Studies of MEMS Acoustic Sensors as Implantable Microphones for Totally Implantable Hearing-Aid Systems

被引:30
|
作者
Ko, Wen H. [1 ]
Zhang, Rui [1 ]
Huang, Ping [1 ]
Guo, Jun [1 ]
Ye, Xuesong [1 ]
Young, Darrin J. [1 ]
Megerian, Cliff A. [2 ]
机构
[1] Case Western Reserve Univ, Dept Elect Engn & Comp Sci, Case Sch Engn, Cleveland, OH USA
[2] Case Western Reserve Univ, Dept Otolaryngol, Sch Med, Cleveland, OH USA
关键词
Displacement sensor; implantable microphone; implantable sensor; microelectromechanical-system (MEMS) acoustic sensors; totally implantable hearing systems; vibration sensor; DEVICE;
D O I
10.1109/TBCAS.2009.2032267
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
There is a need for high-quality implantable microphones for existing semiimplantable middle-ear hearing systems and cochlear prosthesis to make them totally implantable, thus overcoming discomfort, inconvenience, and social stigma. This paper summarizes and compares the results of an in-vitro study on three design approaches and the feasibility of using microelectromechanical system acoustic sensors as implantable microphones to convert the umbo vibration directly into a high-quality sound signal. The requirements of sensors were selected including the ability to withstand large body shocks or sudden changes of air pressure. Umbo vibration characteristics were extracted from literature and laboratory measurement data. A piezoelectric vibration source was built and calibrated to simulate the umbo vibration. Two laboratory models of the acoustic sensor were studied. The model-A device, using electrets-microphone as the sensor, was designed and tested in the laboratory and on temporal bones. The results verify that the laboratory measurement is consistent with the temporal bone characterization and achieves a near flat frequency response with a minimum detectable signal of a 65-dB sound-pressure-level (SPL) at 1 kHz. The model-B sensor was then designed to increase the sensitivity and provide an easy mounting on umbo. The model-B device can detect 40-dB SPL sound in the 1-2 kHz region, with 100-Hz channel bandwidth. The results of model-A and model-B displacement sensors and the acceleration sensor are summarized and compared. A preliminary design of the implantable displacement sensor for totally implantable hearing-aid systems is also presented.
引用
收藏
页码:277 / 285
页数:9
相关论文
共 50 条
  • [31] ADVANTAGES OF DIRECTIONAL HEARING-AID MICROPHONES RELATED TO ROOM ACOUSTICS
    LEEUW, AR
    DRESCHLER, WA
    AUDIOLOGY, 1991, 30 (06): : 330 - 344
  • [32] Fault Modeling of Implantable MEMS Sensors
    Miguel, J. A.
    Lechuga, Y.
    Martinez, M.
    Berrazueta, J. R.
    BIODEVICES: PROCEEDINGS OF THE 2015 INTERNATIONAL CONFERENCE ON BIOMEDICAL ELECTRONICS AND DEVICES, 2015, : 162 - 167
  • [33] RADIO HEARING-AID SYSTEMS
    NOLAN, M
    TEACHER OF THE DEAF, 1983, 7 (04): : 105 - 121
  • [34] Human studies of a piezoelectric transducer and a microphone for a totally implantable electronic hearing device
    Zenner, HP
    Leysieffer, H
    Maassen, M
    Lehner, R
    Lenarz, T
    Baumann, J
    Keiner, S
    Plinkert, PK
    McElveen, JT
    AMERICAN JOURNAL OF OTOLOGY, 2000, 21 (02): : 196 - 204
  • [35] Fully implantable hearing systems
    Maurer, J.
    HNO, 2009, 57 (03) : 199 - +
  • [36] ACOUSTIC-EMISSION - A HEARING-AID FOR HYDRIL
    不详
    MATERIALS EVALUATION, 1981, 39 (11) : 1008 - 1009
  • [37] Optimized flextensional transducers for an implantable hearing aid
    Schmitt, D
    Weber, PK
    Pelinescu, I
    Park, SE
    2002 IEEE ULTRASONICS SYMPOSIUM PROCEEDINGS, VOLS 1 AND 2, 2002, : 1273 - 1276
  • [38] IMPLANTABLE HEARING-AID USING AN OSSICULAR VIBRATOR COMPOSED OF A PIEZOELECTRIC CERAMIC BIMORPH - APPLICATION TO 4 PATIENTS
    YANAGIHARA, N
    YAMANAKA, E
    GYO, K
    AMERICAN JOURNAL OF OTOLOGY, 1987, 8 (03): : 213 - 219
  • [39] Totally implantable hearing system: Five-year hearing results
    Shohet, Jack A.
    Kraus, Eric M.
    Catalano, Peter J.
    Toh, Elizabeth
    LARYNGOSCOPE, 2018, 128 (01): : 210 - 216
  • [40] Modeling of Faulty Implantable MEMS Pressure Sensors
    Miguel, J. A.
    Lechuga, Y.
    Martinez, M.
    Bracho, S.
    2014 26TH INTERNATIONAL CONFERENCE ON MICROELECTRONICS (ICM), 2014, : 80 - 83