Micro Electrical Impedance Spectroscopy (mEIS) Fabricated on the Curved Surface of a Fine Needle for Biotissue Discrimination

被引:9
|
作者
Kang, Giseok [1 ]
Yun, Joho [1 ]
Cho, Jin-Seong [2 ]
Yoon, JungHan [2 ]
Lee, Jong-Hyun [1 ,3 ]
机构
[1] GIST, Dept Med Syst Engn, Gwangju 500712, South Korea
[2] Chonnam Natl Univ, Coll Med, Dept Surg, Gwangju 501757, South Korea
[3] GIST, Dept Mechatron, Gwangju 500712, South Korea
关键词
Electrical impedance spectroscopy (EIS); EIS-on-a-needle (EoN); Microfabrication; Flexible photomask; Biotissue discrimination; CELL; PARYLENE;
D O I
10.1002/elan.201500591
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This work presents a novel micro electrical impedance spectroscopy (mEIS) technique that can measure and discriminate the electrical signal responses of biotissues in real time. An EoN (EIS-on-a-needle), EIS on the surface of a fine needle (400 mm in diameter), was fabricated using a newly developed flexible photomask film. The base material of the photomask is parylene-C, which allows uniform contact on the curved surface of the needle; thus, the designed electrode patterns of the photomask can be transferred onto the needle surface with a high resolution (2.95% or less in dimensional error). To validate the developed EoN as an electrical sensor, ex vivo experiments with various biotissues-butchered pork (skin, fat, and muscle) and human breast tissues (normal and cancerous)-were conducted by measuring real-time electrical impedance during a frequency sweep. The conductivities (relative permittivity) of the pork tissues were evaluated by electrical equivalent circuit analysis: 56.6 mS/m (37,800), 68.0 mS/m (74,755), and 74.9 mS/m (26,145) for the skin, fat, and muscle, respectively. Moreover, the normal and cancerous tissues were well distinguished by electrical resistance at 4.04 kHz for various cancer grades (Elston grades 1, 2, and 3). Analysis of the electrical impedance suggests that the EoN can be utilized to diagnose the physiological states of biotissues in clinical use.
引用
收藏
页码:733 / 741
页数:9
相关论文
共 50 条
  • [11] A micro-fabricated electrical-impedance-based in-flight hematology analyzer
    Layton, Bradley E.
    Jamieson, Brian G.
    Sullivan, Stephanie M.
    ICMM 2005: 3rd International Conference on Microchannels and Minichannels, Pt B, 2005, : 579 - 583
  • [12] ELECTRICAL-IMPEDANCE SPECTROSCOPY OF SILICON SURFACE-STATES
    VISCOR, P
    VEDDE, J
    SURFACE SCIENCE, 1993, 287 : 510 - 513
  • [13] Electrochemical impedance spectroscopy of the cardiomyocyte adhesion to extracellular matrix on micro-fabricated biosensors
    Qiu, Yiling
    Liao, Ronglih
    Zhang, Xin
    TRANSDUCERS '07 & EUROSENSORS XXI, DIGEST OF TECHNICAL PAPERS, VOLS 1 AND 2, 2007,
  • [14] Ex vivo identification of thyroid cancer tissue using electrical impedance spectroscopy on a needle
    Yun, Joho
    Hong, Yong-Tae
    Hong, Ki-Hwan
    Lee, Jong-Hyun
    SENSORS AND ACTUATORS B-CHEMICAL, 2018, 261 : 550 - 557
  • [15] Flexible Micro Sensor for Intravascular Vulnerable Plaque Diagnostic with Electrical Impedance Spectroscopy
    Zhang, Xiaoxiao
    Xu, Lei
    Lee, Juhyun
    Li, Rongsong
    Liu, Yang
    Shapero, Aubrey
    Hsiai, Tzung
    Tai, Yu-Chong
    2015 IEEE 10TH INTERNATIONAL CONFERENCE ON NANO/MICRO ENGINEERED AND MOLECULAR SYSTEMS (NEMS), 2015, : 243 - 247
  • [16] A micro system using dielectrophoresis and electrical impedance spectroscopy for cell manipulation and analysis
    Mohanty, SK
    Ravula, SK
    Engisch, KL
    Frazier, AB
    BOSTON TRANSDUCERS'03: DIGEST OF TECHNICAL PAPERS, VOLS 1 AND 2, 2003, : 1055 - 1058
  • [17] Biopsy Needle Integrated with Electrical Impedance Sensing Microelectrode Array towards Real-time Needle Guidance and Tissue Discrimination
    Jaeho Park
    Won-Mook Choi
    Kyuyoung Kim
    Won-Il Jeong
    Joon-Beom Seo
    Inkyu Park
    Scientific Reports, 8
  • [18] Biopsy Needle Integrated with Electrical Impedance Sensing Microelectrode Array towards Real-time Needle Guidance and Tissue Discrimination
    Park, Jaeho
    Choi, Won-Mook
    Kim, Kyuyoung
    Jeong, Won-Il
    Seo, Joon-Beom
    Park, Inkyu
    SCIENTIFIC REPORTS, 2018, 8
  • [19] DIFFERENTIATION BETWEEN NORMAL AND CANCEROUS RENAL TISSUES USING ELECTRICAL IMPEDANCE SPECTROSCOPY NEEDLE.
    Kim, Hyeon Woo
    Yun, Joho
    Park, Yangkyu
    Shin, Dong Gil
    Lee, Jeong Zoo
    Lee, Jong-Hyun
    Lee, Wan
    Lee, Chang Yell
    JOURNAL OF UROLOGY, 2016, 195 (04): : E1169 - E1169
  • [20] Micro-fabricated High-impedance Surface for Millimeter Wave Beam Steering Applications
    Chicherin, Dmitry
    Dudorov, Sergey
    Sterner, Mikael
    Oberhammer, Joachim
    Raisanen, Antti V.
    2008 33RD INTERNATIONAL CONFERENCE ON INFRARED, MILLIMETER AND TERAHERTZ WAVES, VOLS 1 AND 2, 2008, : 574 - +