共 50 条
  • [1] Comparative micromotion of fully and proximally cemented femoral stems
    Bachus, KN
    Bloebaum, RD
    Jones, RE
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1999, (366) : 248 - 257
  • [2] Initial stability of cemented femoral stems as a function of surface finish, collar, and stem size
    Ebramzadeh, E
    Sangiorgio, SN
    Longjohn, DB
    Buhari, CE
    Dorr, LD
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2004, 86A (01): : 106 - 115
  • [3] In vitro stability of cemented and cementless femoral stems with compaction
    Chareancholvanich, K
    Bourgeault, CA
    Schmidt, AH
    Gustilo, RB
    Lew, WD
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2002, (394) : 290 - 302
  • [4] Influence of stem geometry on the stability of polished tapered cemented femoral stems
    Glyn-Jones, S
    Gill, HS
    Beard, DJ
    McLardy-Smith, P
    Murray, DW
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2005, 87B (07): : 921 - 927
  • [5] The Paradox of the Decline of Cemented Femoral Stems
    Dujardin, Franck
    De Sulauze, Pierre
    Lalevee, Matthieu
    Matsoukis, Jean
    Curado, Jonathan
    [J]. ARTHROPLASTY TODAY, 2023, 19
  • [6] Cemented femoral stems in patients with DDH
    Garcia-Cimbrelo, E.
    [J]. HIP INTERNATIONAL, 2007, 17 (02) : S128 - S133
  • [7] RETRIEVAL OF FRACTURED CEMENTED FEMORAL STEMS
    WELCH, RB
    MCGANN, WA
    FLOYD, T
    PICETTI, GD
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1989, 71A (06): : 918 - 919
  • [8] Corrosion of cemented titanium femoral stems
    Thomas, SR
    Shukla, D
    Latham, PD
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2004, 86B (07): : 974 - 978
  • [9] METHOD DEVELOPMENT FOR DETERMINATION OF INITIAL STABILITY OF CEMENTLESS FEMORAL STEMS
    Scholl, Laura Yanoso
    Schmidig, Gregg
    Thakore, Mayur
    Hayes, Daniel
    [J]. PROCEEDINGS OF THE ASME SUMMER BIOENGINEERING CONFERENCE - 2009, PT A AND B, 2009, : 885 - 886
  • [10] Cemented femoral stems are not all created equal!
    Barrack, RL
    [J]. ORTHOPEDICS, 2002, 25 (09) : 939 - 940