Design and validation of a robotic needle positioning system for small animal imaging applications

被引:0
|
作者
Waspe, Adam C. [1 ,2 ]
Cakiroglu, H. Jason [1 ,2 ]
Lacefield, James C. [1 ,2 ]
Fenster, Aaron [1 ,2 ]
机构
[1] Univ Western Ontario, London, ON N6A 5B9, Canada
[2] John P Robarts Res Inst, London, ON N6A 5K8, Canada
基金
加拿大创新基金会; 加拿大自然科学与工程研究理事会;
关键词
image-guided intervention; medical robotics; percutaneous needle insertion; three-dimensional imaging; preclinical imaging;
D O I
暂无
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A needle-positioning robot has been developed for image-guided interventions in small animals. The device is designed to position a needle with an error < 100 mu m. The robot has two rotational axes (pitch and roll) to control needle orientation, and one linear axis to perform needle insertion. The three axes intersect at a single point, creating a remote center of motion. Needle positioning error was quantified at ten target locations for each rotational plane. The measured needle positioning accuracy in free space was 54 +/- 12 mu m and 91 +/- 21 mu m for the pitch and roll axes, respectively. The device's accuracy compares favorably with the SiZes of iypical interventional targets in mouse models.
引用
收藏
页码:237 / 240
页数:4
相关论文
共 50 条
  • [31] An integrated fibered confocal microscopy system for in vivo and in situ fluorescence imaging -: Applications to endoscopy in small animal imaging
    Perchant, A
    Le Goualher, G
    Genet, M
    Viellerobe, B
    Berier, F
    2004 2ND IEEE INTERNATIONAL SYMPOSIUM ON BIOMEDICAL IMAGING: MACRO TO NANO, VOLS 1 AND 2, 2004, : 692 - 695
  • [32] Design and validation of a medical robotic device system to control two collaborative robots for ultrasound-guided needle insertions
    Berger, Johann
    Unger, Michael
    Keller, Johannes
    Reich, C. Martin
    Neumuth, Thomas
    Melzer, Andreas
    FRONTIERS IN ROBOTICS AND AI, 2022, 9
  • [33] Design of γ-ray optics for small animal imaging.
    Pivovaroff, MJ
    Craig, W
    Ziock, K
    Barber, W
    Iwata, K
    Hasegawa, BH
    JOURNAL OF NUCLEAR MEDICINE, 2002, 43 (05) : 231P - 232P
  • [34] Pinhole aperture design for small-animal imaging
    Hu, CM
    Chen, JC
    Liu, RS
    Small Animal SPECT Imaging, 2005, : 183 - 187
  • [35] Design and characterization of a CT scanner for small animal imaging
    Pekar, J.
    Patterson, M.
    MEDICAL PHYSICS, 2006, 33 (07) : 2663 - 2663
  • [36] Integration of the thermal imaging system into technological and robotic applications
    Pancík, J
    Scepko, P
    Havlík, S
    Sekerka, V
    ISMCR '98: PROCEEDINGS OF THE EIGHTH INTERNATIONAL SYMPOSIUM ON MEASUREMENT AND CONTROL IN ROBOTICS, 1998, : 139 - 144
  • [37] MARESye: A hybrid imaging system for underwater robotic applications
    Pinto, Andry Maykol
    Matos, Anibal C.
    INFORMATION FUSION, 2020, 55 : 16 - 29
  • [38] Robotic Fish Wireless Communication and Positioning System Design Based on ZigBee
    Hu, Qingsong
    Zhou, Shuxin
    Xu, Lihong
    ADVANCES IN ELECTRONIC COMMERCE, WEB APPLICATION AND COMMUNICATION, VOL 2, 2012, 149 : 375 - +
  • [39] A collaborative robotic uterine positioning system for laparoscopic hysterectomy: Design and experiments
    Yip, Hiu Man
    Wang, Zerui
    Navarro-Alarcon, David
    Li, Peng
    Cheung, Tak Hong
    Greiffenhagen, Christian
    Liu, Yun-hui
    INTERNATIONAL JOURNAL OF MEDICAL ROBOTICS AND COMPUTER ASSISTED SURGERY, 2020, 16 (04):
  • [40] An Image-Guided Robot for Needle Insertion in Small Animal. Accurate Needle Positioning using Visual Servoing.
    Ayadi, A.
    Bayle, B.
    Graebling, P.
    Gangloff, J.
    2008 IEEE/RSJ INTERNATIONAL CONFERENCE ON ROBOTS AND INTELLIGENT SYSTEMS, VOLS 1-3, CONFERENCE PROCEEDINGS, 2008, : 1453 - 1458