A comparison of the dose distributions from three proton treatment planning systems in the planning of meningioma patients with single-field uniform dose pencil beam scanning

被引:8
|
作者
Doolan, Paul J. [1 ]
Alshaikhi, Jailan [1 ,2 ]
Rosenberg, Ivan [2 ]
Ainsley, Christopher G. [2 ]
Gibson, Adam [1 ]
D'Souza, Derek [2 ]
Bentefour, El Hassane [3 ]
Royle, Gary [1 ]
机构
[1] UCL, Dept Med Phys & Bioengn, London WC1E 6BT, England
[2] Univ Coll London Hosp, Dept Radiotherapy, London, England
[3] Ion Beam Applicat, Louvain La Neuve, Belgium
来源
基金
英国工程与自然科学研究理事会;
关键词
proton therapy; particle therapy; treatment planning; planning comparison; PROSTATE-CANCER; X-RAY; RADIOTHERAPY TREATMENT; RADIATION-THERAPY; INTENSITY; OPTIMIZATION; IMRT; UNCERTAINTIES; DELIVERY; SENSITIVITY;
D O I
10.1120/jacmp.v16i1.4996
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
With the number of new proton centers increasing rapidly, there is a need for an assessment of the available proton treatment planning systems (TPSs). This study compares the dose distributions of complex meningioma plans produced by three proton TPSs: Eclipse, Pinnacle(3), and XiO. All three systems were commissioned with the same beam data and, as best as possible, matched configuration settings. Proton treatment plans for ten patients were produced on each system with a pencil beam scanning, single-field uniform dose approach, using a fixed horizontal beamline. All 30 plans were subjected to identical dose constraints, both for the target coverage and organ at risk (OAR) sparing, with a consistent order of priority. Beam geometry, lateral field margins, and lateral spot resolutions were made consistent across all systems. Few statistically significant differences were found between the target coverage and OAR sparing of each system, with all optimizers managing to produce plans within clinical tolerances (D2 < 107% of prescribed dose, D5 < 105%, D95 > 95%, D99 > 90%, and OAR maximum doses) despite strict constraints and overlapping structures.
引用
收藏
页码:86 / 99
页数:14
相关论文
共 50 条
  • [1] Impact of varying planning parameters on proton pencil beam scanning dose distributions in four commercial treatment planning systems
    Alshaikhi, Jailan
    Doolan, Paul J.
    D'Souza, Derek
    Holloway, Stacey McGowan
    Amos, Richard A.
    Royle, Gary
    [J]. MEDICAL PHYSICS, 2019, 46 (03) : 1150 - 1162
  • [2] Evaluation of a Pencil Beam Scanning Proton Dose Model for a Commercial Treatment Planning System
    Patel, D.
    Poenisch, F.
    Sahoo, N.
    Zhu, X.
    Gillin, M.
    Janson, M.
    Mohan, R.
    Titt, U.
    [J]. MEDICAL PHYSICS, 2018, 45 (06) : E261 - E261
  • [3] A comparison of two pencil beam scanning treatment planning systems for proton therapy
    Langner, Ulrich W.
    Mundis, Michelle
    Strauss, Dan
    Zhu, Mingyao
    Mossahebi, Sina
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2018, 19 (01): : 156 - 163
  • [4] Pseudo Proton Radiography Beam Validation of Monte Carlo Dose Calculation in Two Pencil Beam Scanning Treatment Planning Systems
    Chang, C.
    Zhou, J.
    Yang, X.
    Dhabaan, A.
    Zhang, R.
    Liu, T.
    McDonald, M.
    Langen, K.
    Lin, L.
    [J]. MEDICAL PHYSICS, 2019, 46 (06) : E592 - E592
  • [5] A standardized commissioning framework of Monte Carlo dose calculation algorithms for proton pencil beam scanning treatment planning systems
    Chang, Chih-Wei
    Huang, Sheng
    Harms, Joseph
    Zhou, Jun
    Zhang, Rongxiao
    Dhabaan, Anees
    Slopsema, Roelf
    Kang, Minglei
    Liu, Tian
    McDonald, Mark
    Langen, Katja
    Lin, Liyong
    [J]. MEDICAL PHYSICS, 2020, 47 (04) : 1545 - 1557
  • [6] Automated treatment planning for proton pencil beam scanning using deep learning dose prediction and dose-mimicking optimization
    Maes, Dominic
    Holmstrom, Mats
    Helander, Rasmus
    Saini, Jatinder
    Fang, Christine
    Bowen, Stephen R.
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2023, 24 (10):
  • [7] A single-field integrated boost treatment planning technique for spot scanning proton therapy
    Zhu, Xiaorong Ronald
    Poenisch, Falk
    Li, Heng
    Zhang, Xiaodong
    Sahoo, Narayan
    Wu, Richard Y.
    Li, Xiaoqiang
    Lee, Andrew K.
    Chang, Eric L.
    Choi, Seungtaek
    Pugh, Thomas
    Frank, Steven J.
    Gillin, Michael T.
    Mahajan, Anita
    Grosshans, David R.
    [J]. RADIATION ONCOLOGY, 2014, 9
  • [8] A single-field integrated boost treatment planning technique for spot scanning proton therapy
    Xiaorong Ronald Zhu
    Falk Poenisch
    Heng Li
    Xiaodong Zhang
    Narayan Sahoo
    Richard Y Wu
    Xiaoqiang Li
    Andrew K Lee
    Eric L Chang
    Seungtaek Choi
    Thomas Pugh
    Steven J Frank
    Michael T Gillin
    Anita Mahajan
    David R Grosshans
    [J]. Radiation Oncology, 9
  • [9] A Golden Beam Data Commissioning Framework of Monte Carlo Dose Calculation Algorithms of Two Pencil Beam Scanning Proton Therapy Treatment Planning Systems
    Chang, C.
    Harms, J.
    Zhang, R.
    Zhou, J.
    Lin, Y.
    Slopsema, R.
    Dhabaan, A.
    Liu, T.
    McDonald, M.
    Langen, K.
    Lin, L.
    [J]. MEDICAL PHYSICS, 2019, 46 (06) : E167 - E167
  • [10] Comparison of Full Intensity Modulated and Single Field Uniform Dose Proton Beam Scanning Lung Treatments
    Dowdell, S.
    Grassberger, C.
    Paganetti, H.
    [J]. MEDICAL PHYSICS, 2013, 40 (06)