Geometric Calibration Using Line Fiducials for Cone-Beam CT with General, Non-Circular Source-Detector Trajectories

被引:4
|
作者
Jacobson, M. W. [1 ]
Ketcha, M. [1 ]
Uneri, A. [1 ]
Goerres, J. [1 ]
De Silva, T. [1 ]
Reaungamornrat, S. [2 ]
Vogt, S. [3 ]
Kleinszig, G. [3 ]
Siewerdsen, J. H. [1 ]
机构
[1] Johns Hopkins Univ, Dept Biomed Engn, Baltimore, MD 21205 USA
[2] Johns Hopkins Univ, Dept Comp Sci, Baltimore, MD 21218 USA
[3] Siemens Healthineers XP Div, Erlangen, Germany
关键词
D O I
10.1117/12.2255724
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
Purpose: Traditional BB-based geometric calibration methods for cone-beam CT (CBCT) rely strongly on foreknowledge of the scan trajectory shape. This is a hindrance to the implementation of variable trajectory CBCT systems, normally requiring a dedicated calibration phantom or software algorithm for every scan orbit of interest. A more flexible method of calibration is proposed here that accommodates multiple orbit types-including strongly noncircular trajectories-with a single phantom and software routine. Methods: The proposed method uses a calibration phantom consisting of multiple line-shaped wire segments. Geometric models relating the 3D line equations of the wires to the 2D line equations of their projections are used as the basis for system geometry estimation. This method was tested using a mobile C-arm CT system and comparisons were made to standard BB-based calibrations. Simulation studies were also conducted using a sinusoid-on-sphere orbit. Calibration performance was quantified in terms of Point Spread Function (PSF) width and back projection error. Visual image quality was assessed with respect to spatial resolution in trabecular bone in an anthropomorphic head phantom. Results: The wire-based calibration method performed equal to or better than BB-based calibrations in all evaluated metrics. For the sinusoidal scans, the method provided reliable calibration, validating its application to non-circular trajectories. Furthermore, the ability to improve image quality using non-circular orbits in conjunction with this calibration method was demonstrated. Conclusion: The proposed method has been shown feasible for conventional circular CBCT scans and offers a promising tool for non-circular scan orbits that can improve image quality, reduce dose, and extend field of view.
引用
收藏
页数:7
相关论文
共 50 条
  • [1] Fully automatic online geometric calibration for non-circular cone-beam CT orbits using fiducials with unknown placement
    Ma, Yiqun Q.
    Reynolds, Tess
    Ehtiati, Tina
    Weiss, Clifford
    Hong, Kelvin
    Theodore, Nicholas
    Gang, Grace J.
    Stayman, J. Webster
    [J]. MEDICAL PHYSICS, 2024, 51 (05) : 3245 - 3264
  • [2] An analytic geometric calibration method for circular cone-beam CT
    Xu, Jingyan
    Tsui, Benjamin
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2012, 53
  • [3] A line fiducial method for geometric calibration of cone-beam CT systems with diverse scan trajectories
    Jacobson, M. W.
    Ketcha, M. D.
    Capostagno, S.
    Martin, A.
    Uneri, A.
    Goerres, J.
    De Silva, T.
    Reaungamornrat, S.
    Han, R.
    Manbachi, A.
    Stayman, J. W.
    Vogt, S.
    Kleinszig, G.
    Siewerdsen, J. H.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (02):
  • [4] Geometric Calibration of Cone-beam CT with a Flat-panel Detector
    Wu, Dufan
    Li, Liang
    Zhang, Li
    Xing, Yuxiang
    Chen, Zhiqiang
    Xiao, Yongshun
    [J]. 2011 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2011, : 2952 - 2955
  • [5] A consistency condition for cone-beam CT with general source trajectories
    Sidky, EY
    Zou, Y
    Xia, D
    Pan, XC
    [J]. Medical Imaging 2005: Image Processing, Pt 1-3, 2005, 5747 : 2042 - 2047
  • [6] Non-circular Cone Beam CT Trajectories: A Preliminary Investigation on a Clinical Scanner
    Pearson, Erik A.
    Cho, Seungryong
    Pelizzari, Charles A.
    Pan, Xiaochuan
    [J]. 2010 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD (NSS/MIC), 2010, : 3172 - 3175
  • [7] A paddle-wheel phantom for geometric calibration of cone-beam CT with offset detector
    da Silva, Joakim
    Chen, Maxin
    Eriksson, Markus
    [J]. MEDICAL IMAGING 2022: PHYSICS OF MEDICAL IMAGING, 2022, 12031
  • [8] Cone-beam CT sampling incompleteness: analytical and empirical studies of emerging systems and source-detector orbits
    Wu, Pengwei
    Tersol, Aina
    Clackdoyle, Rolf
    Boone, John M.
    Siewerdsen, Jeffrey H.
    [J]. JOURNAL OF MEDICAL IMAGING, 2023, 10 (03)
  • [9] Automated Geometric Calibration and Reconstruction in Circular Cone-Beam Tomography
    Mennessier, C.
    Clackdoyle, R.
    [J]. 2008 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (2008 NSS/MIC), VOLS 1-9, 2009, : 4347 - 4351
  • [10] Iterative geometric calibration in circular cone-beam computed tomography
    Zhang, Feng
    Du, Jianping
    Jiang, Hua
    Li, Lei
    Guan, Min
    Yan, Bin
    [J]. OPTIK, 2014, 125 (11): : 2509 - 2514