Determination of cortisol and cortisone in human saliva by a liquid chromatography-tandem mass spectrometry method

被引:12
|
作者
Henriques Vieira, Jose Gilberto [1 ]
Nakamura, Odete Hirata [1 ]
Carvalho, Valdemir Melechko [1 ]
机构
[1] Steroid Labs, Grp Fleury, Sao Paulo, Brazil
关键词
Salivary cortisol; salivary cortisone; mass spectrometry; radioimmunoassay; hypercortisolism screening; CORTICOSTEROID-BINDING GLOBULIN; FREE HORMONE HYPOTHESIS; URINARY FREE CORTISOL; CUSHINGS-SYNDROME; RADIOIMMUNOASSAY; DEXAMETHASONE; SPECIFICITY; DIAGNOSIS; STEROIDS; PLASMA;
D O I
10.1590/0004-2730000003347
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective: Salivary cortisol measurement plays an important role in the evaluation of adrenal function. Its high correlation with free serum cortisol, the easy of sampling and the limited presence of interfering steroids, generated multiple recent studies of its application, in special in the screening of adrenal hyperfunction. In this paper we present our experience in the development of a high pressure liquid chromatography tandem mass spectrometry (HPLC-MS/MS) method for salivary cortisol and cortisone measurement. Materials and methods: For this study we used 181 saliva samples from our routine diagnostic laboratory. The HPLC-MS/MS method was based on a Waters Quattro Premier tandem mass spectrometer with an electrospray probe. After derivatization with hydroxylamine transitions monitored included cortisol and cortisone. An in-house radioimmunoassay (RIA) was used for salivary cortisol results comparison. Results: Functional sensitivity was 24 ng/dL for cortisol and linearity from 24 to 1929 ng/dL. Saliva cortisol values obtained in the 181 samples presented a median of 52 ng/dL with 5-95% percentile of 24 and 374 ng/dL. With the RIA the results were 86, 25 and 436 ng/dL, respectively, with values for RIA being significantly higher (P < 0.0001) and high correlation (r = 0.8312, P < 0.0001). Cortisone measured in 159 samples showed a median of 278 ng/dL, with 5-95% percentile of 100 and 1,133 ng/dL. Correlation with cortisol values was significant (r = 0.820, P < 0.0001). Conclusion: We conclude that the HPLC-MS/MS method compares favorably with the RIA for salivary cortisol measurement, with the additional possibility of concomitant cortisone measurement and the evaluation of 11 beta HSD2 activity.
引用
收藏
页码:844 / 850
页数:7
相关论文
共 50 条
  • [1] Simultaneous Determination of Cortisol and Cortisone from Human Serum by Liquid Chromatography-Tandem Mass Spectrometry
    Lee, Sanghoo
    Lim, Hwan-Sub
    Shin, Hye-Jin
    Kim, Seol-A
    Park, Jimyeong
    Kim, Hyun-Chul
    Kim, Hyogyeong
    Kim, Hyung Joo
    Kim, Yun-Tae
    Lee, Kyoung-Ryul
    Kim, Young-Jin
    [J]. JOURNAL OF ANALYTICAL METHODS IN CHEMISTRY, 2014, 2014
  • [2] A Simple Ultraperformance Liquid Chromatography-Tandem Mass Spectrometry Method for Measurement of Cortisol Level in Human Saliva
    Alvi, Syed N.
    Hammami, Muhammad M.
    [J]. INTERNATIONAL JOURNAL OF ANALYTICAL CHEMISTRY, 2019, 2019
  • [3] Determination of loperamide in human plasma and saliva by liquid chromatography-tandem mass spectrometry
    Arafat, Tawfiq
    Arafat, Basil
    Awad, Riad
    Abu Awwad, Ahmad
    [J]. JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2014, 972 : 81 - 88
  • [4] Determination of salivary cortisol by liquid chromatography-tandem mass spectrometry
    Turpeinen, Ursula
    Valimaki, Matti Juhani
    Hamalainen, Esa
    [J]. SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 2009, 69 (05): : 592 - 597
  • [5] Simultaneous measurement of cortisol and cortisone in human saliva using liquid chromatography-tandem mass spectrometry: Application in basal and stimulated conditions
    Perogamvros, Ilias
    Owen, Laura J.
    Newell-Price, John
    Ray, David W.
    Trainer, Peter J.
    Keevil, Brian G.
    [J]. JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2009, 877 (29): : 3771 - 3775
  • [6] Determination of cyclosporine in saliva using liquid chromatography-tandem mass spectrometry
    Mendonza, A
    Gohh, R
    Akhlaghi, F
    [J]. THERAPEUTIC DRUG MONITORING, 2004, 26 (05) : 569 - 575
  • [7] Validation of a high-throughput liquid chromatography-tandem mass spectrometry method for urinary cortisol and cortisone
    Taylor, RL
    Machacek, D
    Singh, RJ
    [J]. CLINICAL CHEMISTRY, 2002, 48 (09) : 1511 - 1519
  • [8] Determination of urinary free cortisol by liquid chromatography-tandem mass spectrometry
    Turpeinen, U
    Stenman, UH
    [J]. SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 2003, 63 (02): : 143 - 150
  • [9] Determination of cortisol in human saliva using liquid chromatography-electrospray tandem mass spectrometry
    Jönsson, BAG
    Malmberg, B
    Amilon, A
    Garde, AH
    Orbæk, P
    [J]. JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2003, 784 (01): : 63 - 68
  • [10] Analysis of Perchlorate in Human Saliva by Liquid Chromatography-Tandem Mass Spectrometry
    Oldi, John F.
    Kannan, Kurunthachalam
    [J]. ENVIRONMENTAL SCIENCE & TECHNOLOGY, 2009, 43 (01) : 142 - 147