Iodine Vapor Staining for Atomic Number Contrast in Backscattered Electron and X-Ray Imaging

被引:37
|
作者
Boyde, Alan [1 ]
Mccorkell, Fergus A. [2 ]
Taylor, Graham K. [2 ]
Bomphrey, Richard J. [3 ]
Doube, Michael [4 ]
机构
[1] Queen Mary Univ London, Barts & London Sch Med & Dent, London E1 4NS, England
[2] Univ Oxford, Dept Zool, Oxford Anim Flight Grp, Oxford OX1 3PS, England
[3] Royal Vet Coll, Dept Comparat Biomed Sci, Struct & Mot Lab, Hatfield AL9 7TA, Herts, England
[4] Univ London Royal Vet Coll, Dept Comparat Biomed Sci, London NW1 0TU, England
基金
英国惠康基金; 英国工程与自然科学研究理事会;
关键词
SEM; microCT; desublimation; attenuation; segmentation; soft tissue; histology; MICRO-COMPUTED TOMOGRAPHY; SHRINKAGE; TISSUES; AGENTS;
D O I
10.1002/jemt.22435
中图分类号
R602 [外科病理学、解剖学]; R32 [人体形态学];
学科分类号
100101 ;
摘要
Iodine imparts strong contrast to objects imaged with electrons and X-rays due to its high atomic number (53), and is widely used in liquid form as a microscopic stain and clinical contrast agent. We have developed a simple technique which exploits elemental iodine's sublimation-deposition state-change equilibrium to vapor stain specimens with iodine gas. Specimens are enclosed in a gas-tight container along with a small mass of solid I-2. The bottle is left at ambient laboratory conditions while staining proceeds until empirically determined completion (typically days to weeks). We demonstrate the utility of iodine vapor staining by applying it to resin-embedded tissue blocks and whole locusts and imaging them with backscattered electron scanning electron microscopy (BSE SEM) or X-ray microtomography (XMT). Contrast is comparable to that achieved with liquid staining but without the consequent tissue shrinkage, stain pooling, or uneven coverage artefacts associated with immersing the specimen in iodine solutions. Unmineralized tissue histology can be read in BSE SEM images with good discrimination between tissue components. Organs within the locust head are readily distinguished in XMT images with particularly useful contrast in the chitin exoskeleton, muscle and nerves. Here, we have used iodine vapor staining for two imaging modalities in frequent use in our laboratories and on the specimen types with which we work. It is likely to be equally convenient for a wide range of specimens, and for other modalities which generate contrast from electron- and photon-sample interactions, such as transmission electron microscopy and light microscopy. Microsc. Res. Tech. 77:1044-1051, 2014. (c) 2014 The Authors. Microscopy Research Technique published by Wiley Periodocals, Inc.
引用
收藏
页码:1044 / 1051
页数:8
相关论文
共 50 条
  • [1] Can X-ray Spectrum Imaging (XSI) Replace Backscattered Electron (BSE) Imaging for Compositional Contrast in the Scanning Electron Microscope?
    Newbury, Dale E.
    MICROSCOPY AND MICROANALYSIS, 2009, 15 : 666 - 667
  • [2] Can X-ray Spectrum Imaging Replace Backscattered Electrons for Compositional Contrast in the Scanning Electron Microscope?
    Newbury, Dale E.
    Ritchie, Nicholas W. M.
    SCANNING, 2011, 33 (03) : 174 - 192
  • [3] Comparison of Rhenium and Iodine as Contrast Agents in X-Ray Imaging
    De la Vega, Jose Carlos
    Esquinas, Pedro Luis
    Gill, Jovan Kaur
    Jessa, Selin
    Gill, Bradford
    Thakur, Yogesh
    Saatchi, Katayoun
    Hafeli, Urs O.
    CONTRAST MEDIA & MOLECULAR IMAGING, 2021, 2021
  • [4] Mean Atomic Number Quantitative Assessment in Backscattered Electron imaging
    Sanchez, E.
    Torres Deluigi, M.
    Castellano, G.
    MICROSCOPY AND MICROANALYSIS, 2012, 18 (06) : 1355 - 1361
  • [5] Phase contrast: the frontier of x-ray and electron imaging PREFACE
    Hwu, Y.
    Margaritondo, G.
    JOURNAL OF PHYSICS D-APPLIED PHYSICS, 2013, 46 (49)
  • [6] Refraction contrast in X-ray imaging
    Keyriläinen, J
    Fernández, M
    Suortti, P
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2002, 488 (1-2): : 419 - 427
  • [7] X-Ray Imaging with Phase Contrast
    Bonse, Ulrich
    Beckmann, Felix
    INTERNATIONAL CONFERENCE ON ADVANCED PHASE MEASUREMENT METHODS IN OPTICS AN IMAGING, 2010, 1236 : 189 - +
  • [8] Phase contrast X-ray imaging
    Weon, Byung Mook
    Je, Jung Ho
    Hwu, Yeukuang
    Margaritondo, Giorgio
    INTERNATIONAL JOURNAL OF NANOTECHNOLOGY, 2006, 3 (2-3) : 280 - 297
  • [9] X-ray phase contrast imaging
    Xiong, SS
    Chen, JW
    Gao, HY
    Xie, HL
    Li, RX
    Xu, ZZ
    ISTM/2003: 5TH INTERNATIONAL SYMPOSIUM ON TEST AND MEASUREMENT, VOLS 1-6, CONFERENCE PROCEEDINGS, 2003, : 3715 - 3718
  • [10] Phase contrast X-ray imaging
    Biomedical Imaging Center, Department of Materials Science and Engineering, Pohang University of Science and Technology, Pohang, 790-784, Korea, Republic of
    不详
    不详
    不详
    不详
    不详
    不详
    不详
    不详
    不详
    不详
    Int. J. Nanotechnol., 2006, 2-3 (280-297):