Assessment of an instrument for automated reagent and specimen dispensing in blood banking

被引:0
|
作者
Coovadia, AS [1 ]
Armstrong, C [1 ]
Chan, R [1 ]
Pinkerton, PH [1 ]
机构
[1] Sunnybrook Hlth Sci Ctr, Dept Lab Med, Div Clin Pathol, N York, ON M4N 3M51, Canada
来源
TRANSFUSION SCIENCE | 1997年 / 18卷 / 04期
关键词
D O I
暂无
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
We have evaluated a semi-automated computer controlled dispensing device in conjunction with microtube technology for the performance of routine blood grouping and antibody screening procedures. A total of 787 specimens have been tested, 78 with unexpected antibodies previously identified by manual microtube methods. All were tested in duplicate in different order; there was complete agreement in blood group (ABO, RhD) determinations and the unexpected antibodies were found in the appropriate microtube location in all duplicate sets of results. We conclude that the device accurately and reproducibly dispenses cells, sera and reagents. The savings of technologists' time over traditional manual tube methods exceed 75%, excluding the time to perform elements common to both methods. The combination of automation and microtube technology offers the opportunity for substantial savings in technologist time with accuracy in reagent and specimen dispensing. (C) 1997 Elsevier Science Ltd.
引用
收藏
页码:517 / 522
页数:6
相关论文
共 50 条
  • [21] IMPROVEMENT OF PROCEDURE FOR STIRRING AND DISPENSING TO PREPARE ANTI-COAGULATED FRESH BLOOD SAMPLES FOR EXTERNAL QUALITY ASSESSMENT OF AUTOMATED HEMATOLOGY ANALYZER
    Kondo, Hiroshi
    Takahira, Izuru
    Ikeda, Hisataka
    Kondo, Tamiaki
    Sirakami, Atsushi
    Nagai, Yutaka
    Kawai, Yohko
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2016, 38 : 56 - 56
  • [22] Relevance of pediatric control assessment in cord blood banking
    Díaz, A
    Reboredo, NM
    Castro, A
    Cervilla, JR
    Fraga, JM
    Areal, C
    PROCEEDINGS OF THE XIX EUROPEAN CONGRESS OF PERINATAL MEDICINE, 2004, : 421 - 424
  • [23] Automated dispensing of medicines registered in EPSM: assessment of the reason for not administering its cost impact
    Renaud, A.
    Wattecamps, C.
    Blondiaux, S.
    Moens, C.
    Zawadzki, E.
    EUROPEAN PSYCHIATRY, 2014, 29 (08) : 647 - 648
  • [24] PREPARATION OF AN AUTHENTIC WARM AUTOIMMUNE HEMOLYTIC-ANEMIA SPECIMEN FOR INSTRUCTIONAL PURPOSES IN BLOOD BANKING
    CALLICOAT, PA
    STURDIVANT, DL
    SMITH, DM
    JOURNAL OF MEDICAL TECHNOLOGY, 1984, 1 (05): : 398 - 398
  • [25] Assessment of magnetic bead-based automated whole blood RNA-isolation from a validated RNA stabilization reagent (Tempus Blood RNA)
    Vordenbaeumen, Stefan
    Bleck, Ellen
    Brinks, Ralph
    Schneider, Matthias
    JOURNAL OF IMMUNOLOGICAL METHODS, 2014, 406 : 148 - 149
  • [26] ASSESSMENT OF THE CORVAC BLOOD COLLECTION TUBE FOR DRUG SPECIMEN PROCESSING
    DEVINE, JE
    THERAPEUTIC DRUG MONITORING, 1986, 8 (02) : 241 - 243
  • [27] Evaluation of the Galileo Echo™ Automated Blood Bank Instrument for Pretransfusion Testing
    Bouillon, P. A.
    Lowery, J.
    Johnson, A. N.
    Bonner, H.
    Fugate, L.
    TRANSFUSION, 2009, 49 : 145A - 145A
  • [28] EVALUATION OF THE CELLAVISION HEMAPREP SEMI-AUTOMATED BLOOD SMEAR INSTRUMENT
    Ho, E.
    Baetens, D.
    Van Breda, K.
    Malfait, R.
    Van den Bossche, J.
    ACTA CLINICA BELGICA, 2010, 65 (03): : 218 - 218
  • [29] Screening for syphilis and CMV antibodies on an automated blood bank instrument.
    Stone, DL
    Sinor, LT
    Young, R
    Radtke, F
    TRANSFUSION, 1996, 36 (09) : S136 - S136
  • [30] Thermogenesis AXP autoxpress platform and bioarchive system for automated cord blood banking
    Dobrila, L.
    Chapman, J.
    Marr, D.
    Kumar, V.
    Coelho, P.
    Rubinstein, P.
    VOX SANGUINIS, 2006, 91 : 243 - 244