Accuracy of dual-source computed tomography coronary angiography: evaluation with a standardised protocol for cardiac surgeons

被引:9
|
作者
Plass, Andre [1 ]
Azemaj, Naim [1 ]
Scheffel, Hans [2 ]
Desbiolles, Lotus [2 ]
Alkadhi, Hatem [2 ]
Genoni, Michele [1 ]
Falk, Volkmar [1 ]
Grunenfelder, Jurg [1 ]
机构
[1] Univ Zurich Hosp, Cardiovasc Surg Clin, CH-8091 Zurich, Switzerland
[2] Univ Zurich Hosp, Inst Diagnost Radiol, CH-8091 Zurich, Switzerland
关键词
Coronary artery imaging; Dual-source computed tomography; Cardiac surgery; MULTISLICE SPIRAL CT; ARTERY-DISEASE; TEMPORAL RESOLUTION; DIAGNOSTIC-ACCURACY; 1ST EXPERIENCE; HEART; RECONSTRUCTION; QUANTIFICATION; CARDIOLOGY;
D O I
10.1016/j.ejcts.2009.06.027
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: This study assesses the accuracy of the new dual-source computed tomography (DSCT) for detection of coronary artery disease (CAD) compared with invasive coronary angiography (ICA) with a specifically designed data presentation protocol for cardiac surgeons. Methods: Forty patients (30 males/10 females) underwent ICA and DSCT. Best-quality images were prepared by radiologists. Evaluation of 12 segments of significant coronary stenosis was done by two cardiac surgeons with a data presentation protocol including different coronary views in two-/three-dimensional (2D/3D) images. No beta-blockers were administered prior to DSCT. Results: ICA revealed CAD in 21 patients and valvular disease but no CAD in 19 patients. In DSCT, 20/21 patients were diagnosed with CAD (at least one significant stenosis per patient). In 11/21 patients, all 12 segments were assessed correctly; in 7/21 patients one segment and in 3/21 patients two segments were evaluated incorrectly. Of all 21 patients with CAD, 239/252 segments (95%) were correctly evaluated. In 18/19 patients without CAD, DSCT correctly ruled-out the ICA results in 226/228 segments (99%). In total, 465/480 segments were correctly assessed (97%). Of 480 segments, only six were considered not assessable. DSCT assessments of the segments showed a sensitivity of 91%, specificity of 99%, a positive predictive value of 92% and a negative predictive value of 99%. Conclusions: The accuracy of DSCT coronary angiography especially for exclusion of CAD is promising. The introduced data presentation protocol allows for the independent evaluation by cardiac surgeons after pre-arrangement from the radiologists. (C) 2009 European Association for Cardio-Thoracic Surgery. Published by Elsevier B.V. All rights reserved.
引用
收藏
页码:1011 / 1017
页数:7
相关论文
共 50 条
  • [1] Accuracy of dual-source computed tomography coronary angiography: evaluation with a standardised protocol for cardiac surgeons
    Bedi, Harinder Singh
    EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2011, 39 (02) : 282 - 282
  • [2] Accuracy of dual-source computed tomography coronary angiography: evaluation with a standardised protocol for cardiac surgeons Reply
    Plass, Andre
    EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2011, 39 (02) : 283 - 283
  • [3] Safety and accuracy of dual-source coronary computed tomography angiography in the pediatric population
    Han, B. Kelly
    Lindberg, Jana
    Overman, David
    Schwartz, Robert S.
    Grant, Katharine
    Lesser, John R.
    JOURNAL OF CARDIOVASCULAR COMPUTED TOMOGRAPHY, 2012, 6 (04) : 252 - 259
  • [4] Influence of heart rate on the diagnostic accuracy of dual-source computed tomography coronary angiography
    Ropers, Ulrike
    Ropers, Dieter
    Pflederer, Tobias
    Anders, Katharina
    Kuettner, Axel
    Stilianakis, Nikolaos I.
    Komatsu, Sei
    Kalender, Willi
    Bautz, Werner
    Daniel, Werner G.
    Achenbach, Stephan
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2007, 50 (25) : 2393 - 2398
  • [5] Diagnostic accuracy of dual-source coronary computed tomography angiography in patients after bypass grafting
    Kepka, Cezary
    Opolski, Maksymilian
    Kruk, Mariusz
    Pregowski, Jerzy
    Kowalewska, Anna
    Wilkowska, Paulina
    Dzielinska, Zofia
    Karcz, Maciej
    Ruzyllo, Witold
    Demkow, Marcin
    POSTEPY W KARDIOLOGII INTERWENCYJNEJ, 2012, 8 (03): : 183 - 192
  • [6] Radiation dose estimates in dual-source computed tomography coronary angiography
    Stolzmann, Paul
    Scheffel, Hans
    Schertler, Thomas
    Frauenfelder, Thomas
    Leschka, Sebastian
    Husmann, Lars
    Flohr, Thomas G.
    Marincek, Borut
    Kaufmann, Philipp A.
    Alkadhi, Hatem
    EUROPEAN RADIOLOGY, 2008, 18 (03) : 592 - 599
  • [7] Radiation dose estimates in dual-source computed tomography coronary angiography
    Paul Stolzmann
    Hans Scheffel
    Thomas Schertler
    Thomas Frauenfelder
    Sebastian Leschka
    Lars Husmann
    Thomas G. Flohr
    Borut Marincek
    Philipp A. Kaufmann
    Hatem Alkadhi
    European Radiology, 2008, 18 : 592 - 599
  • [8] Coronary artery anomalies, detection by dual-source computed tomography angiography
    Vliegen, H. W.
    NETHERLANDS HEART JOURNAL, 2010, 18 (10): : 464 - 465
  • [9] Coronary artery anomalies, detection by dual-source computed tomography angiography
    H. W. Vliegen
    Netherlands Heart Journal, 2010, 18 : 464 - 465
  • [10] Diagnostic accuracy of dual-source coronary computed tomography angiography in patients after bypass grafting: comparison with invasive coronary angiography
    Kepka, C.
    Opolski, M. P.
    Pregowski, J.
    Kruk, M.
    Kowalewska, A.
    Wilkonska, P.
    Witkowski, A.
    Ruzyllo, W.
    Demkow, M.
    EUROPEAN HEART JOURNAL, 2012, 33 : 411 - 411