Fluid dynamics of a pediatric ventricular assist device

被引:45
|
作者
Bachmann, C
Hugo, G
Rosenberg, G
Deutsch, S
Fontaine, A
Tarbell, JM [1 ]
机构
[1] Penn State Univ, Dept Chem Engn, Fenske Lab 155, University Pk, PA 16802 USA
[2] Penn State Univ, Dept Bioengn, University Pk, PA 16802 USA
[3] Penn State Univ, Milton S Hershey Med Ctr, Div Artificial Organs, Hershey, PA 17033 USA
[4] Penn State Univ, Appl Res Lab, University Pk, PA 16802 USA
关键词
pediatric ventricular assist device; fluid mechanics; laser Doppler velocimetry; dimensional analysis; artificial heart;
D O I
10.1046/j.1525-1594.2000.06536.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The number of pediatric patients requiring some form of mechanical circulatory assistance is growing throughout the world because of new surgical procedures and the success of pediatric cardiac transplantation. However, the salvage rate for those patients requiring circulatory support may be as low as 25%. Despite the fact that Penn State's 70 cc pneumatic ventricular assist device has been used with a success rate of over 90% in more than 250 patients worldwide, efforts to scale down the pump have encountered difficulties. Animal experiments with a 15 cc version were unsuccessful, with explanted pumps showing extensive thrombus deposition within the pumping chamber. The materials used to fabricate the smaller pump as well as the basic operating principles are identical to the successful adult-sized version. It is therefore believed that reducing the size of the pump altered the internal flow field, and that fluid dynamic factors were responsible for the high degree of thrombus observed with the implanted devices. A dimensional analysis was conducted that revealed significant differences in both Reynolds (Re) and Strouhal (St) numbers between the successful and unsuccessful pumps. Two component laser Doppler velocimetry was then used to characterize the internal flow field quantitatively. Comparison with data from the 70 cc pump showed a reduction in wall shear stress and turbulence levels in the 15 cc pump that would yield an environment conducive to clot formation.
引用
收藏
页码:362 / 372
页数:11
相关论文
共 50 条
  • [41] Computational Fluid Dynamics Analysis of Thrombosis Potential in Left Ventricular Assist Device Drainage Cannulae
    Fraser, Katharine H.
    Zhang, Tao
    Taskin, M. Ertan
    Griffith, Bartley P.
    Wu, Zhongjun J.
    ASAIO JOURNAL, 2010, 56 (03) : 157 - 163
  • [42] Characterization and Development of Universal Ventricular Assist Device: Computational Fluid Dynamics Analysis of Advanced Design
    Goodin, Mark S.
    Showalter, Michael S.
    Horvath, David J.
    Kuban, Barry D.
    Flick, Christine R.
    Polakowski, Anthony R.
    Fukamachi, Kiyotaka
    Karimov, Jamshid H.
    ASAIO JOURNAL, 2022, 68 (08) : 1024 - 1035
  • [43] Design and transient computational fluid dynamics study of a continuous axial flow ventricular assist device
    Song, XW
    Untaroiu, A
    Wood, HG
    Allaire, PE
    Throckmorton, AL
    Day, SW
    Olsen, DB
    ASAIO JOURNAL, 2004, 50 (03) : 215 - 224
  • [44] Transient and quasi-steady computational fluid dynamics study of a left ventricular assist device
    Song, XW
    Throckmorton, AL
    Wood, HG
    Allaire, PE
    Olsen, DB
    ASAIO JOURNAL, 2004, 50 (05) : 410 - 417
  • [45] Dynamics of Progenitor Cells and Ventricular Assist Device Intervention
    Tsiavou, Anastasia
    Manginas, Athanassios
    JOURNAL OF CARDIOVASCULAR TRANSLATIONAL RESEARCH, 2010, 3 (02) : 147 - 152
  • [46] Dynamics of Progenitor Cells and Ventricular Assist Device Intervention
    Anastasia Tsiavou
    Athanassios Manginas
    Journal of Cardiovascular Translational Research, 2010, 3 : 147 - 152
  • [47] Thrombus Prediction in Adult and Pediatric Pulsatile Ventricular Assist Devices: The Role of Experimental Fluid Dynamics
    Nanna, J. C.
    Roszelle, B. N.
    Cooper, B. T.
    Yang, N.
    Reddy, V.
    Deutsch, S.
    Manning, K. B.
    2010 IEEE 36TH ANNUAL NORTHEAST BIOENGINEERING CONFERENCE, 2010,
  • [48] Mechanical thrombectomy for pediatric acute stroke and ventricular assist device
    Stowe, Robert Clinton
    Kan, Peter
    Breen, Day Burruss
    Agarwal, Sonika
    BRAIN & DEVELOPMENT, 2018, 40 (01): : 81 - 84
  • [49] Biventricular support with HeartWare ventricular assist device in a pediatric patient
    Deshpande, Shriprasad R.
    Carroll, Mallory M.
    Mao, Chad
    Mahle, William T.
    Kanter, Kirk
    PEDIATRIC TRANSPLANTATION, 2018, 22 (01)
  • [50] In Vivo Evaluation of the "TinyPump" as a Pediatric Left Ventricular Assist Device
    Kitao, Takashi
    Ando, Yusuke
    Yoshikawa, Masaharu
    Kobayashi, Mariko
    Kimura, Taro
    Ohsawa, Hideyuki
    Machida, Shinya
    Yokoyama, Naoyuki
    Sakota, Daisuke
    Konno, Tomohiro
    Ishihara, Kazuhiko
    Takatani, Setsuo
    ARTIFICIAL ORGANS, 2011, 35 (05) : 543 - 553