Continuous measurement of cardiac output using stochastic system identification techniques

被引:0
|
作者
Yelderman, M [1 ]
机构
[1] Monterey Med Solut, CEO, Monterey, CA USA
关键词
system theory; patient monitoring; continuous cardiac output; stochastic;
D O I
暂无
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
Indicator dilutions techniques offer the most reliable methods of determining clinical cardiac output because of the elastic nature of the cardiac vessels. A catheter-mounted heating filament affords a simple means of supplying "heat" indicator, but is power and temperature limited because of possible patient injury. A stochastic signal processing method using pseudorandom binary infusion of heat offers a process of enhancing the signal to noise sufficiently to facilitate a computation of cardiac output over a reasonable time period (5 min) with a clinically acceptable error.
引用
收藏
页码:5360 / 5362
页数:3
相关论文
共 50 条
  • [1] CONTINUOUS MEASUREMENT OF CARDIAC-OUTPUT WITH THE USE OF STOCHASTIC-SYSTEM IDENTIFICATION TECHNIQUES
    YELDERMAN, M
    JOURNAL OF CLINICAL MONITORING, 1990, 6 (04): : 322 - 332
  • [2] Influence of continuous renal replacement therapy on cardiac output measurement using thermodilution techniques
    Heise, D.
    Faulstich, M.
    Moerer, O.
    Braeuer, A.
    Quintel, M.
    MINERVA ANESTESIOLOGICA, 2012, 78 (03) : 315 - 321
  • [3] Validation study of PulseCO system for continuous cardiac output measurement
    Berberian, G
    Quinn, TA
    Vigilance, DW
    Park, DY
    Cabreriza, SE
    Curtis, LJ
    Spotnitz, HM
    ASAIO JOURNAL, 2005, 51 (01) : 37 - 40
  • [4] A CATHETER MOUNTED DOPPLER SYSTEM FOR THE CONTINUOUS MEASUREMENT OF CARDIAC-OUTPUT
    ROTH, AC
    KALITA, S
    SUCHY, H
    PROCEEDINGS OF THE ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, PTS 1-4, 1988, : 1778 - 1778
  • [5] Intraoperative validation of a new system for invasive continuous cardiac output measurement
    Giuseppe D’Ancona
    Matteo Parrinello
    Gianluca Santise
    Domenico Biondo
    Francesco Pirone
    Sergio Sciacca
    Marco Turrisi
    Antonio Arcadipane
    Michele Pilato
    Intensive Care Medicine, 2009, 35 : 943 - 947
  • [6] Intraoperative validation of a new system for invasive continuous cardiac output measurement
    D'Ancona, Giuseppe
    Parrinello, Matteo
    Santise, Gianluca
    Biondo, Domenico
    Pirone, Francesco
    Sciacca, Sergio
    Turrisi, Marco
    Arcadipane, Antonio
    Pilato, Michele
    INTENSIVE CARE MEDICINE, 2009, 35 (05) : 943 - 947
  • [7] LINEAR STOCHASTIC SYSTEM IDENTIFICATION USING CORRELATION TECHNIQUES
    BOOZER, DD
    MCDANIEL, WL
    JOURNAL OF THE FRANKLIN INSTITUTE-ENGINEERING AND APPLIED MATHEMATICS, 1973, 296 (01): : 59 - 69
  • [8] Validation of a continuous cardiac output measurement using arterial pressure waveforms
    W McGee
    J Horswell
    G Janvier
    Critical Care, 9 (Suppl 1):
  • [9] CONTINUOUS MEASUREMENT OF CARDIAC-OUTPUT USING PULSE CONTOUR ANALYSIS
    IRLBECK, M
    FORST, H
    BRIEGEL, J
    HALLER, M
    PETER, K
    ANAESTHESIST, 1995, 44 (07): : 493 - 500
  • [10] CONTINUOUS THERMAL MEASUREMENT OF CARDIAC-OUTPUT
    PHILIP, JH
    LONG, MC
    QUINN, MD
    NEWBOWER, RS
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1984, 31 (05) : 393 - 400