Indirect oscillometric and direct blood pressure measurements in anesthetized and conscious neonatal foals

被引:27
|
作者
Nout, YS
Corley, KTT
Donaldson, LL
Furr, MO
机构
[1] Virginia Maryland Reg Coll Vet Med, Marison DuPont Scott Equine Med Ctr, Leesburg, VA USA
[2] Ohio State Univ, Columbus, OH 43210 USA
[3] Univ London Royal Vet Coll, Hatfield, Herts, England
来源
关键词
animals newborn; arterial catheterization; automated sphygmomanometry; comparative study; horses; indirect blood pressure;
D O I
10.1046/j.1435-6935.2002.00027.x
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
Objective: To investigate the agreement between indirect oscillometric and direct blood pressure measurement in the equine neonate. Design: Prospective observational study. Setting: University Veterinary Teaching Hospital. Animals: Ten crossbred foals of 30-46 hours of age. Interventions: Six animals (Group 1) were anesthetized. Four animals (Group 2) were restrained on a mat. All animals were instrumented with a catheter in the greater metatarsal artery and an oscillometric blood pressure cuff over the coccygeal artery. Blood pressure was varied with dobutamine, phenylephrine, nitroprusside, and increased depth of anesthesia (Group 1) or dopamine (Group 2). Measurements and main results: Simultaneous direct and indirect blood pressure measurements were obtained from the greater metatarsal artery and the coccygeal artery, respectively. There was good agreement between the 2 methods for mean and diastolic blood pressures in both groups, but not for systolic pressure. The agreement was best in mean blood pressure of anesthetized foals (mean bias -1.07; limits of agreement -9.39, 7.25 mmHg). Conclusions: Indirect oscillometry appears lobe an acceptable method for measuring mean arterial blood pressure in both anesthetized and conscious neonatal foals, and may be a valid method of monitoring critically ill foals.
引用
收藏
页码:75 / 80
页数:6
相关论文
共 50 条
  • [1] Comparison of direct and indirect (oscillometric) measurements of arterial blood pressure in conscious dogs
    Bodey, AR
    Michell, AR
    Bovee, KC
    Buranakurl, C
    Garg, T
    [J]. RESEARCH IN VETERINARY SCIENCE, 1996, 61 (01) : 17 - 21
  • [2] Comparison of the indirect oscillometric and direct arterial methods for blood pressure measurements in anesthetized dogs
    Meurs, KM
    Miller, MW
    Slater, MR
    [J]. JOURNAL OF THE AMERICAN ANIMAL HOSPITAL ASSOCIATION, 1996, 32 (06) : 471 - 475
  • [3] Evaluation of oscillometric and Doppler ultrasonic devices for blood pressure measurements in anesthetized and conscious dogs
    Vachon, Catherine
    Belanger, Marie C.
    Burns, Patrick M.
    [J]. RESEARCH IN VETERINARY SCIENCE, 2014, 97 (01) : 111 - 117
  • [4] A COMPARISON OF DIRECT AND INDIRECT (OSCILLOMETRIC) MEASUREMENTS OF ARTERIAL BLOOD-PRESSURE IN ANESTHETIZED DOGS, USING TAIL AND LIMB CUFFS
    BODEY, AR
    YOUNG, LE
    BARTRAM, DH
    DIAMOND, MJ
    MICHELL, AR
    [J]. RESEARCH IN VETERINARY SCIENCE, 1994, 57 (03) : 265 - 269
  • [5] COMPARISON OF DIRECT AND INDIRECT BLOOD-PRESSURE MEASUREMENTS IN ANESTHETIZED DOGS
    GAINS, MJ
    GRODECKI, KM
    JACOBS, RM
    DYSON, D
    FOSTER, RA
    [J]. CANADIAN JOURNAL OF VETERINARY RESEARCH-REVUE CANADIENNE DE RECHERCHE VETERINAIRE, 1995, 59 (03): : 238 - 240
  • [6] Accuracy of indirect measurement of blood pressure in neonatal foals
    Giguère, S
    Knowles, HA
    Valverde, A
    Bucki, E
    Young, L
    [J]. JOURNAL OF VETERINARY INTERNAL MEDICINE, 2005, 19 (04) : 571 - 576
  • [7] EVALUATION OF AGREEMENT BETWEEN OSCILLOMETRIC AND DIRECT BLOOD PRESSURE MEASUREMENTS IN ANESTHETIZED TIGERS (PANTHERA TIGRIS)
    Smith, Christopher K.
    Chow, Natalie
    Zhu, Xiaojuan
    Cushing, Andrew C.
    [J]. JOURNAL OF ZOO AND WILDLIFE MEDICINE, 2022, 53 (04) : 777 - 784
  • [8] Agreement between oscillometric and direct blood pressure measurements in anesthetized captive chimpanzees (Pan troglodytes)
    Smith, Christopher K.
    Ashley, Anthony L.
    Zhu, Xiaojuan
    Cushing, Andrew C.
    [J]. AMERICAN JOURNAL OF VETERINARY RESEARCH, 2021, 82 (12) : 963 - 969
  • [9] Evaluation of an indirect oscillometric blood pressure monitor in normotensive and hypotensive anesthetized dogs
    Shih, Andre
    Robertson, Sheilah
    Vigani, Alessio
    da Cunha, Anderson
    Pablo, Luisito
    Bandt, Carsten
    [J]. JOURNAL OF VETERINARY EMERGENCY AND CRITICAL CARE, 2010, 20 (03) : 313 - 318
  • [10] Doppler and oscillometric mean blood pressure best represent direct blood pressure measurements in anesthetized rhesus macaques (Macaca mulatta)
    Kang, Stacey Carol
    Jampachaisri, Katechan
    Pacharinsak, Cholawat
    [J]. JOURNAL OF MEDICAL PRIMATOLOGY, 2019, 48 (02) : 123 - 128