MRI-Guided Breast Biopsy: Influence of Choice of Vacuum Biopsy System on the Mode of Biopsy of MRI-Only Suspicious Breast Lesions

被引:25
|
作者
Schrading, Simone [1 ]
Simon, Birgit [1 ]
Braun, Michael [2 ]
Wardelmann, Eva [3 ]
Schild, Hans H. [1 ]
Kuhl, Christiane K. [1 ]
机构
[1] Univ Bonn, Dept Radiol, D-53105 Bonn, Germany
[2] Univ Bonn, Dept Gynecol, D-53105 Bonn, Germany
[3] Univ Bonn, Dept Pathol, D-53105 Bonn, Germany
关键词
breast cancer; breast imaging; MRI; MRI-guided intervention; vacuum biopsy; CLINICAL-EXPERIENCE; DEVICE; NEEDLE; LOCALIZATION;
D O I
10.2214/AJR.09.2550
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
OBJECTIVE. The purpose of this study was to evaluate two systems of MRI-guided vacuum-assisted biopsy and to investigate the influence of the choice of system on the choice of biopsy mode in the care of patients with lesions found only at MRI. MATERIALS AND METHODS. Over a period of 3 years, a total 349 patients underwent MRI-guided tissue sampling of 475 lesions found only at MRI. The lesions were sampled by needle localization plus excisional biopsy or by vacuum-assisted biopsy. Two different systems were used for MRI-guided vacuum-assisted biopsy. During the first half of the study period, a handheld system was used, and during the second half, a console system was used. The procedural advantages and disadvantages, size of lesions biopsied, and time needed for vacuum-assisted biopsy were recorded. The distribution of the type of intervention (needle localization or vacuum-assisted biopsy) used to manage MRI-only lesions during the two study periods also was assessed. RESULTS. The average diameter of lesions sampled with vacuum-assisted biopsy was 19.2 mm with the handheld system and 10.4 mm with the console system (p < 0.039). The average biopsy time was 69 minutes for the handheld system and 39 minutes for the console system (p < 0.005). Of the total of 170 MRI-only lesions biopsied with the handheld system, 121 (71%) were sampled by localization and 49 (29%) by vacuum-assisted biopsy. Of the total 305 MRI-only lesions biopsied with the console system, 38 (12%) were sampled by localization and 267 (88%) by vacuum-assisted biopsy. CONCLUSION. Because of the procedural advantages of use of the console-based system, smaller lesions were biopsied in less time and with higher operator confidence. This result translated into a major shift in the care of patients with MRI-only lesions away from lesion localization toward increased use of MRI-guided vacuum-assisted biopsy.
引用
收藏
页码:1650 / 1657
页数:8
相关论文
共 50 条
  • [21] MRI-Guided Breast Biopsy: A Series Review with Histologic Correlation
    Courtney, Michael
    O'Keeffe, Sylvia
    Maguire, Aoife
    Doyle, Aoife
    BREAST CANCER RESEARCH, 2019, 21
  • [22] MRI-Guided Breast Biopsy: Outcomes and Effect on Patient Management
    Myers, Kelly S.
    Kamel, Ihab R.
    Macura, Katarzyna J.
    CLINICAL BREAST CANCER, 2015, 15 (02) : 143 - 152
  • [23] Lesion Nonvisualization at MRI-Guided Breast Biopsy: Now What?
    Neal, Colleen H.
    ACADEMIC RADIOLOGY, 2018, 25 (09) : 1099 - 1100
  • [24] Complete Excision of the MRI Target Lesion at MRI-Guided Vacuum-Assisted Biopsy of Breast Cancer
    Lee, Jung-Min
    Kaplan, Jennifer B.
    Murray, Melissa P.
    Liberman, Laura
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2008, 191 (04) : 1198 - 1202
  • [25] The utility of interval MRI after benign and concordant MRI-guided breast biopsy
    Sasaki, Jennifer
    Kowzun, Maria
    Teng, Bo
    Potdevin, Lindsay
    Eladoumikdachi, Firas
    Kumar, Shicha
    ANNALS OF SURGICAL ONCOLOGY, 2018, 25 : 511 - 511
  • [26] Imaging-Histologic Discordance at MRI-Guided Vacuum-Assisted Breast Biopsy
    Lee, J.
    Murray, M. P.
    Liberman, L.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2007, 188 (05)
  • [27] Contrast- enhanced spectral mammography (CESM)-guided breast biopsy as an alternative to MRI-guided biopsy
    James, Jonathan
    BRITISH JOURNAL OF RADIOLOGY, 2022, 95 (1132):
  • [28] Optimal Design of Bionic Flexible Fixation System for MRI-Guided Breast Biopsy
    Zhang, Tianxue
    Liu, Yun-hui
    JOURNAL OF BIONIC ENGINEERING, 2019, 16 (06) : 1116 - 1126
  • [29] Optimal Design of Bionic Flexible Fixation System for MRI-Guided Breast Biopsy
    Tianxue Zhang
    Yun-hui Liu
    Journal of Bionic Engineering, 2019, 16 : 1116 - 1126
  • [30] High-Risk Lesions at MRI-Guided Breast Biopsy: Frequency and Rate of Underestimation
    Lourenca, Ana P.
    Khalil, Henan
    Sanford, Matthew
    Donegan, Linda
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2014, 203 (03) : 682 - 686