An in vitro verification of strength estimation for moving an 125I source during implantation in brachytherapy

被引:3
|
作者
Tanaka, Kenichi [1 ]
Kajimoto, Tsuyoshi [1 ]
Hayashi, Takahiro [1 ]
Asanuma, Osamu [2 ]
Hori, Masakazu [3 ]
Kamo, Ken-ichi [3 ]
Sumida, Iori [4 ]
Takahashi, Yutaka [4 ]
Tateoka, Kunihiko [3 ]
Bengua, Gerard [5 ]
Sakata, Koh-ichi [3 ]
Endo, Satoru [1 ]
机构
[1] Hiroshima Univ, Grad Sch Engn, Quantum Energy Applicat, 1-4-1 Kagamiyama, Higashihiroshima, Hiroshima 7398527, Japan
[2] Sapporo Med Univ Hosp, Div Radiol & Nucl Med, Chuo Ward, South 1,West 17, Sapporo, Hokkaido 0608556, Japan
[3] Sapporo Med Univ, Grad Sch Med, Dept Radiat Oncol, Chuo Ward, South 1,West 16, Sapporo, Hokkaido 0608543, Japan
[4] Osaka Univ, Grad Sch Med, Dept Radiol, 2-2 Yamada Oka, Suita, Osaka 5650871, Japan
[5] Auckland City Hosp, Pk Rd, Auckland 1142, New Zealand
基金
日本学术振兴会; 日本科学技术振兴机构;
关键词
brachytherapy; I-125; source strength; moving source; verification; INTERSTITIAL BRACHYTHERAPY; DOSIMETRY;
D O I
10.1093/jrr/rry021
中图分类号
Q [生物科学];
学科分类号
07 ; 0710 ; 09 ;
摘要
This study aims to demonstrate the feasibility of a method for estimating the strength of a moving brachytherapy source during implantation in a patient. Experiments were performed under the same conditions as in the actual treatment, except for one point that the source was not implanted into a patient. The brachytherapy source selected for this study was I-125 with an air kerma strength of 0.332 U (mu Gym(2)h(-1)), and the detector used was a plastic scintillator with dimensions of 10 cm x 5 cm x 5 cm. A calibration factor to convert the counting rate of the detector to the source strength was measured and then the accuracy of the proposed method was investigated for a manually driven source. The accuracy was found to be under 10% when the shielding effect of additional needles for implantation at other positions was corrected, and about 30% when the shielding was not corrected. Even without shielding correction, the proposed method can detect dead/dropped source, implantation of a source with the wrong strength, and a mistake in the number of the sources implanted. Furthermore, when the correction was applied, the achieved accuracy came close to within 7% required to find the Oncoseed 6711 (I-125 seed with unintended strength among the commercially supplied values of 0.392, 0.462 and 0.533 U).
引用
收藏
页码:484 / 489
页数:6
相关论文
共 50 条
  • [21] Minimizing the number of implantation needles for prostate 125I brachytherapy: An investigation of possibilities and implications
    Steggerda, Marcel J.
    van der Poel, Henk G.
    Moonen, Luc M. F.
    BRACHYTHERAPY, 2010, 9 (04) : 319 - 327
  • [22] Corrections for photons attenuation and scattering for the extrapolation chamber for measurement of air-kerma strength of 125I brachytherapy source
    Liu, C.
    Song, M.
    Liu, Y.
    Wang, H.
    Wei, K.
    Hang, Z.
    JOURNAL OF INSTRUMENTATION, 2021, 16 (03)
  • [23] Monte Carlo Simulation of Dose Calculation Parameter of Radioactive 125I Brachytherapy Source
    Liu C.
    Song M.
    Wang H.
    Ni N.
    Wei K.
    Liu Y.
    Yuanzineng Kexue Jishu/Atomic Energy Science and Technology, 2020, 54 (11): : 2217 - 2222
  • [24] Anisotropy Function Verification of I-125 Brachytherapy Source with Film Dosimetry
    Zhang, D.
    Gupta, N.
    Lu, L.
    Zhang, H.
    MEDICAL PHYSICS, 2010, 37 (06)
  • [25] Evaluation of the visibility of a new thinner 125I radioactive source for permanent prostate brachytherapy
    Roberts, Gemma
    Al-Qaisieh, Bashar
    Bownes, Peter
    BRACHYTHERAPY, 2012, 11 (06) : 460 - 467
  • [26] Preliminary Research of Tissue Heterogeneity Correction for Dose Distribution of 125I Brachytherapy Source
    Sun, Liang
    Yu, Hongliang
    JOURNAL OF NUCLEAR SCIENCE AND TECHNOLOGY, 2009, 46 (04) : 331 - 338
  • [27] Comparison of methods for preparation of 125I brachytherapy source cores for the treatment of eye cancer
    Manolkar, RB
    Sane, SU
    Pillai, KT
    Majali, MA
    APPLIED RADIATION AND ISOTOPES, 2003, 59 (2-3) : 145 - 150
  • [28] Dose distribution along the transverse axis of a new 125I source for interstitial brachytherapy
    Nath, R
    Yue, N
    MEDICAL PHYSICS, 2000, 27 (11) : 2536 - 2540
  • [29] Experimental and theoretical determination of dosimetric characteristics of IsoAid ADVANTAGE™ 125I brachytherapy source
    Meigooni, AS
    Hayes, JL
    Zhang, HL
    Sowards, K
    MEDICAL PHYSICS, 2002, 29 (09) : 2152 - 2158
  • [30] Chinese expert consensus on radioactive 125I seeds interstitial implantation brachytherapy for pancreatic cancer
    Gai, Baodong
    Zhang, Fujun
    JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2018, 14 (07) : 1455 - 1462