Corrosion of cemented titanium femoral stems

被引:28
|
作者
Thomas, SR [1 ]
Shukla, D [1 ]
Latham, PD [1 ]
机构
[1] Kettering & Dist Gen Hosp, Northants, England
来源
关键词
D O I
10.1302/0301-620X.86B7.14812
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Cemented titanium stems in hip arthroplasty are associated with proximal cement-stem debonding and early failure. This was well publicised with the 3M Capital hip. However, corrosion in this setting has been reported with only one stem design and is less widely accepted. We present a series of 12 cemented titanium Furlong Straight Stems which required revision at a mean of 78 months for thigh pain. At revision the stems were severely corroded in a pattern which was typical of crevice corrosion. Symptoms were eliminated after revision to an all-stainless steel femoral prosthesis of the same design. We discuss the likely causes for the corrosion. The combination of a titanium stem and cement appears to facilitate crevice corrosion.
引用
收藏
页码:974 / 978
页数:5
相关论文
共 50 条
  • [1] Galvanically enhanced fretting-crevice corrosion of cemented femoral stems
    Bryant, M.
    Farrar, R.
    Freeman, R.
    Brummitt, K.
    Nolan, J.
    Neville, A.
    [J]. JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2014, 40 : 275 - 286
  • [2] Crevice corrosion of cemented titanium alloy stems in total hip replacements
    Willert, HG
    Broback, LG
    Buchhorn, GH
    Jensen, PH
    Koster, G
    Lang, I
    Ochsner, P
    Schenk, R
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1996, (333) : 51 - 75
  • [3] In-vitro characteristics of cemented titanium femoral stems with a smooth surface finish
    Akiyama, Haruhiko
    Yamamoto, Koji
    Kaneuji, Ayumi
    Matsumoto, Tadami
    Nakamura, Takashi
    [J]. JOURNAL OF ORTHOPAEDIC SCIENCE, 2013, 18 (01) : 29 - 37
  • [4] The Paradox of the Decline of Cemented Femoral Stems
    Dujardin, Franck
    De Sulauze, Pierre
    Lalevee, Matthieu
    Matsoukis, Jean
    Curado, Jonathan
    [J]. ARTHROPLASTY TODAY, 2023, 19
  • [5] Cemented femoral stems in patients with DDH
    Garcia-Cimbrelo, E.
    [J]. HIP INTERNATIONAL, 2007, 17 (02) : S128 - S133
  • [6] RETRIEVAL OF FRACTURED CEMENTED FEMORAL STEMS
    WELCH, RB
    MCGANN, WA
    FLOYD, T
    PICETTI, GD
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1989, 71A (06): : 918 - 919
  • [7] Clinical and biological assessment of cemented titanium femoral stems: an 11-year experience
    Patrick Boyer
    Jean-Yves Lazennec
    Joel Poupon
    Marc-Antoine Rousseau
    Philippe Ravaud
    Yves Catonné
    [J]. International Orthopaedics, 2009, 33 : 1209 - 1215
  • [8] Clinical and biological assessment of cemented titanium femoral stems: an 11-year experience
    Boyer, Patrick
    Lazennec, Jean-Yves
    Poupon, Joel
    Rousseau, Marc-Antoine
    Ravaud, Philippe
    Catonne, Yves
    [J]. INTERNATIONAL ORTHOPAEDICS, 2009, 33 (05) : 1209 - 1215
  • [9] Cemented femoral stems are not all created equal!
    Barrack, RL
    [J]. ORTHOPEDICS, 2002, 25 (09) : 939 - 940
  • [10] Loosening of matt and polished cemented femoral stems
    Howie, DW
    Middleton, RG
    Costi, K