Diagnostic Performance of PET Versus SPECT Myocardial Perfusion Imaging in Patients with Smaller Left Ventricles: A Substudy of the 18F-Flurpiridaz Phase III Clinical Trial

被引:16
|
作者
Packard, Rene R. Sevag [1 ,2 ,3 ]
Lazewatsky, Joel L. [4 ]
Orlandi, Cesare [4 ]
Maddahi, Jamshid [1 ,5 ]
机构
[1] Univ Calif Los Angeles, David Geffen Sch Med, Dept Med, Div Cardiol, Los Angeles, CA 90095 USA
[2] Ronald Reagan UCLA Med Ctr, Los Angeles, CA 90095 USA
[3] Vet Affairs West LA Med Ctr, Los Angeles, CA 90073 USA
[4] Lantheus Med Imaging, North Billerica, MA USA
[5] Univ Calif Los Angeles, David Geffen Sch Med, Dept Mol & Med Pharmacol, Nucl Med Clin, Los Angeles, CA 90095 USA
关键词
diagnostic performance; left ventricle size; SPECT MPI; PET MPI; flurpiridaz; CORONARY-ARTERY-DISEASE; CARDIOVASCULAR MAGNETIC-RESONANCE; POSITRON-EMISSION-TOMOGRAPHY; COMPUTED-TOMOGRAPHY; EJECTION FRACTION; RISK-ASSESSMENT; BLOOD-FLOW; RB-82; PET; ACCURACY; ANGIOGRAPHY;
D O I
10.2967/jnumed.120.252007
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The performance of SPECT myocardial perfusion imaging (MPI) may deteriorate in smaller hearts, primarily because of the lower resolution of conventional Anger cameras. F-18-flurpiridaz is a novel PET MPI agent with superior image and defect resolution. We sought to determine the diagnostic performance of Tc-99m-labeled SPECT MPI compared with F-18-flurpiridaz PET MPI according to left ventricle (LV) size. Methods: We conducted a substudy of the phase III clinical trial of flurpiridaz (n = 750) and stratified diagnostic performance according to the median PET LV end-diastolic volume (LVEDV), with smaller LVs defined as having an LVEDV of less than 113 mL (n = 369) and larger LVs defined as having an LVEDV of at least 113 mL (n = 381). Images were interpreted by the majority rule of 3 independent masked readers. The reference standard was quantitative invasive angiography, with at least 50% stenosis in at least 1 coronary artery considered significant. Results: SPECT performance decreased significantly from an area under the curve (AUC) of 0.75 in larger LVs to 0.67 in smaller LVs (P = 0.03), whereas PET performance was similar in larger and smaller LVs (AUC, 0.79 vs. 0.77, P = 0.49). Accordingly, in smaller LVs, PET had a higher AUC (0.77) than the SPECT AUC (0.67) (P < 0.0001), a phenomenon driven by female patients (P < 0.0001). In smaller LVs, there was a degradation of SPECT sensitivity that was highly significant (P < 0.001), whereas there was no significant change in PET sensitivity according to LV size (P = 0.07). Overall, PET had significantly higher sensitivity than SPECT in both smaller LVs (67% vs. 43%, P < 0.001) and larger LVs (76% vs. 61%, P < 0.001). The specificities of PET and SPECT were similar in larger LVs (76% vs. 83%, P = 0.11). Although SPECT specificity improved in smaller compared with larger LVs (90% vs. 83%, P = 0.03), the PET specificity did not change with LV size (76% vs. 76%, P = 0.9). Conclusion: The diagnostic performance of F-18-flurpiridaz PET MPI is not affected by LV size and is superior to SPECT MPI in patients with smaller LVs, highlighting the importance of appropriate test selection in these patients.
引用
收藏
页码:849 / 854
页数:6
相关论文
共 50 条
  • [1] Diagnostic performance of PET myocardial perfusion imaging with Flurpiridaz F18 is unaffected by cardiac size and is superior to SPECT in patients with smaller left ventricles: a sub-study of the Flurpiridaz F18 phase III clinical trial
    Packard, Rene
    Lazewatsky, Joel
    Orlandi, Cesare
    Maddahi, Jamshid
    JOURNAL OF NUCLEAR MEDICINE, 2017, 58
  • [2] 18F-Flurpiridaz Myocardial Perfusion PET, What Is it That Matters? Diagnostic or Prognostic Accuracy?
