The presence of pseudotumours, which are soft-tissue masses relating to the hip, after metal-on-metal hip resurfacing arthroplasty has been associated with elevated levels of metal ions in serum, suggesting that pseudotumours occur when there is increased wear. We aimed to quantify the wear in vivo of implants revised for pseudotumours (eight) and of a control group of implants (22) revised for other reasons of failure. We found that the implant group with pseudotumours had a significantly higher rate of median linear wear of the femoral component at 8.1 mu m/year (2.75 to 25.4) than the 1.79 mu m/year (0.82 to 4.15; p = 0.002) of the non-pseudotumour group. For the acetabular component a significantly higher rate of median linear wear of 7.36 mu m/year (1.61 to 24.9) was observed in the pseudotumour group compared with 1.28 mu m/year (0.81 to 3.33, p = 0.001) in the other group. Wear of the acetabular component in the pseudotumour group always involved the edge of the implant, indicating that edge-loading had occurred. Our findings are the first direct evidence that pseudotumour is associated with increased wear at the metal-on-metal articulation. Furthermore, edge-loading with the loss of fluid-film lubrication may be an important mechanism of generation of wear in patients with a pseudotumour.
引用
收藏
页码:356 / 361
页数:6
相关论文
共 50 条
  • [1] What is the natural history of asymptomatic pseudotumours in metal-on-metal hip resurfacing patients?
    Matharu, Gulraj S.
    Ostlere, Simon J.
    Pandit, Hemant G.
    Murray, David W.
    [J]. HIP INTERNATIONAL, 2016, 26 (06) : 522 - 530
  • [2] Pseudotumours associated with metal-on-metal hip resurfacing: 10-year Newcastle experience
    Malviya, Ajay
    Holland, James P.
    [J]. ACTA ORTHOPAEDICA BELGICA, 2009, 75 (04): : 477 - 483
  • [3] Analysis of 118 second-generation metal-on-metal retrieved hip implants
    Sieber, HP
    Rieker, CB
    Köttig, P
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1999, 81B (01): : 46 - 50
  • [4] In vivo evaluation of edge-loading in metal-on-metal hip resurfacing patients with pseudotumours
    Kwon, Y-M.
    Mellon, S. J.
    Monk, P.
    Murray, D. W.
    Gill, H. S.
    [J]. BONE & JOINT RESEARCH, 2012, 1 (04): : 42 - 49
  • [5] Local and systemic consequences of metal-on-metal hip resurfacing implants
    Campbell, Patricia
    Takamura, Karren
    [J]. ANNALS OF JOINT, 2020, 5 (01):
  • [6] A study of the wear of explanted metal-on-metal resurfacing hip prostheses
    Joyce, Thomas J.
    Langton, David J.
    Nargol, Antoni V. F.
    [J]. TRIBOLOGY INTERNATIONAL, 2011, 44 (05) : 517 - 522
  • [7] Wear Analysis of 39 Conserve Plus Metal-on-Metal Hip Resurfacing Retrievals
    Takamura, Karren M.
    Amstutz, Harlan C.
    Lu, Zhen
    Campbell, Pat A.
    Ebramzadeh, Edward
    [J]. JOURNAL OF ARTHROPLASTY, 2014, 29 (02): : 410 - 415
  • [8] Metal-on-metal hip resurfacing
    Roberts, J
    Meek, RMD
    Roberts, P
    Grigoris, P
    [J]. SCOTTISH MEDICAL JOURNAL, 2005, 50 (01) : 10 - 12
  • [9] Metal-on-Metal Hip Resurfacing
    Quesada, Mario J.
    Marker, David R.
    Mont, Michael A.
    [J]. JOURNAL OF ARTHROPLASTY, 2008, 23 (07): : 69 - 73
  • [10] Metal-on-metal hip resurfacing
    Ebied, A
    Journeaux, S
    [J]. CURRENT ORTHOPAEDICS, 2002, 16 (06): : 420 - 425