Current source design for electrical impedance tomography

被引:105
|
作者
Ross, AS [1 ]
Saulnier, GJ
Newell, JC
Isaacson, D
机构
[1] Rensselaer Polytech Inst, Dept Biomed Engn, Troy, NY 12180 USA
[2] Rensselaer Polytech Inst, Dept Elect & Comp Syst Engn, Troy, NY 12180 USA
[3] Rensselaer Polytech Inst, Dept Math Sci, Troy, NY 12180 USA
关键词
electrical impedance imaging; electrical impedance tomography; current source; generalized impedance converter;
D O I
10.1088/0967-3334/24/2/361
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Questions regarding the feasibility of using electrical impedance tomography (EIT) to detect breast cancer may be answered by building a sufficiently precise multiple frequency EIT instrument. Current sources are desirable for this application, yet no current source designs have been reported that have the required precision at the multiple frequencies needed. We have designed an EIT current source using an enhanced Howland topology in parallel with a generalized impedance converter (GIC). This combination allows for nearly independent adjustment of output resistance and output capacitance, resulting in simulated output impedances in excess of 2 GOmega between 100 Hz and I MHz. In this paper, the theoretical operation of this current source is explained, and experimental results demonstrate the feasibility of creating a high precision, multiple frequency, capacitance compensated current source for EIT applications.
引用
收藏
页码:509 / 516
页数:8
相关论文
共 50 条
  • [1] Design of current source for multi-frequency simultaneous electrical impedance tomography
    Han, Bing
    Xu, Yanbin
    Dong, Feng
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2017, 88 (09):
  • [2] Precision constant current source for Electrical Impedance Tomography
    Lee, JW
    Oh, TI
    Paek, SM
    Lee, JS
    Woo, EJ
    PROCEEDINGS OF THE 25TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-4: A NEW BEGINNING FOR HUMAN HEALTH, 2003, 25 : 1066 - 1069
  • [3] Single Source Current Drive Patterns for Electrical Impedance Tomography
    Zhang, Lifeng
    Wang, Huaxiang
    Zhang, Lifeng
    2010 IEEE INTERNATIONAL INSTRUMENTATION AND MEASUREMENT TECHNOLOGY CONFERENCE I2MTC 2010, PROCEEDINGS, 2010,
  • [4] DSP-based current source for electrical impedance tomography
    Saulnier, Gary J.
    Abdelwahab, Ahmed
    Shishvan, Omid Rajabi
    PHYSIOLOGICAL MEASUREMENT, 2020, 41 (06)
  • [5] The Design of Voltage Controlled Current Source (VCCS) for Single Frequency Electrical Impedance Tomography (EIT)
    Umbu, Ari Bangkit Sanjaya
    Endarko
    2017 INTERNATIONAL SEMINAR ON SENSORS, INSTRUMENTATION, MEASUREMENT AND METROLOGY (ISSIMM), 2017, : 30 - 36
  • [6] Current approaches to analogue instrumentation design in electrical impedance tomography
    Boone, KG
    Holder, DS
    PHYSIOLOGICAL MEASUREMENT, 1996, 17 (04) : 229 - 247
  • [7] A floating wide-band current source for electrical impedance tomography
    Khalighi, M.
    Mikaeili, M.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2018, 89 (08):
  • [8] Electrode Configuration with Howland Current Source and Demodulator for Electrical Impedance Tomography
    Nasiri, Hossein
    Benetti, Rafael
    Takimoto, Rogerio Y.
    Cavalheiro, Andre C. M.
    Duran, Guilherme C.
    Martins, Thiago C.
    Tsuzuki, Marcos S. G.
    IFAC PAPERSONLINE, 2021, 54 (15): : 79 - 84
  • [9] Study of Voltage Control Current Source in Electrical Impedance Tomography System
    Li Zhangyong
    Xu Zhui
    Ren Chaoshi
    Wang Wei
    Zhao Dechun
    Zhang Huiquan
    2010 4TH INTERNATIONAL CONFERENCE ON BIOINFORMATICS AND BIOMEDICAL ENGINEERING (ICBBE 2010), 2010,
  • [10] Verifying of the Multi-Frequency Current Source for Electrical Impedance Tomography
    Huang, J. J.
    Wu, R. J.
    Cheng, K. S.
    4TH KUALA LUMPUR INTERNATIONAL CONFERENCE ON BIOMEDICAL ENGINEERING 2008, VOLS 1 AND 2, 2008, 21 (1-2): : 694 - +