Non-Invasive Cardiac and Respiratory Activity Assessment From Various Human Body Locations Using Bioimpedance

被引:0
|
作者
Sel, Kaan [1 ]
Osman, Deen [1 ]
Jafari, Roozbeh [1 ]
机构
[1] Texas A&M Univ, College Stn, TX 77843 USA
来源
IEEE OPEN JOURNAL OF ENGINEERING IN MEDICINE AND BIOLOGY | 2021年 / 2卷
关键词
Bioimpedance; Blood; Sensors; Carotid arteries; Biomedical monitoring; Electrodes; Morphology; hemodynamics; impedance cardiography; non-invasive; respiration; BLOOD-FLOW; IMPEDANCE;
D O I
10.1109/OJEMB.2021.3085482
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Objective: Bioimpedance sensing is a powerful technique that measures the tissue impedance and captures important physiological parameters including blood flow, lung movements, muscle contractions, body fluid shifts, and other cardiovascular parameters. This paper presents a comprehensive analysis of the modality at different arterial (ulnar, radial, tibial, and carotid arteries) and thoracic (side-rib cage and top thoracolumbar fascia) body regions and offers insights into the effectiveness of capturing various cardiac and respiratory activities. Methods: We assess the bioimpedance performance in estimating inter-beat (IBI) and inter -breath intervals (IBrI) on six-hours of data acquired in a pilot-study from five healthy participants at rest. Results: Overall, we achieve mean errors as low as 0.003 +/- 0.002 and 0.67 +/- 0.28 seconds for IBI and IBrI estimations, respectively. Conclusions: The results show that bioimpedance can be effectively used to monitor cardiac and respiratory activities both at limbs and upper body and demonstrate a strong potential to be adopted by wearables that aim to provide high-fidelity physiological sensing to address precision medicine needs.
引用
收藏
页码:210 / 217
页数:8
相关论文
共 50 条
  • [1] NON-INVASIVE CARDIAC-OUTPUT MEASUREMENT BY BIOIMPEDANCE - IS IT RELIABLE
    JEWKES, C
    VERHOEFF, F
    SEAR, JW
    FOEX, P
    BRITISH JOURNAL OF ANAESTHESIA, 1989, 63 (05) : P619 - P619
  • [2] Myths and methodologies: Invasive and non-invasive assessment of respiratory muscle activity in humans
    Hudson, Anna L.
    Jolley, Caroline J.
    Gandevia, Simon C.
    Butler, Jane E.
    EXPERIMENTAL PHYSIOLOGY, 2025,
  • [3] ASSESSMENT OF RESPIRATORY DEPRESSION USING A NEW NON-INVASIVE TECHNIQUE
    HERSH, E
    DESJARDINS, P
    SIMPSER, M
    DADZIE, C
    CHUDY, D
    JOURNAL OF DENTAL RESEARCH, 1983, 62 : 214 - 214
  • [4] Non-invasive assessment of cardiac haemodynamics
    Vanjani, C. V.
    INTERNATIONAL JOURNAL OF CARDIOLOGY, 2007, 122 : 20 - 20
  • [5] A new non-invasive bioimpedance system to monitor cardiac function continuously
    Heethaar, RM
    Goovaerts, HG
    De Waal, EEC
    Hoetink, AE
    Marcus, TJT
    Faes, TJC
    INTENSIVE CARE MEDICINE, 2002, 28 : S175 - S175
  • [6] Non-invasive techniques for respiratory assessment in horses
    不详
    VETERINARY RECORD, 2010, 166 (12) : 348 - 348
  • [7] Non-invasive assessment of human baroreflex during different body positions
    Iida, R
    Hirayanagi, K
    Iwasaki, K
    Ogawa, S
    Suzuki, H
    Yajima, K
    JOURNAL OF THE AUTONOMIC NERVOUS SYSTEM, 1999, 75 (2-3): : 164 - 170
  • [8] Thoracic electrical bioimpedance: a non-invasive measure of cardiac output for porcine research
    Broomhead, CJ
    Wright, SJ
    Kiff, KM
    Withington, PS
    LABORATORY ANIMALS, 1998, 32 (03) : 324 - 329
  • [9] Non-invasive assessment of cardiac output in children
    Richardson, JR
    Ferguson, J
    Hiscox, J
    Rawles, J
    JOURNAL OF ACCIDENT & EMERGENCY MEDICINE, 1998, 15 (05): : 304 - 307
  • [10] Non-invasive cardiac output assessment in children
    Childs, C
    Dark, P
    JOURNAL OF ACCIDENT & EMERGENCY MEDICINE, 1999, 16 (02): : 158 - 158