Simultaneous spectrophotometric determination of iron, titanium, and aluminum by partial least-squares calibration method in micellar medium

被引:17
|
作者
Safavi, A [1 ]
Mirzaee, M [1 ]
Abdollahi, H [1 ]
机构
[1] Shiraz Univ, Dept Chem, Fac Sci, Shiraz 71454, Iran
关键词
simultaneous analysis; PLS method; iron; titanium; aluminum;
D O I
10.1081/AL-120018258
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Simultaneous spectrophotometric determination of iron, titanium and aluminum was carried out using their morin (3,5,7,2',4'-penta hydroxy flavone) complexes in Triton X-100 micellar, medium. Partial least-squares regression was employed for the resolution of ternary mixtures of Fe, Ti, and Al. The presence of Triton X-100 micellar system avoids the solvent extraction step necessary for the formation of soluble morin complexes. The limits of detection obtained for Fe(III), Ti (IV), and for Al (III) were 0.03, 0.04, and 0.02 mug mL(-1), respectively. The relative standard error for applying the method to the analysis of 20 synthetic samples in the concentration ranges of 0.03 to 1.0 mug mL(-1) Al(III), 0.06 to 1.6mug mL(-1) Ti(IV), and 0.04 to 1.5 mug mL(-1) Fe(III) was 5%. Analysis of synthetic and artificial mixtures of solutions of several alloys was successfully performed by the method.
引用
收藏
页码:699 / 712
页数:14
相关论文
共 50 条