Localization of cardiac volume and patient features in inverse geometry x-ray fluoroscopy

被引:0
|
作者
Speidel, Michael A. [1 ,2 ]
Slagowski, Jordan M. [1 ]
Dunkerley, David A. P. [1 ]
Wagner, Martin [1 ]
Funk, Tobias [3 ]
Raval, Amish N. [2 ]
机构
[1] Univ Wisconsin, Dept Med Phys, 1530 Med Sci Ctr, Madison, WI 53706 USA
[2] Univ Wisconsin, Dept Med, Madison, WI USA
[3] Triple Ring Technol Inc, Newark, CA USA
基金
美国国家卫生研究院;
关键词
inverse geometry; x-ray fluoroscopy; scanning-beam digital x-ray; tomosynthesis; interventional cardiology; SELECTION;
D O I
10.1117/12.2254400
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
The scanning-beam digital x-ray (SBDX) system is an inverse geometry x-ray fluoroscopy technology that performs real-time tomosynthesis at planes perpendicular to the source-detector axis. The live display is a composite image which portrays sharp features (e.g. coronary arteries) extracted from a 16 cm thick reconstruction volume. We present a method for automatically determining the position of the cardiac volume prior to acquisition of a coronary angiogram. In the algorithm, a single non-contrast frame is reconstructed over a 44 cm thickness using shift-and-add digital tomosynthesis. Gradient filtering is applied to each plane to emphasize features such as the cardiomediastinal contour, diaphragm, and lung texture, and then sharpness vs. plane position curves are generated. Three sharpness metrics were investigated: average gradient in the bright field, maximum gradient, and the number of normalized gradients exceeding 0.5. A model correlating the peak sharpness in a non-contrast frame and the midplane of the coronary arteries in a contrast-enhanced frame was established using 37 SBDX angiographic loops (64-136 kg human subjects, 0-30 degrees cranial-caudal). The average gradient in the bright field (primarily lung) and the number of normalized gradients >0.5 each yielded peaks correlated to the coronary midplane. The rms deviation between the predicted and true midplane was 1.57 cm. For a 16 cm reconstruction volume and the 5.5-11.5 cm thick cardiac volumes in this study, midplane estimation errors of 2.25-5.25 cm were tolerable. Tomosynthesis-based localization of cardiac volume is feasible. This technique could be applied prior to coronary angiography, or to assist in isocentering the patient for rotational angiography.
引用
收藏
页数:11
相关论文
共 50 条
  • [1] Monoplane Stereoscopic Imaging Method for Inverse Geometry X-ray Fluoroscopy
    Tomkowiak, Michael T.
    Van Lysel, Michael S.
    Speidel, Michael A.
    MEDICAL IMAGING 2013: IMAGE PROCESSING, 2013, 8669
  • [2] Inverse Geometry X-Ray Fluoroscopy Using Scanning-Beam Digital X-Ray Technology
    Speidel, M.
    MEDICAL PHYSICS, 2011, 38 (06) : 3774 - +
  • [3] Three-dimensional tracking of cardiac catheters using an inverse geometry x-ray fluoroscopy system
    Speidel, Michael A.
    Tomkowiak, Michael T.
    Raval, Amish N.
    Van Lysel, Michael S.
    MEDICAL PHYSICS, 2010, 37 (12) : 6377 - 6389
  • [4] Registration of 3D Transesophageal Echocardiography and X-Ray Fluoroscopy Using An Inverse Geometry X-Ray System
    Speidel, M.
    Hatt, C.
    Tomkowiak, M.
    Funk, T.
    Raval, A.
    MEDICAL PHYSICS, 2014, 41 (06) : 471 - 471
  • [5] Depth-resolved registration of transesophageal echo to x-ray fluoroscopy using an inverse geometry fluoroscopy system
    Hatt, Charles R.
    Tomkowiak, Michael T.
    Dunkerley, David A. P.
    Slagowski, Jordan M.
    Funk, Tobias
    Raval, Amish N.
    Speidel, Michael A.
    MEDICAL PHYSICS, 2015, 42 (12) : 7022 - 7033
  • [6] THE RAY DOSAGE OF THE PATIENT IN X-RAY FLUOROSCOPY AND PHOTOGRAPHY
    WEGELIUS, C
    NORDISK MEDICIN, 1946, 32 (47) : 2684 - 2686
  • [7] PATIENT X-RAY EXPOSURE DURING FLUOROSCOPY
    ARDRAN, GM
    PHYSICS IN MEDICINE AND BIOLOGY, 1968, 13 (03): : 469 - &
  • [8] HOW SAFE IS X-RAY AND FLUOROSCOPY FOR THE PATIENT AND THE DOCTOR
    CLEIN, NW
    JOURNAL OF PEDIATRICS, 1954, 45 (03): : 310 - 315
  • [9] Monte Carlo simulation of inverse geometry x-ray fluoroscopy using a modified MC-GPU framework
    Dunkerley, David A. P.
    Tomkowiak, Michael T.
    Slagowski, Jordan M.
    McCabe, Bradley P.
    Funk, Tobias
    Speidel, Michael A.
    MEDICAL IMAGING 2015: PHYSICS OF MEDICAL IMAGING, 2015, 9412
  • [10] A Device Enhancing and Denoising Algorithm for X-ray Cardiac Fluoroscopy
    Bismuth, Vincent
    Vaillant, Regis
    19TH INTERNATIONAL CONFERENCE ON PATTERN RECOGNITION, VOLS 1-6, 2008, : 1802 - 1805