Radiotherapy-induced malfunctions of cardiac implantable electronic devices in cancer patients

被引:13
|
作者
Malavasi, Vincenzo L. [1 ]
De Marco, Giuseppina [2 ]
Imberti, Jacopo F. [1 ]
Placentino, Filippo [1 ]
Vitolo, Marco [1 ]
Mazzeo, Ercole [2 ]
Cicoria, Gianfranco [3 ]
Casali, Edoardo [1 ]
Turco, Vincenzo [1 ]
Lohr, Frank [2 ]
Boriani, Giuseppe [1 ]
机构
[1] Univ Modena & Reggio Emilia, Policlin Modena, Dept Biomed Metab & Neural Sci, Div Cardiol, Via Pozzo 71, I-41124 Modena, Italy
[2] Univ Modena & Reggio Emilia, Policlin Modena, Dept Oncol, Div Radiotherapy, Modena, Italy
[3] Policlin Modena, Dept Med Phys, Modena, Italy
关键词
Cancer; Cardiac implantable electrical device; Cardio-oncology; Oncology; Radiotherapy; CARDIOVERTER-DEFIBRILLATOR MALFUNCTIONS; RADIATION-THERAPY; MANAGEMENT; PACEMAKERS; EXPOSURE; RISK;
D O I
10.1007/s11739-019-02240-y
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The number of patients with cardiac implantable electronic devices (CIEDs) requiring radiation therapy (RT) for cancer treatment is increasing. The purpose of this study is to estimate the prevalence, possible predictors, and clinical impact of RT-related CIEDs malfunctions. We retrospectively reviewed the medical records of all pacemaker (PM)/implantable cardioverter-defibrillator (ICD) patients who underwent RT in the last 14 years. One hundred and twenty-seven patients who underwent 150 separate RT courses were analysed (99 with a PM and 27 with an ICD). Of note, 21/127 (16.6%) patients were PM-dependent. Neutron-producing RT was used in 37/139 (26.6%) courses, whereas non-neutron-producing RT was used in 102/139 (73.4%) courses. The cumulative dose (Dmax) delivered to the CIED exceeded 5 Gy only in 2/132 (1.5%) cases. Device malfunctions were observed in 3/150 (2%) RT courses, but none was life-threatening or led to a major clinical event and all were resolved by CIED reprogramming. In all cases, the Dmax delivered to the CIED was < 2 Gy. Two malfunctions occurred in the 37 patients treated with neutron-producing RT (5.4%), and 1 malfunction occurred in the 102 patients treated with non-neutron-producing RT (1%) (p = 0.17). Device relocation from the RT field was performed in 2/127 (1.6%) patients. RT in patients with CIED is substantially safe if performed in an appropriately organized environment, with uncommon CIEDs malfunctions and no major clinical events. Neutron-producing energies, rather than Dmax, seem to increase the risk of malfunctions. Device interrogation on a regular basis is advised to promptly manage CIED malfunctions.
引用
下载
收藏
页码:967 / 973
页数:7
相关论文
共 50 条
  • [21] MRI in patients with cardiac implantable electronic devices
    Cronin, Edmond M.
    Mahon, Niall
    Wilkoff, Bruce L.
    EXPERT REVIEW OF MEDICAL DEVICES, 2012, 9 (02) : 139 - 146
  • [22] MRI in Patients with Cardiac Implantable Electronic Devices
    Muthalaly, Rahul G.
    Nerlekar, Nitesh
    Ge, Yin
    Kwong, Raymond Y.
    Nasis, Arthur
    RADIOLOGY, 2018, 289 (02) : 281 - 292
  • [23] A review and analysis of stereotactic body radiotherapy and radiosurgery of patients with cardiac implantable electronic devices
    Hossein Aslian
    Tomas Kron
    Francesco Longo
    Roya Rad
    Mara Severgnini
    Australasian Physical & Engineering Sciences in Medicine, 2019, 42 : 415 - 425
  • [24] MRI of Patients with Cardiac Implantable Electronic Devices
    Jessica A. Martinez
    Daniel B. Ennis
    Current Cardiovascular Imaging Reports, 2019, 12
  • [25] A review and analysis of stereotactic body radiotherapy and radiosurgery of patients with cardiac implantable electronic devices
    Aslian, Hossein
    Kron, Tomas
    Longo, Francesco
    Rad, Roya
    Severgnini, Mara
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2019, 42 (02) : 415 - 425
  • [26] Radiotherapy for Patients with Cardiovascular Implantable Electronic Devices: A Review
    Yeung, Cynthia
    Chacko, Sanoj
    Glover, Benedict
    Campbell, Debra
    Crystal, Eugene
    Ben-Dov, Nissan
    Baranchuk, Adrian
    CANADIAN JOURNAL OF CARDIOLOGY, 2018, 34 (03) : 244 - 251
  • [27] Radiotherapy-Induced Fatigue in Breast Cancer Patients
    Kowalczyk, Lidia
    Deutschmann, Christine
    Crevenna, Richard
    Konrad, Stefan
    Singer, Christian F.
    Farr, Alex
    BREAST CARE, 2021, 16 (03) : 236 - 242
  • [28] Management of cardiac implantable devices in patients undergoing radiotherapy
    Dobson, Rebecca
    Wright, David J.
    CURRENT PROBLEMS IN CANCER, 2018, 42 (04) : 443 - 448
  • [29] Incidence of Radiotherapy Induced Cardiac Implantable Electronic Devices Malfunction: Australian -Based Observation Study
    Lin, Y. H.
    Azraai, M.
    Miura, D.
    Rodrigues, T.
    Nadurata, V.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2023, 117 (02): : E408 - E408
  • [30] Safety of electronic massagers in patients with cardiac implantable electronic devices
    Chia, Pow-Li
    Mok, Kwang-How
    Wong, Shiun Woei
    Foo, David
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2021, 44 (01): : 167 - 170