THE IMPACT OF AUTOMATIC WEDGE FILTER ON PHOTONEUTRON AND PHOTON SPECTRA OF AN 18-MV PHOTON BEAM

被引:6
|
作者
Ghavami, Seyyed-Mostafa [2 ]
Mesbahi, Asghar [1 ,3 ]
Mohammadi, Ehsan [4 ]
机构
[1] Tabriz Univ Med Sci, Sch Med, Dept Med Phys, Tabriz, Iran
[2] Tabriz Univ Med Sci, Paramed Sch, Dept Radiol, Tabriz, Iran
[3] Imam Hosp, Dept Radiat Therapy, Tabriz, Iran
[4] Imam khomeini Hosp, Dept Radiat Therapy, Tehran, Iran
关键词
DYNAMIC MLC IMRT; MONTE-CARLO; LINEAR-ACCELERATOR; NEUTRON-SPECTRA; 15; MV; RADIOTHERAPY; LINAC; SIEMENS; MODEL;
D O I
10.1093/rpd/ncp204
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
The effect of all automatic wedge filter oil photon and photoneutron spectra of a medical linae was evaluated using the Monte Carlo method. The head of an Elekta SL75/25 was simulated using the MCNPX Monte Carlo code. The photon and photoneutron spectra for open and wedged beams were calculated at the isocentre with a source to axis distance of 100 cm. For a wedged beam, the neutron fluence was from 3.84 to 7.2 times higher for field sizes from 10 x 10 to 30 x 30 cm(2). The neutron fluence is decreased with field size for open beams and is increased with field size for wedged beams. The photon beam spectra became harder and file mean energy was 6 % higher for a wedged beam, which led to a 4 % increase in relative depth dose and a better skin sparing effect. The results here recommend that the higher photoneutron fluence of the wedged beam should be taken into account in patient dosimetry and shielding calculations.
引用
收藏
页码:123 / 128
页数:6
相关论文
共 50 条
  • [41] The Impact of 6MV Non-Reference Photon Energy Spectra on OSLD Response
    Scarboro, S.
    Followill, D.
    Kerns, J.
    Kry, S.
    MEDICAL PHYSICS, 2011, 38 (06)
  • [42] Therapeutic applications of 2.5 MV photon beam in radiation therapy
    Khaledi, Navid
    Yao, Bonnie
    Khan, Rao
    Grafe, James L.
    MEDICAL PHYSICS, 2022, 49 (08) : 5660 - 5660
  • [43] Treatment planning with a 2.5 MV photon beam for radiation therapy
    Khaledi, Navid
    Hayes, Chris
    Belshaw, Louise
    Grattan, Mark
    Khan, Rao
    Grafe, James L.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2022, 23 (12):
  • [44] Dosimetric Properties of a Transmission Detector in a 6 MV Photon Beam
    Nakaguchi, Y.
    Araki, F.
    Saiga, S.
    MEDICAL PHYSICS, 2010, 37 (06)
  • [45] Measurement of percentage dose at the surface for a 6 MV photon beam
    Galvan De la Cruz, O. O.
    Rodriguez-Avila, M. A.
    Rivera-Montalvo, T.
    Garcia Garduno, O. A.
    REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2019, 24 (06) : 585 - 592
  • [46] The influence of field size and off-axis distance on photoneutron spectra of the 18 MV Siemens Oncor linear accelerator beam
    Brkic, Hrvoje
    Ivkovic, Ana
    Kasabasic, Mladen
    Sovilj, Marina Poje
    Jurkovic, Slaven
    Stimac, Damir
    Rubin, Otmar
    Faj, Dario
    RADIATION MEASUREMENTS, 2016, 93 : 28 - 34
  • [47] MONTE CARLO SIMULATION OF 6, 10 AND 18 MV PHOTON BEAM DOSE DISTRIBUTION IN A BRAIN TUMOR
    Gonzales-Ccoscco, Alberto E.
    Guzman-Calcina, Carmen S.
    Vega-Ramirez, Jose L.
    MOMENTO-REVISTA DE FISICA, 2023, (67): : 22 - 38
  • [48] Characterization of wedge factors and dose distributions in radiotherapy with symmetric and asymmetric physical wedged beams of 6 MV photon beam
    Zabihzadeh M.
    Fadaei M.
    Hoseini S.M.
    Arvandi S.
    Tahmasbi M.
    Iranian Journal of Medical Physics, 2020, 17 (03) : 213 - 219
  • [49] CHARACTERISTICS OF AN 18 MV PHOTON-BEAM FROM A THERAC 20 MEDICAL LINEAR-ACCELERATOR
    PATTERSON, MS
    SHRAGGE, PC
    MEDICAL PHYSICS, 1981, 8 (03) : 312 - 318
  • [50] Dosimetric properties of a flattening filter-free 6-MV photon beam: A Monte Carlo study
    Mesbahi A.
    Mehnati P.
    Keshtkar A.
    Farajollahi A.
    Radiation Medicine, 2007, 25 (7): : 315 - 324