A comparison of the dose distributions from three proton treatment planning systems in the planning of meningioma patients with single-field uniform dose pencil beam scanning

被引:8
|
作者
Doolan, Paul J. [1 ]
Alshaikhi, Jailan [1 ,2 ]
Rosenberg, Ivan [2 ]
Ainsley, Christopher G. [2 ]
Gibson, Adam [1 ]
D'Souza, Derek [2 ]
Bentefour, El Hassane [3 ]
Royle, Gary [1 ]
机构
[1] UCL, Dept Med Phys & Bioengn, London WC1E 6BT, England
[2] Univ Coll London Hosp, Dept Radiotherapy, London, England
[3] Ion Beam Applicat, Louvain La Neuve, Belgium
来源
基金
英国工程与自然科学研究理事会;
关键词
proton therapy; particle therapy; treatment planning; planning comparison; PROSTATE-CANCER; X-RAY; RADIOTHERAPY TREATMENT; RADIATION-THERAPY; INTENSITY; OPTIMIZATION; IMRT; UNCERTAINTIES; DELIVERY; SENSITIVITY;
D O I
10.1120/jacmp.v16i1.4996
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
With the number of new proton centers increasing rapidly, there is a need for an assessment of the available proton treatment planning systems (TPSs). This study compares the dose distributions of complex meningioma plans produced by three proton TPSs: Eclipse, Pinnacle(3), and XiO. All three systems were commissioned with the same beam data and, as best as possible, matched configuration settings. Proton treatment plans for ten patients were produced on each system with a pencil beam scanning, single-field uniform dose approach, using a fixed horizontal beamline. All 30 plans were subjected to identical dose constraints, both for the target coverage and organ at risk (OAR) sparing, with a consistent order of priority. Beam geometry, lateral field margins, and lateral spot resolutions were made consistent across all systems. Few statistically significant differences were found between the target coverage and OAR sparing of each system, with all optimizers managing to produce plans within clinical tolerances (D2 < 107% of prescribed dose, D5 < 105%, D95 > 95%, D99 > 90%, and OAR maximum doses) despite strict constraints and overlapping structures.
引用
收藏
页码:86 / 99
页数:14
相关论文
共 50 条
  • [41] Proton therapy dose distribution comparison between Monte Carlo and a treatment planning system for pediatric patients with ependymoma
    Jia, Yingcui
    Beltran, Chris
    Indelicato, Daniel J.
    Flampouri, Stella
    Li, Zuofeng
    Merchant, Thomas E.
    [J]. MEDICAL PHYSICS, 2012, 39 (08) : 4742 - 4747
  • [42] Measurements of lateral penumbra for uniform scanning proton beams under various beam delivery conditions and comparison to the XiO treatment planning system
    Rana, Suresh
    Zeidan, Omar
    Ramirez, Eric
    Rains, Michael
    Gao, Junfang
    Zheng, Yuanshui
    [J]. MEDICAL PHYSICS, 2013, 40 (09)
  • [43] Searching for Practical Monte Carlo Dose Calculation Parameters of Line-Scanning Proton Beam Therapy in a Commercial Treatment Planning System
    Kim, H.
    Chung, K.
    Han, Y.
    Park, W.
    Park, H.
    Lim, D.
    Choi, D.
    [J]. MEDICAL PHYSICS, 2022, 49 (06) : E872 - E873
  • [44] Analysis of Differences in Tumor Control Probability Calculated From Monte Carlo and Treatment Planning System Dose Distributions for Proton Therapy
    Peeler, C. R.
    Titt, U.
    Grosshans, D. R.
    Liao, Z.
    Mohan, R.
    Mirkovic, D.
    [J]. MEDICAL PHYSICS, 2017, 44 (06)
  • [45] Implementation of a simplified analytical random walk model dose calculation algorithm with nuclear interaction for treatment planning of scanning-beam proton therapy
    Yao, Weiguang
    Moskvin, Vadim
    Zhao, Li
    Lukose, Renin
    Hua, Chia-Ho
    Tsiamas, Panagiotis
    Axente, Marian
    Merchant, Thomas E.
    Farr, Jonathan B.
    [J]. BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2018, 4 (03):
  • [46] Comparison between proton and neutron dose distributions from single-scattering and dual-scattering systems in ocular proton therapy
    Koch, N
    Newhauser, W
    [J]. MEDICAL PHYSICS, 2005, 32 (06) : 2028 - 2028
  • [47] Evaluation of IMRT Plans From Three Treatment Planning Systems Based On Independent Monte Carlo Dose Calculation
    Chi, Z.
    Li, J.
    Jin, L.
    Fan, J.
    Lin, T.
    Xu, Q.
    Ma, C.
    [J]. MEDICAL PHYSICS, 2008, 35 (06)
  • [48] Use of single-energy proton pencil beam scanning Bragg peak for intensity-modulated proton therapy FLASH treatment planning in liver-hypofractionated radiation therapy
    Wei, Shouyi
    Lin, Haibo
    Shi, Chengyu
    Xiong, Weijun
    Chen, Chin-Cheng
    Huang, Sheng
    Press, Robert H.
    Hasan, Shaakir
    Chhabra, Arpit M.
    Choi, J. Isabelle
    Simone, Charles B., II
    Kang, Minglei
    [J]. MEDICAL PHYSICS, 2022, 49 (10) : 6560 - 6574
  • [49] SBRT of lung tumours: Monte Carlo simulation with PENELOPE of dose distributions including respiratory motion and comparison with different treatment planning systems
    Panettieri, Vanessa
    Wennberg, Berit
    Gagliardi, Giovanna
    Duch, Maria Amor
    Ginjaume, Merce
    Lax, Ingmar
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (14): : 4265 - 4281
  • [50] Evaluation of dose distribution differences from five algorithms implemented in three commercial treatment planning systems for lung SBRT
    Sarkar, Vikren
    Paxton, Adam
    Rassiah, Prema
    Kokeny, Kristine E.
    Hitchcock, Ying J.
    Salter, Bill J.
    [J]. JOURNAL OF RADIOSURGERY AND SBRT, 2020, 7 (01): : 57 - 66