Dose reduction in Full-Field Digital Electromagnetic Radiation Mammography using Noise reduction Method: Phantom Study

被引:0
|
作者
Choi, Seokyoon [1 ]
Seo, Jeongmin [1 ]
机构
[1] Catholic Univ Pusan, Dept Radiol Sci, Busan 46252, South Korea
关键词
high energy electromagnetic beam; electromagnetic beam image; electromagnetic radiation; mammography;
D O I
10.4283/JMAG.2019.24.4.744
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
The purpose of study is to investigate optimal exposure parameters for full-field digital electromagnetic radiation mammography (FFDM) and to reduce the average glandular dose (AGD) by applying a noise reduction algorithm. To find an optimal exposure parameter, filtration-target material combinations (Mo/Rh, Mo/Mo, W/ Rh) were applied by each tube voltages (23-35 kVp). AGD, Noise, Figure-of-merit (FOM) were measured. The experimental results, using 29 kVp of W/Rh was found to be best in all cases before and after applying the denoising filter. With a denoising filter, noise decreased 76.44 % and SNR increased 76.18 %, which improved image quality. If a denoising method is suggested for images of 35 kVp of W/Rh (SNR: 3.58, AGD: 0.67 mGy) using the smallest AGD when using FFDM, better picture quality (SNR 9.73, ADG: 0.67 mGy) could be gotten than 23 kVp of Mo/Mo combination. From this study, it is possible to obtain good quality images with a lower dose than FFDM used in the clinic.
引用
收藏
页码:744 / 751
页数:8
相关论文
共 50 条
  • [21] Diagnostic performance of digital breast tomosynthesis and full-field digital mammography with new reconstruction and new processing for dose reduction
    Tokiko Endo
    Takako Morita
    Mikinao Oiwa
    Namiko Suda
    Yasuyuki Sato
    Shu Ichihara
    Misaki Shiraiwa
    Kazuaki Yoshikawa
    Takao Horiba
    Hirotoshi Ogawa
    Yukie Hayashi
    Tomonari Sendai
    Takahisa Arai
    [J]. Breast Cancer, 2018, 25 : 159 - 166
  • [22] Diagnostic performance of digital breast tomosynthesis and full-field digital mammography with new reconstruction and new processing for dose reduction
    Endo, Tokiko
    Morita, Takako
    Oiwa, Mikinao
    Suda, Namiko
    Sato, Yasuyuki
    Ichihara, Shu
    Shiraiwa, Misaki
    Yoshikawa, Kazuaki
    Horiba, Takao
    Ogawa, Hirotoshi
    Hayashi, Yukie
    Sendai, Tomonari
    Arai, Takahisa
    [J]. BREAST CANCER, 2018, 25 (02) : 159 - 166
  • [23] The effect of the antiscatter grid on full-field digital mammography phantom images
    Chakraborty, DP
    [J]. JOURNAL OF DIGITAL IMAGING, 1999, 12 (01) : 12 - 22
  • [24] The effect of the antiscatter grid on full-field digital mammography phantom images
    Dev P. Chakraborty
    [J]. Journal of Digital Imaging, 1999, 12 : 12 - 22
  • [25] Eye radiation dose from breast cancer screening using full field digital mammography and digital breast tomosynthesis: A phantom study
    Ali, R. M. K. M.
    Hogg, P.
    [J]. RADIOGRAPHY, 2024, 30 (01) : 141 - 144
  • [26] A clinical survey of patient dose in full-field digital mammography
    Hermann, K
    Obenauer, S
    Marten, K
    Kehbel, S
    Fischer, U
    Grabbe, EH
    [J]. RADIOLOGY, 2002, 225 : 269 - 269
  • [27] Patient dose in full-field digital mammography: an Italian survey
    Gisella Gennaro
    Paola Baldelli
    Angelo Taibi
    Cosimo Di Maggio
    Mauro Gambaccini
    [J]. European Radiology, 2004, 14 : 645 - 652
  • [28] Patient dose in full-field digital mammography: an Italian survey
    Gennaro, G
    Baldelli, P
    Taibi, A
    Di Maggio, C
    Gambaccini, M
    [J]. EUROPEAN RADIOLOGY, 2004, 14 (04) : 645 - 652
  • [29] Comparison of a new full-field digital mammography system and conventional film-screen mammography: A phantom study
    Bick, U
    Diekmann, F
    Marth, F
    Le Roux, A
    Juran, R
    Friedrich, M
    [J]. RADIOLOGY, 1999, 213P : 152 - 152
  • [30] X-ray spectrum optimization of full-field digital mammography: Simulation and phantom study
    Bernhardt, Philipp
    Mertelmeier, Thomas
    Hoheisel, Martin
    [J]. MEDICAL PHYSICS, 2006, 33 (11) : 4337 - 4349