EXPERIMENTAL AND MONTE CARLO INVESTIGATIONS OF BCF-12 SMALL-AREA PLASTIC SCINTILLATION DETECTORS FOR NEUTRON PINHOLE CAMERA

被引:0
|
作者
Bielecki, J. [1 ]
Drozdowicz, K. [1 ]
Dworak, D. [1 ]
Igielski, A. [1 ]
Janik, W. [1 ]
Kulinska, A. [1 ]
Marciniak, L. [1 ]
Scholz, M. [1 ]
Turzanski, M. [1 ]
Wiacek, U. [1 ]
Woznicka, U. [1 ]
Wojcik-Gargula, A. [1 ]
机构
[1] Polish Acad Sci IFJ PAN, Inst Nucl Phys, PL-31342 Krakow, Poland
关键词
PLASMA-FOCUS; FUSION;
D O I
10.1093/rpd/ncx277
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
Plastic organic scintillators such as the blue-emitting BCF-12 are versatile and inexpensive tools. Recently, BCF-12 scintillators have been foreseen for investigation of the spatial distribution of neutrons emitted from dense magnetized plasma. For this purpose, small-area (5 mm x 5 mm) detectors based on BCF-12 scintillation rods and Hamamatsu photomultiplier tubes were designed and constructed at the Institute of Nuclear Physics. They will be located inside the neutron pinhole camera of the PF-24 plasma focus device. Two different geometrical layouts and approaches to the construction of the scintillation element were tested. The aim of this work was to determine the efficiency of the detectors. For this purpose, the experimental investigations using a neutron generator and a Pu-Be source were combined with Monte Carlo computations using the Geant4 code.
引用
收藏
页码:427 / 431
页数:5
相关论文
共 4 条
  • [1] Comparison of photon counting and conventional scintillation detectors in a pinhole SPECT system for small animal imaging: Monte carlo simulation studies
    Young-Jin Lee
    Su-Jin Park
    Seung-Wan Lee
    Dae-Hong Kim
    Ye-Seul Kim
    Hee-Joung Kim
    Journal of the Korean Physical Society, 2013, 62 : 1317 - 1322
  • [2] Comparison of photon counting and conventional scintillation detectors in a pinhole SPECT system for small animal imaging: Monte carlo simulation studies
    Lee, Young-Jin
    Park, Su-Jin
    Lee, Seung-Wan
    Kim, Dae-Hong
    Kim, Ye-Seul
    Kim, Hee-Joung
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2013, 62 (09) : 1317 - 1322
  • [3] Study of the response of plastic scintillation detectors in small-field 6 MV photon beams by Monte Carlo simulations
    Wang, Lilie L. W.
    Beddar, Sam
    MEDICAL PHYSICS, 2011, 38 (03) : 1596 - 1599
  • [4] Some experimental and Monte Carlo investigations of the plastic scintillators for the current mode measurements at pulsed neutron sources
    Rogov, A
    Pepyolyshev, Y
    Carta, M
    d'Angelo, A
    ADVANCED MONTE CARLO FOR RADIATION PHYSICS, PARTICLE TRANSPORT SIMULATION AND APPLICATIONS, 2001, : 899 - 904