Validating and Calibrating the Nintendo Wii Balance Board to Derive Reliable Center of Pressure Measures

被引:87
|
作者
Leach, Julia M. [1 ]
Mancini, Martina [2 ]
Peterka, Robert J. [3 ]
Hayes, Tamara L. [1 ]
Horak, Fay B. [2 ]
机构
[1] OHSU, Dept Biomed Engn, Portland, OR 97239 USA
[2] OHSU, Dept Neurol, Portland, OR 97239 USA
[3] OHSU, Dept Biomed Engn, Beaverton, OR 97006 USA
来源
SENSORS | 2014年 / 14卷 / 10期
关键词
Wii balance board; force plate; validation; calibration; postural sway; balance; POSTURAL CONTROL; PARKINSONS-DISEASE; OLDER-ADULTS; FALLS; RISK; INSTABILITY; IMPAIRMENT; FREQUENCY; STABILITY; PLATFORM;
D O I
10.3390/s141018244
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The Nintendo Wii balance board (WBB) has generated significant interest in its application as a postural control measurement device in both the clinical and (basic, clinical, and rehabilitation) research domains. Although the WBB has been proposed as an alternative to the "gold standard" laboratory-grade force plate, additional research is necessary before the WBB can be considered a valid and reliable center of pressure (CoP) measurement device. In this study, we used the WBB and a laboratory-grade AMTI force plate (AFP) to simultaneously measure the CoP displacement of a controlled dynamic load, which has not been done before. A one-dimensional inverted pendulum was displaced at several different displacement angles and load heights to simulate a variety of postural sway amplitudes and frequencies (<1 Hz). Twelve WBBs were tested to address the issue of inter-device variability. There was a significant effect of sway amplitude, frequency, and direction on the WBB's CoP measurement error, with an increase in error as both sway amplitude and frequency increased and a significantly greater error in the mediolateral (ML) (compared to the anteroposterior (AP)) sway direction. There was no difference in error across the 12 WBB's, supporting low inter-device variability. A linear calibration procedure was then implemented to correct the WBB's CoP signals and reduce measurement error. There was a significant effect of calibration on the WBB's CoP signal accuracy, with a significant reduction in CoP measurement error (quantified by root-mean-squared error) from 2-6 mm (before calibration) to 0.5-2 mm (after calibration). WBB-based CoP signal calibration also significantly reduced the percent error in derived (time-domain) CoP sway measures, from -10.5% (before calibration) to -0.05% (after calibration) (percent errors averaged across all sway measures and in both sway directions). In this study, we characterized the WBB's CoP measurement error under controlled, dynamic conditions and implemented a linear calibration procedure for WBB CoP signals that is recommended to reduce CoP measurement error and provide more reliable estimates of time-domain CoP measures. Despite our promising results, additional work is necessary to understand how our findings translate to the clinical and rehabilitation research domains. Once the WBB's CoP measurement error is fully characterized in human postural sway (which differs from our simulated postural sway in both amplitude and frequency content), it may be used to measure CoP displacement in situations where lower accuracy and precision is acceptable.
引用
收藏
页码:18244 / 18267
页数:24
相关论文
共 50 条
  • [1] Reliability and Sensitivity Analysis on the Center of Pressure Measures in Healthy Young Adults Using Nintendo Wii Balance Board
    Qu, Xiaoting
    Nan, Wenya
    Wong, Chi Man
    Wan, Feng
    Hu, Yong
    [J]. 2015 IEEE INTERNATIONAL CONFERENCE ON COMPUTATIONAL INTELLIGENCE AND VIRTUAL ENVIRONMENTS FOR MEASUREMENT SYSTEMS AND APPLICATIONS (CIVEMSA), 2015, : 1 - 4
  • [2] Accuracy of force and center of pressure measures of the Wii Balance Board
    Bartlett, Harrison L.
    Ting, Lena H.
    Bingham, Jeffrey T.
    [J]. GAIT & POSTURE, 2014, 39 (01) : 224 - 228
  • [3] Nintendo Wii Balance Board for Balance Disorders
    Gil-Gomez, Jose-Antonio
    Lozano, Jose-Antonio
    Alcaniz, Mariano
    Albiol Perez, Sergio
    [J]. 2009 VIRTUAL REHABILITATION INTERNATIONAL CONFERENCE, 2009, : 213 - 213
  • [4] Can the Nintendo WII Balance Board be Used for a Reliable Assessment of the Initiation of Gait?
    Verdini, Federica
    Tigrini, Andrea
    Fioretti, Sandro
    Mengarelli, Alessandro
    [J]. IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2022, 71
  • [5] Evaluating center of pressure (CoP) measures of postural stability in low and normal vision using the Nintendo Wii Balance Board™ (WBB)
    Jeter, Pamela
    Gu, Jialiang
    Goldstein, Judith
    Bittner, Ava
    Dagnelie, Gislin
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2013, 54 (15)
  • [6] Validity of the Nintendo Wii Balance Board for the Assessment of Balance Measures in the Functional Reach Test
    Mengarelli, Alessandro
    Cardarelli, Stefano
    Strazza, Annachiara
    Di Nardo, Francesco
    Fioretti, Sandro
    Verdini, Federica
    [J]. IEEE TRANSACTIONS ON NEURAL SYSTEMS AND REHABILITATION ENGINEERING, 2018, 26 (07) : 1400 - 1406
  • [7] Balance Evaluation in Hemophilic Patients by Nintendo Wii Balance Board
    Perez-Alenda, S.
    Carrasco, J. J.
    Casana-Granell, J.
    Alakhdar-Mohamara, Y.
    Cuesta-Barriuso, R.
    Torres-Ortuno, A.
    Cruz-Montecinos, C.
    Querol, F.
    [J]. HAEMOPHILIA, 2015, 21 (06) : E548 - E549
  • [8] Validation of the Nintendo Wii Balance Board for stabilometry measurements
    Nagymate, Gergely
    Vamos, Bence
    Kiss, Rita M.
    [J]. 2016 INTERNATIONAL SYMPOSIUM ON SMALL-SCALE INTELLIGENT MANUFACTURING SYSTEMS (SIMS), 2016, : 111 - 114
  • [9] Isometric hand grip strength measured by the Nintendo Wii Balance Board – a reliable new method
    A. W. Blomkvist
    S. Andersen
    E. D. de Bruin
    M. G. Jorgensen
    [J]. BMC Musculoskeletal Disorders, 17
  • [10] Balance Trainability Using the Nintendo Wii Balance Board in Sportive People
    Paukowits, S.
    Stoeggl, T.