Early surgical and audiologic outcomes of active, transcutaneous, osseointegrated bone-conduction hearing device (Osia 2® system) placement

被引:18
|
作者
Florentine, Michelle M. [1 ,2 ]
Virbalas, Jordan [1 ]
Chan, Dylan K. [1 ]
机构
[1] Univ Calif San Francisco, Dept Otolaryngol Head & Neck Surg, 2233 Post St Box 1225, San Francisco, CA 94115 USA
[2] Tel Aviv Univ, Sackler Sch Med, Tel Aviv, Israel
关键词
Bone conduction hearing device; Conductive hearing loss; Single-sided deafness; Aural atresia; Osia 2 (R) system; AURAL ATRESIA; PERFORMANCE; MULTICENTER;
D O I
10.1016/j.ijporl.2022.111114
中图分类号
R76 [耳鼻咽喉科学];
学科分类号
100213 ;
摘要
Objective: To assess the outcomes of pediatric Osia 2 (R) System placements. Methods: We performed a retrospective chart review of primary and revision Osia 2 (R) System surgical cases at two tertiary academic children's hospitals. Operative details and post-operative surgical and audiologic outcomes were recorded. Results: 18 cases were performed on 14 children (mean age: 11.5 years, range 7-16) and included 9 primary surgeries and 9 revisions from BAHA Attract (R), Connect (R) and Sophono (R) implants. Surgical planning for revision surgeries was complex due to prior incisions, implants, and bony contour. Post-operative aided audiograms revealed pure tone average (0.5-4 kHz) of 26.2 +/- 2.5 dB HL (mean +/- SD), with no high frequency roll-off (8 kHz aided threshold: 23.8 +/- 7.5 dB HL). Two minor post-operative complications were identified. Conclusion: Bone-conduction hearing devices (BCHDs) are used to improve access to sound for children with conductive hearing loss, single-sided deafness, and aural atresia. Traditional passive, percutaneous abutment-based and transcutaneous magnet-based surgical BCHDs can be limited by skin complications and high-frequency acoustic attenuation. Recent availability of active, transcutaneous osseointegrated BCHD systems presents potential for improvement on both of these traditional limitations. Initial experience with the Osia 2 (R) System demonstrates overall successful, uncomplicated placement with excellent audiologic outcomes. Revision cases require careful surgical planning. Further follow-up and comparative studies with other BCHDs are necessary to fully evaluate the effectiveness of the Osia 2 (R) System.
引用
收藏
页数:6
相关论文
共 50 条
  • [41] The effect of anatomical variables and use of the Lifts system on hearing outcomes after implantation of an active transcutaneous bone conduction device in bilateral congenital conductive hearing loss
    Jinsong Yang
    Chunli Zhao
    Yujie Liu
    Mengdie Gao
    Ran Ren
    Danni Wang
    Zhigang Huang
    Shouqin Zhao
    Journal of Otolaryngology - Head & Neck Surgery, 49
  • [42] A comparative study of audiological outcomes and compliance between the Osia system and other bone conduction hearing implants
    Kim, Yehree
    Choe, Goun
    Oh, Heonjeong
    Choi, Byung Yoon
    EUROPEAN ARCHIVES OF OTO-RHINO-LARYNGOLOGY, 2023, 280 (05) : 2217 - 2224
  • [43] A comparative study of audiological outcomes and compliance between the Osia system and other bone conduction hearing implants
    Yehree Kim
    Goun Choe
    Heonjeong Oh
    Byung Yoon Choi
    European Archives of Oto-Rhino-Laryngology, 2023, 280 : 2217 - 2224
  • [44] A Systematic Review of Surgical Characteristics and Adverse Events of an Active, Transcutaneous Bone Conduction Device
    Jukic, Alma
    Munhall, Christopher C.
    Stevens, Shawn M.
    ANNALS OF OTOLOGY RHINOLOGY AND LARYNGOLOGY, 2024, 133 (11): : 956 - 966
  • [45] Medical, Technical and Audiological Outcomes of Hearing Rehabilitation with the Bonebridge Transcutaneous Bone-Conduction Implant: A Single-Center Experience
    Brkic, Faris F.
    Riss, Dominik
    Scheuba, Katharina
    Arnoldner, Christoph
    Gstoettner, Wolfgang
    Baumgartner, Wolf-Dieter
    Vyskocil, Erich
    JOURNAL OF CLINICAL MEDICINE, 2019, 8 (10)
  • [46] Dura and sinus compression with a transcutaneous bone conduction device - hearing outcomes and safety in 38 patients
    Vyskocil, E.
    Riss, D.
    Arnoldner, C.
    Hamzavi, J. S.
    Liepins, R.
    Kaider, A.
    Honeder, C.
    Fumicz, J.
    Gstoettner, W.
    Baumgartner, W. D.
    CLINICAL OTOLARYNGOLOGY, 2017, 42 (05) : 1033 - 1038
  • [47] Early Outcomes of a New Active Transcutaneous Bone Conduction Implant in Pediatric Patients
    You, Peng
    Choi, Alexander
    Drob, Jennifer
    Hunsaker, Sabrina Marciante
    Liu, Yi-Chun Carol
    Silva, Rodrigo
    OTOLOGY & NEUROTOLOGY, 2022, 43 (02) : 212 - 218
  • [48] Retrosigmoidal placement of an active transcutaneous bone conduction implant: surgical and audiological perspectives in a multicentre study
    Canzi, Pietro
    Avato, Irene
    Beltrame, Millo
    Bianchin, Giovanni
    Perotti, Marco
    Tribi, Lorenzo
    Gioia, Barbara
    Aprile, Federico
    Malpede, Stefano
    Scribante, Andrea
    Manfrin, Marco
    Benazzo, Marco
    ACTA OTORHINOLARYNGOLOGICA ITALICA, 2021, 41 (01) : 91 - 99
  • [49] Early Results With the New Active Bone-Conduction Hearing Implant: A Systematic Review and Meta-Analysis
    Lein, Alexander
    Baumgartner, Wolf-Dieter
    Riss, Dominik
    Gstoettner, Wolfgang
    Landegger, Lukas D.
    Liu, David T.
    Thurner, Thomas
    Vyskocil, Erich
    Brkic, Faris F.
    OTOLARYNGOLOGY-HEAD AND NECK SURGERY, 2024, 170 (06) : 1630 - 1647
  • [50] Evaluation of surgery and the short-term benefits of a new active bone conduction hearing implant - the Osia® system
    Gawecki, Wojciech
    Gibasiewicz, Renata
    Marszal, Joanna
    Blaszczyk, Magdalena
    Gawlowska, Maria
    Wierzbicka, Malgorzata
    BRAZILIAN JOURNAL OF OTORHINOLARYNGOLOGY, 2022, 88 (03) : 289 - 295