Localization of cardiac volume and patient features in inverse geometry x-ray fluoroscopy

被引:0
|
作者
Speidel, Michael A. [1 ,2 ]
Slagowski, Jordan M. [1 ]
Dunkerley, David A. P. [1 ]
Wagner, Martin [1 ]
Funk, Tobias [3 ]
Raval, Amish N. [2 ]
机构
[1] Univ Wisconsin, Dept Med Phys, 1530 Med Sci Ctr, Madison, WI 53706 USA
[2] Univ Wisconsin, Dept Med, Madison, WI USA
[3] Triple Ring Technol Inc, Newark, CA USA
基金
美国国家卫生研究院;
关键词
inverse geometry; x-ray fluoroscopy; scanning-beam digital x-ray; tomosynthesis; interventional cardiology; SELECTION;
D O I
10.1117/12.2254400
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
The scanning-beam digital x-ray (SBDX) system is an inverse geometry x-ray fluoroscopy technology that performs real-time tomosynthesis at planes perpendicular to the source-detector axis. The live display is a composite image which portrays sharp features (e.g. coronary arteries) extracted from a 16 cm thick reconstruction volume. We present a method for automatically determining the position of the cardiac volume prior to acquisition of a coronary angiogram. In the algorithm, a single non-contrast frame is reconstructed over a 44 cm thickness using shift-and-add digital tomosynthesis. Gradient filtering is applied to each plane to emphasize features such as the cardiomediastinal contour, diaphragm, and lung texture, and then sharpness vs. plane position curves are generated. Three sharpness metrics were investigated: average gradient in the bright field, maximum gradient, and the number of normalized gradients exceeding 0.5. A model correlating the peak sharpness in a non-contrast frame and the midplane of the coronary arteries in a contrast-enhanced frame was established using 37 SBDX angiographic loops (64-136 kg human subjects, 0-30 degrees cranial-caudal). The average gradient in the bright field (primarily lung) and the number of normalized gradients >0.5 each yielded peaks correlated to the coronary midplane. The rms deviation between the predicted and true midplane was 1.57 cm. For a 16 cm reconstruction volume and the 5.5-11.5 cm thick cardiac volumes in this study, midplane estimation errors of 2.25-5.25 cm were tolerable. Tomosynthesis-based localization of cardiac volume is feasible. This technique could be applied prior to coronary angiography, or to assist in isocentering the patient for rotational angiography.
引用
收藏
页数:11
相关论文
共 50 条
  • [21] Calibration-Free Coronary Artery Measurements for Interventional Device Sizing using Inverse Geometry X-ray Fluoroscopy: In Vivo Validation
    Tomkowiak, Michael T.
    Raval, Amish N.
    Van Lysel, Michael S.
    Funk, Tobias
    Speidel, Michael A.
    MEDICAL IMAGING 2014: PHYSICS OF MEDICAL IMAGING, 2014, 9033
  • [22] THE X-RAY FOVEA, A DEVICE FOR REDUCING X-RAY DOSE IN FLUOROSCOPY - RESPONSE
    WILSON, D
    RZESZOTARSKI, MS
    MEDICAL PHYSICS, 1994, 21 (12) : 1887 - 1887
  • [23] THE X-RAY FOVEA, A DEVICE FOR REDUCING X-RAY DOSE IN FLUOROSCOPY - COMMENT
    WALMSLEY, JJ
    MEDICAL PHYSICS, 1994, 21 (12) : 1885 - 1885
  • [24] X-ray beam equalization for digital fluoroscopy
    Molloi, S
    Tang, J
    Marcin, M
    Zhou, YF
    Anvar, B
    PHYSICS OF MEDICAL IMAGING: MEDICAL IMAGING 1996, 1996, 2708 : 167 - 178
  • [25] Fuzzy segmentation of X-ray fluoroscopy images
    Russakoff, DB
    Rohlfing, T
    Maurer, CR
    MEDICAL IMAGING 2002: IMAGE PROCESSING, VOL 1-3, 2002, 4684 : 146 - 154
  • [26] A STUDY OF MOTION IN GASTROINTESTINAL X-RAY FLUOROSCOPY
    NGUYEN, TC
    ROWLANDS, JA
    MEDICAL PHYSICS, 1989, 16 (04) : 569 - 576
  • [27] Effects of motion blurring in x-ray fluoroscopy
    Xue, P
    Wilson, DL
    MEDICAL PHYSICS, 1998, 25 (05) : 587 - 599
  • [28] An iterative algorithm for X-ray CT fluoroscopy
    Wang, G
    Schweiger, G
    Vannier, MW
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 1998, 17 (05) : 853 - 856
  • [29] An iterative algorithm for X-ray CT fluoroscopy
    Wang, G
    Schweiger, GD
    Vannier, MW
    RADIOLOGY, 1998, 209P : 434 - 434
  • [30] 3D TEE Registration with X-Ray Fluoroscopy for Interventional Cardiac Applications
    Jain, Ameet
    Gutierrez, Luis
    Stanton, Douglas
    FUNCTIONAL IMAGING AND MODELING OF THE HEART, PROCEEDINGS, 2009, 5528 : 321 - 329