Analysis of wall-embedded Langmuir probe signals in different conditions on the Tokamak a Configuration Variable

被引:54
|
作者
Fevrier, O. [1 ]
Theiler, C. [1 ]
De Oliveira, H. [1 ]
Labit, B. [1 ]
Fedorczak, N. [2 ]
Baillod, A. [1 ]
机构
[1] Ecole Polytech Fed Lausanne, SPC, CH-1015 Lausanne, Switzerland
[2] CEA, IRFM, F-13108 St Paul Les Durance, France
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 2018年 / 89卷 / 05期
基金
瑞士国家科学基金会;
关键词
DIII-D DIVERTOR; ALCATOR C-MOD; ENERGY-DISTRIBUTION; MAGNETIC-FIELD; W-FUNCTION; PLASMAS; TCV; RECOMBINATION; DISCHARGES; TARGET;
D O I
10.1063/1.5022459
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
This paper presents the current wall-embedded Langmuir probe system installed on the Tokamak a Configuration Variable (TCV), as well as the analysis tool chain used to interpret the current-voltage characteristic obtained when the probes are operated in swept-bias conditions. The analysis is based on a four-parameter fit combined with a minimum temperature approach. In order to reduce the effect of plasma fluctuations and measurement noise, several current-voltage characteristics are usually averaged before proceeding to the fitting. The impact of this procedure on the results is investigated, as well as the possible role of finite resistances in the circuitry, which could lead to an overestimation of the temperature. We study the application of the procedure in a specific regime, the plasma detachment, where results from other diagnostics indicate that the electron temperature derived from the Langmuir probes might be overestimated. To address this issue, we explore other fitting models and, in particular, an extension of the asymmetric double probe fit, which features effects of sheath expansion. We show that these models yield lower temperatures (up to approximately 60%) than the standard analysis in detached conditions, particularly for a temperature peak observed near the plasma strike point, but a discrepancy with other measurements remains. We explore a possible explanation for this observation, the presence of a fast electron population, and assess how robust the different methods are in such conditions.
引用
收藏
页数:15
相关论文
共 50 条
  • [1] Langmuir probe electronics upgrade on the tokamak a configuration variable
    De Oliveira, H.
    Marmillod, P.
    Theiler, C.
    Chavan, R.
    Fevrier, O.
    Labit, B.
    Lavanchy, P.
    Marletaz, B.
    Pitts, R. A.
    Coda, S.
    Ahn, J.
    Albanese, R.
    Alberti, S.
    Alessi, E.
    Allan, S.
    Anand, H.
    Anastassiou, G.
    Andrebe, Y.
    Angioni, C.
    Ariola, M.
    Bernert, M.
    Beurskens, M.
    Bin, W.
    Blanchard, P.
    Blanken, T. C.
    Boedo, J. A.
    Bolzonella, T.
    Bouquey, F.
    Braunmueller, F. H.
    Bufferand, H.
    Buratti, P.
    Calabro, G.
    Camenen, Y.
    Carnevale, D.
    Carpanese, F.
    Causa, F.
    Cesario, R.
    Chapman, I. T.
    Chellai, O.
    Choi, D.
    Cianfarani, C.
    Ciraolo, G.
    Citrin, J.
    Costea, S.
    Crisanti, F.
    Cruz, N.
    Czarnecka, A.
    Decker, J.
    De Masi, G.
    De Tommasi, G.
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 2019, 90 (08):
  • [2] Comparison of measurements with different types of divertor Langmuir probe in EAST tokamak
    Xu, J. C.
    Wang, L.
    Xu, G. S.
    Yip, C-S
    Liu, J. B.
    Feng, W.
    Meng, L. Y.
    [J]. JOURNAL OF INSTRUMENTATION, 2019, 14
  • [3] Electron thermal transport analysis in Tokamak a Configuration Variable
    Asp, E.
    Kim, J-H
    Horton, W.
    Porte, L.
    Alberti, S.
    Karpushov, A.
    Martin, Y.
    Sauter, O.
    Turri, G.
    Team, T. C. V.
    [J]. PHYSICS OF PLASMAS, 2008, 15 (08)
  • [4] Poloidal mode analysis of magnetic probe data in a spherical tokamak configuration
    Tojo, H.
    Ejiri, A.
    Gryaznevich, M. P.
    Takase, Y.
    Adachi, Y.
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 2008, 79 (10):
  • [5] COMPARATIVE-STUDY OF LANGMUIR PROBE CHARACTERISTICS IN DIFFERENT IONOSPHERIC CONDITIONS
    MARGOTCHAKER, J
    MCNAMARA, AG
    [J]. PLANETARY AND SPACE SCIENCE, 1984, 32 (11) : 1427 - 1437
  • [6] Analysis of magnetic probe signals including effect of cylindrical conducting wall for field-reversed configuration experiment
    Ikeyama, Taeko
    Hiroi, Masanori
    Nemoto, Yuuichi
    Nogi, Yasuyuki
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 2008, 79 (06):
  • [7] Analysis of cylindrical Langmuir probe using experiment and different theories
    Hassouba, M. A.
    Galaly, A. R.
    Rashed, U. M.
    [J]. PLASMA PHYSICS REPORTS, 2013, 39 (03) : 255 - 262
  • [8] Analysis of cylindrical Langmuir probe using experiment and different theories
    M. A. Hassouba
    A. R. Galaly
    U. M. Rashed
    [J]. Plasma Physics Reports, 2013, 39 : 255 - 262
  • [9] A fast-reciprocating probe array for two-dimensional measurements in the divertor region of the Tokamak a configuration variable
    De Oliveira, H.
    Theiler, C.
    Elaian, H.
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 2021, 92 (04):
  • [10] Boundary plasma behavior under different wall conditions on HT-7 tokamak
    Xu, GS
    Song, M
    Wan, BN
    Ling, BL
    Li, CF
    Li, J