Is the use of an additional pace/sense lead the optimal strategy for the avoidance of lead extraction in defibrillation lead failure? A single-centre experience

被引:7
|
作者
Scott, Paul A. [1 ]
Chungh, Aman [1 ]
Zeb, Mehmood [1 ]
Yue, Arthur M. [1 ]
Roberts, Paul R. [1 ]
Morgan, John M. [1 ]
机构
[1] Univ Southampton, Hosp NHS Trust, Wessex Cardiothorac Unit, Southampton SO16 6YD, Hants, England
来源
EUROPACE | 2010年 / 12卷 / 04期
关键词
Implantable cardioverter defibrillators; Lead; Failure; Complications; Defibrillation; PERFORMANCE; MANAGEMENT; DEFECTS;
D O I
10.1093/europace/eup406
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Aims The implantation of an additional pace-sense (P/S) lead is a standard treatment option in the management of an isolated pace-sense problem in a defibrillation (HV-P/S) lead. However, the safety of this management strategy is unclear. We performed a retrospective single-centre study to assess this. Methods and results We studied all patients with an isolated P/S problem in an HV-P/S lead, treated with an additional P/S lead, in our institution. The need for further invasive intervention for a lead-related complication, or death during follow-up, was assessed. From 2000 to 2008, 45 patients were treated with an additional P/S lead. Mean follow-up was 78 +/- 38 months from original device implantation and 28 +/- 17 months following implantation of the additional lead. During follow-up, three patients required an invasive intervention for a lead-related problem. All were successfully treated with lead extraction and device re-implantation. There were five deaths. Following implant of an additional lead, cumulative survival from further lead defects after 6 months, 1, 2, and 3 years was 100, 100, 93, and 87%, respectively. Conclusion In the treatment of an isolated P/S problem in an HV-P/S lead, the placement of an additional P/S lead is a safe management strategy, at least in the short term.
引用
收藏
页码:522 / 526
页数:5
相关论文
共 50 条
  • [31] Outcomes and costs associated with two different lead-extraction approaches: a single-centre study
    Gaubert, Melanie
    Giorgi, Roch
    Franceschi, Frederic
    Koutbi-Franceschi, Linda
    Gitenay, Edouard
    Maille, Baptiste
    Deharo, Jean-Claude
    EUROPACE, 2017, 19 (10): : 1710 - 1716
  • [32] Outcomes of transvenous lead extraction in patients with lead perforation: A single-center experience
    Zhou, Xu
    Ze, Feng
    Li, Ding
    Wang, Long
    Guo, Jihong
    Li, Xuebin
    CLINICAL CARDIOLOGY, 2020, 43 (04) : 386 - 393
  • [33] Performance of Medtronic attain StarFix 4195 lead: a 2-year single-centre experience
    De Cleen, Dieter
    Dilling-Boer, Dagmara
    Vijgen, Johan
    Herbots, Lieven
    Geukens, Robert
    ACTA CARDIOLOGICA, 2009, 64 (05) : 704 - 704
  • [34] Single-centre experience with coronary sinus lead stability and tong-term pacing parameters
    Bulava, Alan
    Luki, Jan
    EUROPACE, 2007, 9 (07): : 523 - 527
  • [35] Incidence of Premature Lead Failure in 2088 Tendril™ Pacing Leads: A Single Centre Experience
    Segan, Louise
    Samuel, Rohit
    Lim, Michael
    Ridley, Daryl
    Sen, Jonathan
    Perrin, Mark
    HEART LUNG AND CIRCULATION, 2021, 30 (07): : 986 - 995
  • [36] Prognostic value of renal failure in patients undergoing transvenous lead extraction: single centre experience and systematic review of the literature
    Massaro, Giulia
    Spadotto, Alberto
    Canovi, Luca
    Martignani, Cristian
    Ziacchi, Matteo
    Angeletti, Andrea
    Galie, Nazzareno
    Boriani, Giuseppe
    Biffi, Mauro
    Diemberger, Igor
    EXPERT REVIEW OF MEDICAL DEVICES, 2022, 19 (11) : 905 - 913
  • [37] Large, Single Center Experience with Single Sheat Lead Extraction
    Platou, E. S.
    Knutsen, T. M.
    Steen, T.
    CARDIOLOGY, 2013, 125 : 22 - 22
  • [38] Large, single center experience with single sheat lead extraction
    Platou, E. S.
    Knutsen, T. M.
    Steen, T.
    EUROPEAN HEART JOURNAL, 2013, 34 : 593 - 593
  • [39] Predictors for adverse events during cardiac lead extraction - Experience from a large single centre
    Xiao, Zengli
    He, Jinshan
    Du, Anqi
    Yang, Dandan
    An, Youzhong
    Li, Xuebin
    INTERNATIONAL JOURNAL OF CARDIOLOGY, 2023, 371 : 167 - 174
  • [40] Lead Extraction for Treatment of Cardiac Device Infection: A 20-Year Single Centre Experience
    Gomes, Sean
    Cranney, Gregory
    Bennett, Michael
    Giles, Robert