A Monte Carlo dosimetry study using Henschke applicator for cervical brachytherapy

被引:12
|
作者
Yu, Pei-Chieh [2 ,4 ]
Chao, Tsi-Chian [1 ]
Lee, Chung-Chi [1 ,3 ]
Wu, Ching-Jung [4 ]
Tung, Chuan-Jong [1 ]
机构
[1] Chang Gung Univ, Dept Med Imaging & Radiol Sci, Tao Yuan 333, Taiwan
[2] Natl Tsing Hua Univ, Dept Biomed Engn & Environm Sci, Hsinchu 30013, Taiwan
[3] Chang Gang Mem Hosp, Dept Radiat Oncol, Tao Yuan 333, Taiwan
[4] Cathay Gen Hosp, Dept Radiat Oncol, Taipei 106, Taiwan
关键词
Brachytherapy; Henschke applicator; MCNP; Mesh tally;
D O I
10.1016/j.nima.2010.01.079
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
In recent years the Henschke applicator has been widely used for gynecologic patients treated by brachytherapy in Taiwan. However, the commercial brachytherapy planning system did not properly evaluate the dose perturbation caused by the Henschke applicator. Since the European Society for Therapeutic Radiology and Oncology advised that the effect of source shielding should be incorporated into the brachytherapy planning system, it required calculation and comparison of the dose distribution around the applicator. This study used the Monte Carlo MCNP code to simulate the dose distribution in a water phantom that contained the Henschke applicator with one tandem and two ovoids. Three dwell positions of a high dose rate Ir-192 source were simulated by including and excluding the applicator. The mesh tally option of the MCNP was applied to facilitate the calculation of a large number of tallies in the phantom. The voxel size effect and the charge particle equilibrium were studied by comparing the results calculated with different tally options. The calculated results showed that the brachytherapy planning system overestimated the rectal dose and that the shielding material in the applicator contributed more than 40% to the rectal dose. (C) 2010 Elsevier B.V. All rights reserved.
引用
收藏
页码:411 / 414
页数:4
相关论文
共 50 条
  • [1] A Monte Carlo Phantom Study for Brachytherapy Dose Perturbation in Henschke Type GYN Applicator
    Yu, P.
    Chao, T.
    Lee, C.
    Tung, C.
    MEDICAL PHYSICS, 2009, 36 (06)
  • [2] Comparative Dosimetry of the Xoft Cervical Applicator and HDR Ir-192 Henschke Applicator
    Holt, R.
    Kelley, L.
    Rusch, T.
    Hausen, H.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [3] Dose assessment for brachytherapy with Henschke applicator
    Yu, Pei-Chieh
    Chao, Tsi-Chian
    Tung, Chuan-Jong
    Wu, Ching-Jung
    Lee, Chung-Chi
    RADIATION MEASUREMENTS, 2011, 46 (12) : 2028 - 2030
  • [4] Experimental and Monte Carlo dosimetry of the Henschke applicator for high dose-rate 192Ir remote afterloading
    Watanabe, Y
    Roy, J
    Harrington, PJ
    Anderson, LL
    MEDICAL PHYSICS, 1998, 25 (05) : 736 - 745
  • [5] Dosimetry and Error Analysis of HDR Brachytherapy Tandem Applicator Using A16 Microchamber and Monte Carlo Simulation
    Amoush, A.
    MEDICAL PHYSICS, 2012, 39 (06) : 3735 - 3735
  • [6] A Monte Carlo dosimetry study of vaginal 192Ir brachytherapy applications with a shielded cylindrical applicator set
    Lymperopoulou, G
    Pantelis, E
    Papagiannis, P
    Rozaki-Mavrouli, H
    Sakelliou, L
    Baltas, D
    Karaiskos, P
    MEDICAL PHYSICS, 2004, 31 (11) : 3080 - 3086
  • [7] Monte Carlo dosimetry for intravascular brachytherapy
    Li, XA
    Wang, RQ
    Yu, C
    Weems, L
    Suntharalingam, M
    USE OF COMPUTERS IN RADIATION THERAPY, 2000, : 489 - 491
  • [8] Monte Carlo dosimetry in HDR brachytherapy
    Martínez-Dávalos, A
    MEDICAL PHYSICS, 2000, 538 : 128 - 133
  • [9] MONTE CARLO APPLICATION IN BRACHYTHERAPY DOSIMETRY
    Perez-Calatayud, Jose
    Granero Cabanero, Domingo
    Ballester Pallares, Facundo
    RADIOTHERAPY AND BRACHYTHERAPY, 2009, : 239 - +
  • [10] Design Optimization of a HDR Ir-192 Brachytherapy Skin Applicator Using Monte Carlo-Based Dosimetry
    Yang, Y.
    Rivard, M. J.
    MEDICAL PHYSICS, 2012, 39 (06) : 3612 - 3612