    Maddahi, Jamshid
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2024, 83 (07) : e69 - e69
  • [3] FLURPIRIDAZ F 18 PET VERSUS TC-99M SPECT FOR MYOCARDIAL PERFUSION IMAGING
    Berman, D. S.
    Maddahi, J.
    CARDIOLOGY, 2013, 125 : 27 - 27
  • [4] Development, diagnostic performance, and interobserver agreement of a 18F-flurpiridaz PET automated perfusion quantitation system
    René R. Sevag Packard
    C. David Cooke
    Kenneth F. Van Train
    John R. Votaw
    James W. Sayre
    Joel L. Lazewatsky
    Kelly M. Champagne
    Cesare Orlandi
    Ernest V. Garcia
    Jamshid Maddahi
    Journal of Nuclear Cardiology, 2022, 29 : 698 - 708
  • [5] Development, diagnostic performance, and interobserver agreement of a 18F-flurpiridaz PET automated perfusion quantitation system
    Packard, Rene R. Sevag
    Cooke, C. David
    Van Train, Kenneth F.
    Votaw, John R.
    Sayre, James W.
    Lazewatsky, Joel L.
    Champagne, Kelly M.
    Orlandi, Cesare
    Garcia, Ernest, V
    Maddahi, Jamshid
    JOURNAL OF NUCLEAR CARDIOLOGY, 2022, 29 (02) : 698 - 708
  • [6] Diagnostic performance comparison of 18F-flurpiridaz PET myocardial blood flow quantitation methods and software packages
    Packard, Rene
    Buckley, Chris
    Cooke, David
    Ficaro, Edward
    Moody, Jonathan
    Slomka, Piotr
    Van Kriekinge, Serge
    Van Train, Kenneth
    Wangerin, Kristen
    Maddahi, Jamshid
    JOURNAL OF NUCLEAR MEDICINE, 2024, 65
  • [7] Phase 2 clinical comparison of flurpiridaz F 18 injection PET and SPECT myocardial perfusion imaging for diagnosis of coronary artery disease
    Maddahi, Jamshid
    Berman, Daniel
    Taillefer, Raymond
    Udelson, James
    Devine, Marybeth
    Lazewatsky, Joel
    Bhat, Gajanan
    Washburn, Dana
    JOURNAL OF NUCLEAR MEDICINE, 2011, 52
  • [8] Improved Myocardial Blood Flow Estimation with Residual Activity and Motion Correction in 18F-flurpiridaz PET myocardial perfusion imaging
    Otaki, Yuka
    Van Kriekinge, Serge
    Wei, Chih-Chun
    Kavanagh, Paul
    Singh, Ananya
    Parekh, Tejas
    DiCarli, Marcelo
    Buckley, Christpher
    Berman, Daniel
    Slomka, Piotr
    JOURNAL OF NUCLEAR MEDICINE, 2021, 62
  • [9] Phase 2 safety and clinical comparison of flurpiridaz F 18 injection PET and SPECT myocardial perfusion imaging for diagnosis of coronary artery disease
    Maddahi, J.
    Berman, D. S.
    Taillefer, R.
    Udelson, J.
    Devine, M.
    Lazewatsky, J.
    Bhat, G.
    Washburn, D.
    EUROPEAN HEART JOURNAL SUPPLEMENTS, 2011, 13 (0A) : A45 - A45
  • [10] Optimization of myocardial perfusion (MP) PET imaging with 18F-flurpiridaz using realistic computer modeling and simulation techniques
    Wiyaporn, Kanyalak
    Tocharoenchai, Chiraporn
    Park, Min Jae
    Higuchi, Takahiro
    Fung, George
    Tsui, Benjamin
    JOURNAL OF NUCLEAR MEDICINE, 2012, 53